1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843193113979), received between Feb 6, 2025 and Mar 30, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Migration” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (P070001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 20, 2026Open, Classified

    Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm.

    Z-2050-2026 · Root cause: Labeling mix-ups · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration2
  • Osseointegration Problem2
  • Device Misassembled During Manufacturing /Shipping1

Patient problems

  • Spinal Column Injury2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Device Misassembled During Manufacturing /Shipping

    A PATIENT (20YO, MALE) WAS IMPLANTED WITH TWO PRODISC C SK DEVICES ON (B)(6) 2026. BOTH IMPLANTS WERE LABELED AS 5MM DEVICES BUT ONCE PLACED IT WAS EVIDENT THAT ONE IMPLANT WAS TALLER THAN THE OTHER. INVESTIGATION FOUND THAT THE ONE IMPLANT WAS 6MM BUT LABELED AS 5MM. THE SURGEON STATED TO LEAVE THE IMPLANT ALONE. HE SAID THE HEIGHT IS FINE AND THAT THE PATIENT COULD TAKE A 5 OR 6MM DEVICE. HE SAID HE WOULD HAVE PULLED IT OUT IF HE THOUGHT IT WOULD HURT THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715993 ↗
  • MalfunctionFeb 6, 2025Osseointegration Problem; Migration; Osseointegration Problem; Migration

    A PATIENT WAS IMPLANTED WITH A PDC VIVO DEVICE AT C4-5 AND TWO PDC SK DEVICES AT C5-6 AND C6-7 ON (B)(6)2024. IT WAS FOUND THAT THE VIVO IMPLANT HAD MIGRATED ANTERIORALLY, THE SK IMPLANT AT C5-6 WAS KYPHOTIC, AND THE SK IMPLANT AT C6-7 HAD MIGRATED POSTERIORLY AND SUBSIDED. A SECOND SURGEON RECOMMENDED REMOVING ALL THREE LEVELS AND FUSING C6-7 AND CONVERTING C4-5 AND C5-6 TO PDC ORIGINAL. NO ADDITIONAL INFORMATION WAS PROVIDED AND IT IS UNKNOWN IF THE PATIENT RECEIVED ANY ADDITIONAL SURGICAL INTERVENTION.

    Patient problems: Spinal Column Injury

    MAUDE report 21327002 ↗
  • MalfunctionFeb 6, 2025Osseointegration Problem; Migration; Osseointegration Problem; Migration

    A PATIENT WAS IMPLANTED WITH A PDC VIVO DEVICE AT C4-5 AND TWO PDC SK DEVICES AT C5-6 AND C6-7 ON (B)(6) 2024. IT WAS FOUND THAT THE VIVO IMPLANT HAD MIGRATED ANTERIORALLY, THE SK IMPLANT AT C5-6 WAS KYPHOTIC, AND THE SK IMPLANT AT C6-7 HAD MIGRATED POSTERIORALLY AND SUBSIDED. A SECOND SURGEON RECOMMENDED REMOVING ALL THREE LEVELS AND FUSING C6-7 AND CONVERTING C4-5 AND C5-6 TO PDC ORIGINAL. NO ADDITIONAL INFORMATION WAS PROVIDED AND IT IS UNKNOWN IF THE PATIENT RECEIVED ANY ADDITIONAL SURGICAL INTERVENTION.

    Patient problems: Spinal Column Injury

    MAUDE report 21326996 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P070001. One submission often covers many models, so check the product and lot details in each record.

  • Z-2051-2026 · Mar 20, 2026 · Open, Classified
    Prodisc C SK U.S. IMPLANT EXTRA LARGE 6MM. Model Number: PDSXL6. Total cervical disc replacement.
  • Z-1426-2026 · Dec 31, 2025 · Open, Classified
    Brand Name: Prodisc Product Name: Prodisc C SK Model/Catalog Number: PDSL6 Software Version: N/A Product Description: Total cervical disc replacement Component: No
  • Z-1406-2021 · Mar 5, 2021 · Terminated
    Prodisc C - prosthesis, intervertebral disc - Product Usage: indicated in skeletally mature patients for reconstruction of the disc from C3-C7 following single-level discectomy for

Other Prosthesis, Intervertebral Disc devices