Dimension Vista® Flex® reagent cartridge E2 10489099Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · LOCI Estradiol Flex® reagent cartridge, 320 Tests · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00842768026461), received between Aug 15, 2018 and Dec 5, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 5, 2022High Test Results; High Test Results
A DISCORDANT FALSELY ELEVATED LOCI ESTRADIOL (E2) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION VISTA® 500 INTELLIGENT LAB SYSTEM. THE FALSELY ELEVATED RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPROCESSED ON THE SAME INSTRUMENT, TWO ALTERNATE SIEMENS INSTRUMENTS AND ON A NON-SIEMENS SYSTEM. THE REPROCESSED RESULTS FROM THE ALTERNATE INSTRUMENTS WERE LOWER THAN THE FALSELY ELEVATED RESULT AND CONSIDERED CORRECT. THE RESULT FROM THE NON-SIEMENS SYSTEM WAS PROVIDED TO THE PHYSICIAN(S) ON A CORRECTED REPORT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY ELEVATED E2 RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15917259 ↗MalfunctionAug 15, 2018High Test Results; High Test Results
A DISCORDANT ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED ON A PATIENT'S SAMPLE ON THE DIMENSION VISTA 1500 SYSTEM. THE SAME SAMPLE WAS TESTED BY TWO ALTERNATE NON-SIEMENS METHODOLOGIES AND LOWER RESULTS WERE OBTAINED CONSISTENT WITH PHYSICIAN EXPECTATIONS. THERE ARE NO REPORTS OF TREATMENT CHANGE OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ELEVATED E2 RESULT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7787093 ↗
Other ENZYME IMMUNOASSAY, INHIBIN-A devices
- Dimension Vista® MULTI 1 SDIL 10469971Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, INHIBIN-A6 reports
- Access Inhibin A A36097Beckman Coulter, Inc. · Enzyme Immunoassay, Inhibin-A3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page