Skip to content

Simplify® Cervical Artificial Disc (REF SM-5)Recalls & reported problems

Nuvasive, Inc. · Simplify® Cervical Artificial Disc Size SM, Height 5 · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843285100023), received between Apr 14, 2023 and Apr 13, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Implant subsidence” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 20, 2022Open, Classified

    Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.

    Z-1453-2022 · Root cause: Labeling Change Control · FDA record ↗

Most reported problems

Device problems

  • Implant subsidence1
  • Migration1
  • Off-Label Use1

Patient problems

  • Neck Pain2
  • Extraskeletal ossification1
  • Limb Pain1
  • No Clinical Signs, Symptoms or Conditions1
  • Osteolysis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2026Implant subsidence

    REVISION OF 1 LEVEL FUSION AND 1 LEVEL CTDR TO 1 LEVEL FUSION. FUSION C5-6 CTDR C6-7. REMOVAL OF SIMPLIFY DISC DUE TO COLLAPSED VERTEBRAE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24866063 ↗
  • InjuryApr 22, 2024Off-Label Use; Off-Label Use

    IN (B)(6) 2020 A PATIENT UNDERWENT A C4-5 CERVICAL TOTAL DISC REPLACEMENT (CTDR) FROM ANOTHER MANUFACTURER. IN (B)(6) 2023 A REVISION TOOK PLACE WHERE THE UNKNOWN C4-5 CTDR WAS REVISED TO A SIMPLIFY DISC AND AN ANTERIOR CERVICAL DISCECTOMY FUSION WAS COMPLETED AT C5-C6. ON (B)(6) 2023 THE PATIENT COMPLAINED OF NECK PAIN AND RADIOGRAPHS IDENTIFIED PERI-IMPLANT LUCENCY AROUND AND A REVISION IS BEING PLANNED.

    Patient problems: Extraskeletal ossification; Neck Pain; Osteolysis

    MAUDE report 19159397 ↗
  • MalfunctionApr 14, 2023Migration; Migration

    ON (B)(6) 2022 , A CT SCAN WAS DONE DUE TO NO RELIEF FROM CONTINUED NECK AND SHOULDER PAIN. NOTING THE DISC SLIPPED OUT OF PLACE AGAIN AT C6/7 AS WELL AS THE DISC AT C5/6 AND WILL HAVE TO HAVE SURGERY AGAIN. ON (B)(6) 2022, A REVISION TOOK PLACE REMOVING BOTH MIGRATED DISCS AT C5/6 AND C6/7 AND REPLACED WITH ALLOGRAFT FOR FUSION. (2 OF 2 (B)(4)).

    Patient problems: Limb Pain; Neck Pain

    MAUDE report 16747776 ↗

Other Prosthesis, Intervertebral Disc devices

All Prosthesis, Intervertebral Disc devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.