Prodisc C SK PDSM5Recalls & reported problems
Centinel Spine, LLC · Prodisc C SK, Medium, 5mm · Class 3 · Implant
All models of Prodisc C SK (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00843193113856), received between Aug 14, 2024 and Nov 12, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
4 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 18, 2024Open, Classified
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
Z-3212-2024 · Root cause: Process control · Recall details · FDA record ↗
- Jul 18, 2024Open, Classified
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
Z-3213-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Malposition of Device2
- Adverse Event Without Identified Device or Use Problem1
- Osseointegration Problem1
Patient problems
- Paralysis2
- Inadequate Pain Relief1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 12, 2025Malposition of Device; Malposition of Device
A PATIENT WAS IMPLANTED WITH TWO PDC SK DEVICES AT C5-6 AND C6-7 ON (B)(6) 2025. PATIENT STATED THAT WHEN SHE AWOKE FROM IMPLANTATION SURGERY, SHE WAS PARALYZED WITH ONLY THE ABILITY TO MOVE HER EYES. PATIENT STATED THAT THE SURGEON TOLD HER THAT THE IMPLANT WAS PUSHED TOO FAR POSTERIOR, AND THAT THE IMPLANT HAD SMASHED HER SPINE, AND HE HAD TO TAKE IT OUT. PATIENT SPENT ONE WEEK IN THE ICU, FOLLOWED BY A MONTH AT A SPINE REHAB CENTER, PATIENT CONTINUES TO GO TO PT THREE TIMES A WEEK AND HAS BEEN UNABLE TO RETURN TO WORK. PATIENT HAD TO RELEARN HOW TO WALK AND CONTINUES TO HAVE ARM AND HAND NUMBNESS.
Patient problems: Paralysis
MAUDE report 23538309 ↗InjuryNov 12, 2025Malposition of Device; Malposition of Device
A PATIENT WAS IMPLANTED WITH TWO PDC SK DEVICES AT C5-6 AND C6-7 ON (B)(6) 2025. PATIENT STATED THAT WHEN SHE AWOKE FROM IMPLANTATION SURGERY, SHE WAS PARALYZED WITH ONLY THE ABILITY TO MOVE HER EYES. PATIENT STATED THAT THE SURGEON TOLD HER THAT THE IMPLANT WAS PUSHED TOO FAR POSTERIOR, AND THAT THE IMPLANT HAD SMASHED HER SPINE, AND HE HAD TO TAKE IT OUT. PATIENT SPENT ONE WEEK IN THE ICU, FOLLOWED BY A MONTH AT A SPINE REHAB CENTER, PATIENT CONTINUES TO GO TO PT THREE TIMES A WEEK AND HAS BEEN UNABLE TO RETURN TO WORK. PATIENT HAD TO RELEARN HOW TO WALK AND CONTINUES TO HAVE ARM AND HAND NUMBNESS.
Patient problems: Paralysis
MAUDE report 23538306 ↗MalfunctionSep 18, 2025Osseointegration Problem; Osseointegration Problem
A PATIENT WAS IMPLANTED WITH A PRODISC C SK DEVICE AT C5-6 ON (B)(6) 2025. AT THE 2 WEEK FOLLOW UP, THE IMAGING SHOWED THAT THE IMPLANT MAY HAVE SLIGHTLY SUBSIDED AND KYPHOSED. THE PATIENT IS EXPERIENCING NO SYMPTOMS AND IS DOING WELL. THERE IS NO INDICATION AT PRESENT THAT ANY INTERVENTION WILL BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23084969 ↗InjuryAug 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT WAS IMPLANTED WITH A PRODISC C SK DEVICE ON (B)(6) 2024. FOLLOWING IMPLANTATION THE PATIENT HAD PERSISTENT ONGOING PAIN. NO MALFUNCTION, MIGRATION, OR PROBLEM WITH THE DEVICE. THE SURGEON REMOVED THE IMPLANT ON (B)(6) 2024 AND FUSED THE PATIENT WITH A PEEK SPACER AND PLATE.
Patient problems: Inadequate Pain Relief
MAUDE report 19982946 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.
- Z-2051-2026 · Mar 20, 2026 · Open, ClassifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
- Z-2050-2026 · Mar 20, 2026 · Open, ClassifiedProdisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc replacement.
- Z-1426-2026 · Dec 31, 2025 · Open, ClassifiedBrand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
- Z-1406-2021 · Mar 5, 2021 · TerminatedProdisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for
Other Prosthesis, Intervertebral Disc devices
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall12 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall4 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page