Skip to content

GEM Microvascular Anastomotic COUPLER GEM2753 IRecalls & reported problems

Synovis Micro Companies Alliance, Inc · The GEM Microvascular Anastomotic COUPLER Device rings are made of polyethylene and surgical grade stainless steel pins. A protective cover and jaw assembly protect the rings and allow for easy loading onto the Anastomotic Instrument. Both the protective cover and jaw assembly are disposable. The GEM Microvascular Anastomotic COUPLER Device is single-use and available in various sizes · Class 2 · Implant

All models of GEM Microvascular Anastomotic COUPLER (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00844735002652), received between Oct 22, 2020 and Apr 14, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Disconnection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection2
  • Detachment of Device or Device Component1
  • Difficult to Open or Close1
  • Incomplete or Inadequate Connection1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Difficult to Open or Close

    IT WAS REPORTED THAT A 2.5MM COUPLER RING MAY OPEN OR FAIL TO MAINTAIN ITS PROPER CLOSURE DURING HANDLING, PREVENTING CORRECT COUPLING OF THE SYSTEM. THIS OCCURRED DURING SURGICAL PROCEDURES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24875204 ↗
  • MalfunctionOct 22, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT TWO (2) 2.5MMCOUPLERS HAD BENT PINS. THIS WAS OBSERVED WHEN TAKING THE COUPLERS OUT OF THE PACKAGING. TO RESOLVE THE EVENT, A NEW COUPLER WAS USED TO CONTINUE PREPARATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23363468 ↗
  • MalfunctionAug 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A 2.5MM COUPLER POPPED OPEN AFTER CLOSURE AND CLAMPING. THE EVENT WAS FURTHER DESCRIBED AS "ANASTOMOSIS STRENGTH WAS NOT STRONG ENOUGH". THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20071148 ↗
  • MalfunctionJun 10, 2024Disconnection; Disconnection

    IT WAS REPORTED THAT THE RINGS FELL OUT OF THE WING JAW ASSEMBLY OF A 2.5MM COUPLER. THIS WAS OBSERVED AFTER THE PRODUCT PACKAGE WAS OPENED, PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19504510 ↗
  • MalfunctionMay 21, 2022Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT THE RING OF A 1.5MM COUPLER WAS NOT ABLE TO FULLY ATTACH AND FAILED TO COMPLETE ANASTOMOSIS. THIS ISSUE WAS IDENTIFIED DURING USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14465756 ↗
  • MalfunctionOct 22, 2020Disconnection; Disconnection

    IT WAS REPORTED THAT THE RING OF A 2.5MM COUPLER DEVICE CAME OUT OF THE CARTRIDGE. THIS ISSUE WAS IDENTIFIED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10722104 ↗

Other Device,anastomotic,microvascular devices

All Device,anastomotic,microvascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page