GEM FLOW COUPLER GEM2754 FCRecalls & reported problems
Synovis Micro Companies Alliance, Inc · The FlowCOUPLER System consists of a FlowCOUPLER Device and a FlowCOUPLER Monitor. The FlowCOUPLER Device includes a 20MHz ultrasonic Doppler transducer (probe) attached to one of the FlowCOUPLER rings, and an external lead. The FlowCOUPLER rings are made of high density polyethylene and surgical grade stainless steel pins. A protective cover and jaw assembly protect the rings and probe which allow for easy loading onto the Anastomotic Instrument. Both the protective cover and jaw assembly are disposable. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00844735006018), received between Jun 25, 2024 and Oct 17, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult to Open or Close1
- Disconnection1
- Failure to Align1
- Incomplete or Inadequate Connection1
- Material Twisted/Bent1
- Misconnection1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2025Misconnection; Misconnection
IT WAS REPORTED THAT A 3.0 MM FLOWCOUPLER RING PINS DID NOT PROPERLY LINE UP TO INSERT PROPERLY IN THE OPPOSING COUPLER RING. THIS WAS OBSERVED DURING A PERIPHERAL FREE FLAP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23327934 ↗MalfunctionOct 1, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM FLOW COUPLER WAS NOT CLOSING ALL THE WAY EVEN WHEN THE DEVICE WAS CLOSED ALL THE WAY. THE SPIKES COULD BE SEEN JUST STARTING TO PENETRATE THE OTHER SIDE; HOWEVER, IT WAS NOT NEAR TO WHERE IT NORMALLY SHOULD BE. THE ISSUE OCCURRED DURING AN INTRA-OPERATION STEP. WHEN SQUEEZING THE FLOW COUPLER, THE DEVICE WOULD NOT COMPLETELY CLOSE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23193419 ↗MalfunctionMar 6, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ONE OF THE PRONGS ON A FLOWCOUPLER SYSTEM WAS BENT. THIS WAS DISCOVERED INTRA-OP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21536695 ↗MalfunctionFeb 13, 2025Failure to Align; Failure to Align
IT WAS REPORTED THAT THE RINGS OF A 3.0MM FLOW COUPLER WERE MISALIGNED DURING A VENOUS ANASTOMOSIS PROCEDURE ON A PATIENT. THE RINGS DID CLOSE; HOWEVER, THE "PRONGS DID NOT MATCH UP PROPERLY FROM ONE SIDE OF THE JAW ASSEMBLY TO THE OTHER RECIPROCAL SIDE RECEIVER". THERE WAS NO REPORT OF PATIENT INJURY; HOWEVER, THE DEVICE WAS EXPLANTED AND REPLACED TO REDO THE ANASTOMOSIS. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21377324 ↗MalfunctionSep 27, 2024Disconnection; Incomplete or Inadequate Connection; Disconnection; Incomplete or Inadequate Connection
IT WAS REPORTED A 3.0MM FLOWCOUPLER FAILED TO HOLD THE VEIN SECURELY, REPEATEDLY SLIPPING OFF THE PRONGS. IT WAS ALSO REPORTED THAT UPON ATTEMPTING TO CLOSE THE COUPLER, THE PRONGS DID NOT ALIGN PROPERLY, PREVENTING THE COUPLER FROM CLOSING. TO RESOLVE THIS ISSUE, THE SURGEON QUICKLY OPENED ANOTHER COUPLER AND DID NOT EXPERIENCE ANY DIFFICULTIES OR ISSUES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20322180 ↗MalfunctionJun 25, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A FLOWCOUPLER PROBE CABLE WAS NOT CONNECTED AND THE RING DISCONNECTED DURING CONDITIONING. THE EVENT OCCURRED DURING AN UNKNOWN PROCESS STEP DURING AN UNSPECIFIED SURGERY. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19607348 ↗
Other Device,anastomotic,microvascular devices
- GEM Microvascular Anastomotic COUPLER GEM2753Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular44 reports
- GEM Microvascular Anastomotic COUPLER GEM2754Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular28 reports
- GEM Microvascular Anastomotic COUPLER GEM2755Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular13 reports
- GEM Microvascular Anastomotic COUPLER GEM2751Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular13 reports
- GEM Microvascular Anastomotic COUPLER GEM2752Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular10 reports
- GEM Microvascular Anastomotic COUPLER GEM2753 ISynovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular6 reports
- GEM Microvascular Anastomotic COUPLER GEM2754 ISynovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular5 reports
- GEM Microvascular Anastomotic COUPLER GEM2750Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page