Dimension Vista® MULTI 1 SDIL 10469971Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · Multi 1 Sample Diluent, 12x2mL · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00842768030314), received between Mar 26, 2018 and Mar 28, 2018 — classified as 6 malfunctions.
The most frequently reported specific device problem is “High Test Results” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results6
Patient problems
- No Known Impact Or Consequence To Patient6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2018High Test Results; High Test Results
A FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED ON A CAP SURVEY SAMPLE ON THE DIMENSION VISTA 500 INSTRUMENT. THE RESULT WAS SCORED AS UNACCEPTABLE BY THE CAP. THERE IS NO INDICATION THAT PATIENT SAMPLES HAVE BEEN AFFECTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7376904 ↗MalfunctionMar 26, 2018High Test Results; High Test Results
A FALSELY ELEVATED ESTRADIOL (E2) RESULT WAS OBTAINED ON A CAP SURVEY SAMPLE ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULT WAS SCORED AS UNACCEPTABLE BY THE CAP. THE CUSTOMER STATED THAT NO PATIENT SAMPLES HAVE BEEN AFFECTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371142 ↗MalfunctionMar 26, 2018High Test Results; High Test Results
DISCORDANT ELEVATED ESTRADIOL (E2) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON THE DIMENSION VISTA 500 INSTRUMENT. THE RESULTS WERE REPORTED TO THE PHYSICIAN(S) AND QUESTIONED. THE ORIGINAL SAMPLES WHICH HAD BEEN TESTED BY AUTO-DILUTION WERE REPEATED BY MANUAL DILUTION AND LOWER RESULTS WERE OBTAINED AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT AUTO-DILUTION RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371141 ↗MalfunctionMar 26, 2018High Test Results; High Test Results
DISCORDANT ELEVATED ESTRADIOL (E2) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE ORIGINAL SAMPLES WHICH HAD BEEN TESTED BY AUTO-DILUTION WERE REPEATED BY MANUAL DILUTION AND LOWER RESULTS WERE OBTAINED AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT AUTO-DILUTION RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371138 ↗MalfunctionMar 26, 2018High Test Results; High Test Results
DISCORDANT ELEVATED ESTRADIOL (E2) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON THE DIMENSION VISTA 1500 INSTRUMENT. THE RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE ORIGINAL SAMPLES WHICH HAD BEEN TESTED BY AUTO-DILUTION WERE REPEATED BY MANUAL DILUTION AND LOWER RESULTS WERE OBTAINED AND REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT AUTO-DILUTION RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371136 ↗MalfunctionMar 26, 2018High Test Results; High Test Results
DISCORDANT ELEVATED ESTRADIOL (E2) RESULTS WERE OBTAINED ON PATIENT SAMPLES ON THE DIMENSION VISTA 500 INSTRUMENT. THE RESULTS WERE REPORTED TO THE PHYSICIAN(S) WHO QUESTIONED THE RESULTS. THE ORIGINAL SAMPLES WHICH HAD BEEN TESTED BY AUTO-DILUTION WERE REPEATED BY MANUAL DILUTION AND LOWER RESULTS WERE OBTAINED. CORRECTED RESULTS WERE REPORTED. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT AUTO-DILUTION RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7371133 ↗
Other ENZYME IMMUNOASSAY, INHIBIN-A devices
- Access Inhibin A A36097Beckman Coulter, Inc. · Enzyme Immunoassay, Inhibin-A3 reports
- Dimension Vista® Flex® reagent cartridge E2 10489099Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, INHIBIN-A2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.