Dade® Innovin® 10445705Recalls & reported problems

Siemens Healthcare Diagnostics Products GmbH · For use in prothrombin time (PT) determinations and prothrombin time-based assays in citrated human plasma · Class 2

All models of Dade Innovin (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768003790), received between Feb 1, 2018 and Jul 1, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED DISCORDANT PROTHROMBIN TIME (PT) RESULTS MEASURED WITH THE BFT II ANALYZER COMPARED TO THE CA-500 SYSTEM AND ABOUT DISCREPANCIES OF RESULTS WITHIN RUN ON BFT II ANALYZER WHEN SAMPLE WAS MEASURED IN TWO-FOLD DETERMINATION. ONLY ONE SPECIFIC PATIENT (DESCRIBED AS PATIENT X) IS AFFECTED. ERRONEOUS RESULTS WERE NOT PROVIDED TO THE PHYSICIAN(S). THE REPEAT RESULTS ON THE CA-500 SYSTEM WERE REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT PROTHROMBIN TIME RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19648653 ↗
  • MalfunctionFeb 1, 2018High Test Results; High Test Results

    A DISCORDANT HIGH PROTHROMBIN TIME (PT)/INTERNATIONAL NORMALIZED RATIO (INR) PATIENT RESULT (PT EXTENDED >160 SECONDS / INR >16.0) WAS GENERATED ON A SYSMEX CA-620 ANALYZER AND REPORTED TO THE PHYSICIAN. THE PHYSICIAN DID NOT QUESTION THE RESULT. PT QUALITY CONTROL (QC) WAS WITHIN RANGE AT THE TIME OF TESTING. THE PATIENT WAS SENT TO THE EMERGENCY DEPARTMENT AND A REDRAW SAMPLE WAS TESTED ON A SYSMEX CA-1500 ANALYZER AT THE HOSPITAL LAB. THIS PT/INR PATIENT RESULT (PT 20.7/ INR 2.0) WAS WITHIN THE THERAPEUTIC RANGE FOR ANTICOAGULANT THERAPY AND REPORTED TO THE PHYSICIAN. THE PATIENT DID NOT RECEIVE ANY ADDITIONAL TREATMENT AND WAS DISCHARGED. THE CUSTOMER REPEATED THE ORIGINAL PATIENT SAMPLE THE NEXT DAY (AFTER REMIX/RE-CENTRIFUGATION) ON THE SAME SYSMEX CA-620 ANALYSER AND A LOWER PT/INR RESULT WAS GENERATED (PT 14.4/INR 1.47). THIS RESULT WAS NOT REPORTED TO THE PHYSICIAN. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HIGH PT/INR PATIENT RESULT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7233793 ↗