Access Inhibin A A36097Recalls & reported problems
Beckman Coulter, Inc. · The Access Inhibin A assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of dimeric inhibin A in human serum and plasma (heparin and EDTA) as an aid in the diagnosis and monitoring of various hormonal reproductive disorders using the Access Immunoassay Systems. · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 15099590203047), received between Dec 27, 2017 and Apr 25, 2021 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Low Test Results3
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2021Low Test Results; Low Test Results
ON 29MARCH2021 THE CUSTOMER REPORTED THAT ON (B)(6) 2021, ERRONEOUS LOW INHIBIN A (ACCESS INHIBIN A, PART NUMBER A36097 AND LOT NUMBER 922981) RESULTS WERE GENERATED ON THE CUSTOMER'S ACCESS 2 (ACCESS 2 IMMUNOASSAY ANALYZER, SERIAL NUMBER (B)(4)) FOR SEVERAL PATIENTS. CUSTOMER DID NOT INDICATE WHETHER THE ERRONEOUS INHIBIN A RESULTS WERE RELEASED FROM THE LABORATORY. THE CUSTOMER DID NOT REPORT INJURY OR CHANGE TO PATIENT TREATMENT AS A RESULT OF THE ERRONEOUS INHIBIN A RESULTS. SYSTEM PERFORMANCE INDICATORS SUCH AS CALIBRATION AND QUALITY CONTROL WERE PASSING WITHIN SPECIFICATIONS AT THE TIME OF THE EVENT. ALTHOUGH THE CUSTOMER REPORTED ISSUES WITH THE SYSTEM CHECK, PASSING SYSTEM CHECK WAS OBTAINED AFTER FAILING SYSTEM CHECK WITH NO INDICATION OF INTERVENTIONS PERFORMED BY CUSTOMER. SAMPLES WERE COLLECTED OFFSITE IN 12X75MM SERUM SEPARATOR TUBES; OTHER SAMPLE HANDLING INFORMATION SUCH AS SAMPLE CLOTTING TIME AND CENTRIFUGATION SPEED AND TIME WERE NOT PROVIDED. CUSTOMER NOTED THE SAMPLES WERE ALIQUOTED AND FROZEN FOR TRANSPORTATION AND THAWED AND TESTED UPON RECEIPT TO THE LABORATORY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11721842 ↗InjurySep 27, 2018Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING LOW REPRODUCTIVE TESTING RESULTS INCLUDING INHIBIN A (ACCESS INHIBIN A), BETA HUMAN CHORIONIC GONADOTROPIN (ACCESS TOTAL BHCG (5TH IS)), UNCONJUGATED ESTRIOL (ACCESS UE3) AND ALPHA-FETOPROTEIN (ACCESS AFP) FOR ONE (1) PATIENT INVOLVING THE UNICEL DXI 800 ACCESS CLINICAL SYSTEM SERIAL NUMBER (B)(4). THE CUSTOMER REANALYZED THE PATIENT'S SAMPLE ON THE SAME UNICEL DXI 800 ACCESS CLINICAL SYSTEM TWO (2) ADDITIONAL TIMES AND OBTAINED REPRODUCIBLE, HIGHER RESULTS. THE INITIAL LOW RESULTS WERE RELEASED FROM THE LABORATORY. THE PATIENT IN QUESTION UNDERWENT AN AMNIOCENTESIS TO RULE OUT TRISOMY 18 DUE TO THE LOW RESULTS OBTAINED. THE RESULTS OF THAT TESTING WERE NORMAL. THERE WAS NO ADDITIONAL INJURY OR REPORT OF DEATH ASSOCIATED WITH THIS INCIDENT. QUALITY CONTROL (QC) AND CALIBRATIONS WERE PERFORMING WITHIN THE ASSAY'S SPECIFICATIONS AT THE TIME OF THE INCIDENT. THERE WERE NO ERROR MESSAGES POSTED TO THE EVENT LOG AT THE TIME OF THE EVENT. THE CUSTOMER DID NOT PROVIDE SPECIFIC INFORMATION REGARDING THE COLLECTION OF THE PATIENT'S SAMPLE. CENTRIFUGATION TIME AND SPEED WERE NOT SUPPLIED BY THE CUSTOMER FOR THIS INCIDENT. NO ISSUES WITH SAMPLE INTEGRITY WERE R
Patient problems: No Code Available
MAUDE report 7914487 ↗MalfunctionDec 27, 2017Low Test Results; Low Test Results
THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE, LOW DIMERIC INHIBIN A (ACCESS INHIBIN A) RESULTS FOR FOUR PATIENT SAMPLES ON THE LABORATORY'S UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM (SERIAL NUMBER (B)(4)). THE INITIAL RESULTS WERE LOWER THAN THE CUSTOMER EXPECTED. THE CUSTOMER WAS ANALYZING THE SAMPLES AS PART OF A MATERNAL SCREEN WHICH INCLUDED ALPHA-FETOPROTEIN (ACCESS AFP) UNCONJUGATED ESTRIOL (ACCESS UNCONJUGATED ESTRIOL) BETA HUMAN CHORIONIC GONADOTROPIN (ACCESS TOTAL BHCG (5TH IS)) AND ACCESS INHIBIN A. THE INITIAL RESULTS FROM THE FOUR ASSAYS RECOVERED BELOW THE CUSTOMER'S EXPECTATIONS FOR THE FOUR PATIENT SAMPLES. AN ADDITIONAL SEVEN PATIENT SAMPLES ANALYZED AT THE SAME TIME RECOVERED WITH INITIAL RESULTS BELOW THE CUSTOMER'S EXPECTATIONS. THE INITIAL RESULTS WERE NOT RELEASED FROM THE LABORATORY. THERE WAS NO CHANGE TO PATIENT CARE OR TREATMENT WHICH OCCURRED IN ASSOCIATION WITH NON-REPRODUCIBLE ACCESS PATIENT RESULTS. THE PATIENTS' SAMPLES WERE REANALYZED ON THE SAME UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM AND RECOVERED WITH HIGHER RESULTS, WITHIN THE CUSTOMER'S EXPECTED VALUES. CALIBRATIONS AND QUALITY CONTROL (QC) WERE PERFORMING WITHIN ASSAY AND INSTRUMENT SPECIFIC
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148746 ↗
FDA clearance and classification
The FDA regulates product code NDR (enzyme immunoassay, inhibin-a) under 21 CFR 862.1260 as a Class I device. 21 CFR 862.1260 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Access Inhibin A A36097
Has the Access Inhibin A A36097 been recalled?
No FDA recall record in our data names this exact device identifier (DI 15099590203047), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Access Inhibin A A36097?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 27, 2017 and Apr 25, 2021. In those reports, the most common specific device problem is “Low Test Results” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Access Inhibin A A36097?
None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Access Inhibin A A36097 FDA approved?
Devices of this type (product code NDR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Beckman Coulter, Inc.
Who makes the Access Inhibin A A36097, and is it still sold?
The FDA's device database lists Beckman Coulter, Inc. as the company responsible for it (brand Access Inhibin A, model A36097). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Enzyme Immunoassay, Inhibin-A devices
- Dimension Vista® MULTI 1 SDIL 10469971Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, INHIBIN-A6 reports
- Dimension Vista® Flex® reagent cartridge E2 10489099Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, INHIBIN-A2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page