Nihon Kohden NKV-440 (UDI 00843685100159)Recalls & reported problems
Nihon Kohden Orangemed, Inc. · Nihon Kohden NKV-440 Ventilator System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843685100159), received between May 22, 2025 and Sep 29, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Component1
- Failure to Clean Adequately1
- Insufficient Device Problem Information1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Hypoxia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Insufficient Device Problem Information
Patient problems: Hypoxia
MAUDE report 50940562 ↗MalfunctionJun 24, 2026Use of Device Problem; Failure to Clean Adequately
IT WAS REPORTED TO NKOM THAT DURING INITIATION OF NEBULIZER TREATMENT, THE VENTILATOR BEGAN GENERATING "CIRCUIT DISCONNECT" AND "CIRCUIT OBSTRUCTION" ALARMS. THE RESPIRATORY THERAPIST (RT) OBSERVED THAT THE DISPLAYED EXHALED TIDAL VOLUME (VTE) AND RESPIRATORY RATE (RR) INTERMITTENTLY FLUCTUATED BETWEEN APPROPRIATE VALUES AND ZERO, DESPITE THE PATIENT CONTINUING TO BREATHE AND TRIGGER THE VENTILATOR. AS A PRECAUTION, THE PATIENT WAS DISCONNECTED FROM THE VENTILATOR AND MANUALLY VENTILATED WITH A RESUSCITATION BAG BEFORE BEING TRANSITIONED TO ANOTHER VENTILATOR. THERE WAS NO PATIENT HARM AS THE PATIENT REMAINED STABLE BEFORE, DURING, AND AFTER THE EVENT. FOLLOW-UP INFORMATION FROM THE RT ON (B)(6) 2026, STATED THAT THE PATIENT RECEIVED DUONEB NEBULIZER TREATMENTS EVERY 6 HOURS AND WAS LAVAGE SUCTIONED AS NEEDED. TROUBLESHOOTING WAS PERFORMED, INCLUDING INSPECTION OF THE PATIENT CIRCUIT FOR DISCONNECTIONS, LEAKS, OR OCCLUSIONS, AND LAVAGE SUCTIONING OF THE PATIENT, WHICH YIELDED AN INSIGNIFICANT AMOUNT OF SECRETIONS. ADDITIONALLY, THE VENTILATOR WAS REPORTED TO EMIT AN INTERMITTENT MOTORIZED OR WHIRRING NOISE. WHEN THE PATIENT WAS PLACED ON ANOTHER VENTILATOR, THE PATIENT TOLERATED TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25615103 ↗MalfunctionMay 22, 2025Defective Component; Defective Component
IT WAS REPORTED THAT THE VENTILATOR DISPLAYED AN ¿INTERNAL AIRFLOW SENSOR FAILURE¿ AND A HIGH FIO2 ALARM WHILE VENTILATING A PATIENT. THE SET FIO2 WAS AT 50%, BUT THE MONITORED VALUE SHOWED FIO2 AT 100%. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22075428 ↗
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page