Skip to content

TSI AL-0103Recalls & reported problems

Tedan Surgical Innovations, L.L.C. · RING CLAMP · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00843059100945), received between Apr 12, 2024 and May 15, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 5 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking6

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON 18 MARCH 2024. IT WAS REPORTED THAT SEVERAL AL-0103 RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON 17 APRIL 2024 THE AL-0103 RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:1, P/N: AL-0103, LOT:1222) & (QTY:5, P/N: AL-0103, LOT:0423). THIS REPORT IS 1 OUT OF 5 FOR LOT 0423. THIS EVENT WAS ORIGINALLY REPORTED UNDER 3012120772-2024-00009. THIS REPORT WAS FILED TO DOCUMENT THE DIFFERENT LOT NUMBERS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19326157 ↗
  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON (B)(6) 2024. IT WAS REPORTED THAT SEVERAL AL-0103 RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON 17 APRIL 2024 THE AL-0103 RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:(B)(4), P/N: (B)(6), LOT:1222) & (QTY:(B)(4), P/N: (B)(6), LOT:0423). THIS REPORT IS 5 OUT OF 5 FOR LOT 0423. THIS EVENT WAS ORIGINALLY REPORTED UNDER 3012120772-2024-00009. THIS REPORT WAS FILED TO DOCUMENT THE DIFFERENT LOT NUMBERS RECEIVED. ON 30 MAY 2024 THE EVALUATION WAS COMPLETED AND AS A RESULT THIS SUPPLEMENTAL REPORT WAS SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19326064 ↗
  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON (B)(6) 2024. IT WAS REPORTED THAT SEVERAL AL-0103 RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON 17 APRIL 2024 THE AL-0103 RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:1, P/N: AL-0103, LOT:1222) & (QTY:5, P/N: AL-0103, LOT:0423). THIS REPORT IS 4 OUT OF 5 FOR LOT 0423. THIS EVENT WAS ORIGINALLY REPORTED UNDER 3012120772-2024-00009. THIS REPORT WAS FILED TO DOCUMENT THE DIFFERENT LOT NUMBERS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19326061 ↗
  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON (B)(6) 2024. IT WAS REPORTED THAT SEVERAL (B)(6) RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON (B)(6) 2024 THE (B)(6) RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:(B)(4), P/N: (B)(6), LOT:1222) & (QTY:(B)(4), P/N: (B)(6), LOT:0423). THIS REPORT IS 3 OUT OF 5 FOR LOT 0423. THIS EVENT WAS ORIGINALLY REPORTED UNDER 3012120772-2024-00009. THIS REPORT WAS FILED TO DOCUMENT THE DIFFERENT LOT NUMBERS RECEIVED. ON 30 MAY 2024 THE EVALUATION WAS COMPLETED AND AS A RESULT THIS SUPPLEMENTAL REPORT WAS SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19326054 ↗
  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON (B)(6) 2024. IT WAS REPORTED THAT SEVERAL AL-0103 RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON 17 APRIL 2024 THE AL-0103 RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:1, P/N: AL-0103, LOT:1222) & (QTY:5, P/N: AL-0103, LOT:0423). THIS REPORT IS 2 OUT OF 5 FOR LOT 0423. THIS EVENT WAS ORIGINALLY REPORTED UNDER 3012120772-2024-00009. THIS REPORT WAS FILED TO DOCUMENT THE DIFFERENT LOT NUMBERS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19326046 ↗
  • MalfunctionApr 12, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 18 MARCH 2024, SEASPINE WAS MADE AWARE OF AN INTRA-OPERATIVE MALFUNCTION INVOLVING THE ALIF RETRACTOR SYSTEM. THE CASE OCCURRED ON (B)(6) 2024. IT WAS REPORTED THAT SEVERAL AL-0103 RING CLAMPS FROZE MID-CASE WHICH MADE CHANGING THE BLADES DIFFICULT. THIS MALFUNCTION CAUSED A SURGICAL DELAY OF APPROXIMATELY 45-60 MINUTES. THE CASE WAS COMPLETED AFTER THE PROPER RETRACTION WAS ACHIEVED. NO PATIENT INJURY WAS REPORTED. ON (B)(6) 2024 THE AL-0103 RING CLAMPS WERE RECEIVED FOR EVALUATION. THE RETURNED PARTS ARE AS FOLLOW: (QTY:1, P/N: AL-0103, LOT:1222) & (QTY:5, P/N: AL-0103, LOT:0423). THIS REPORT IS FOR LOT 1222.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19102136 ↗

Other Operating Room Accessories Table Tray devices

All Operating Room Accessories Table Tray devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.