Dade® Innovin® 10445705Recalls & reported problems
Siemens Healthcare Diagnostics Products GmbH · For use in prothrombin time (PT) determinations and prothrombin time-based assays in citrated human plasma · Class 2
All models of Dade Innovin (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768003790), received between Feb 1, 2018 and Jul 1, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED DISCORDANT PROTHROMBIN TIME (PT) RESULTS MEASURED WITH THE BFT II ANALYZER COMPARED TO THE CA-500 SYSTEM AND ABOUT DISCREPANCIES OF RESULTS WITHIN RUN ON BFT II ANALYZER WHEN SAMPLE WAS MEASURED IN TWO-FOLD DETERMINATION. ONLY ONE SPECIFIC PATIENT (DESCRIBED AS PATIENT X) IS AFFECTED. ERRONEOUS RESULTS WERE NOT PROVIDED TO THE PHYSICIAN(S). THE REPEAT RESULTS ON THE CA-500 SYSTEM WERE REPORTED, AS THE CORRECT RESULTS, TO THE PHYSICIAN(S). THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT PROTHROMBIN TIME RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19648653 ↗MalfunctionFeb 1, 2018High Test Results; High Test Results
A DISCORDANT HIGH PROTHROMBIN TIME (PT)/INTERNATIONAL NORMALIZED RATIO (INR) PATIENT RESULT (PT EXTENDED >160 SECONDS / INR >16.0) WAS GENERATED ON A SYSMEX CA-620 ANALYZER AND REPORTED TO THE PHYSICIAN. THE PHYSICIAN DID NOT QUESTION THE RESULT. PT QUALITY CONTROL (QC) WAS WITHIN RANGE AT THE TIME OF TESTING. THE PATIENT WAS SENT TO THE EMERGENCY DEPARTMENT AND A REDRAW SAMPLE WAS TESTED ON A SYSMEX CA-1500 ANALYZER AT THE HOSPITAL LAB. THIS PT/INR PATIENT RESULT (PT 20.7/ INR 2.0) WAS WITHIN THE THERAPEUTIC RANGE FOR ANTICOAGULANT THERAPY AND REPORTED TO THE PHYSICIAN. THE PATIENT DID NOT RECEIVE ANY ADDITIONAL TREATMENT AND WAS DISCHARGED. THE CUSTOMER REPEATED THE ORIGINAL PATIENT SAMPLE THE NEXT DAY (AFTER REMIX/RE-CENTRIFUGATION) ON THE SAME SYSMEX CA-620 ANALYSER AND A LOWER PT/INR RESULT WAS GENERATED (PT 14.4/INR 1.47). THIS RESULT WAS NOT REPORTED TO THE PHYSICIAN. THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT HIGH PT/INR PATIENT RESULT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7233793 ↗
Other TEST, TIME, PROTHROMBIN devices
- COAGUCHEK XS PT TEST 6 STRIP 4625374160Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN3 deaths1,681 reports
- CoaguChek XS PT strips 4625315160Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN604 reports
- CoaguChek XS Patient Self Testing Care Kit 03666794001Roche Diagnostics GmbH · Test, Time, Prothrombin212 reports
- CoaguChek XS Patient Self Testing Care Kit 4837738001Roche Diagnostics GmbH · TEST, TIME, PROTHROMBIN134 reports
- i-STAT 03P89-24Abbott Point Of Care Inc. · TEST, TIME, PROTHROMBIN127 reports
- HemosIL ReadiPlasTin (20 mL) 00020301400Instrumentation Laboratory Company · TEST, TIME, PROTHROMBIN2 recalls2 reports
- STA® - Néoplastine® CI 10 00666Diagnostica Stago · TEST, TIME, PROTHROMBIN1 recall0 reports
- STA® - Néoplastine® CI Plus 10 00667Diagnostica Stago · TEST, TIME, PROTHROMBIN1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page