GEM Microvascular Anastomotic COUPLER GEM2754Recalls & reported problems
Synovis Micro Companies Alliance, Inc · The GEM Microvascular Anastomotic COUPLER Device rings are made of polyethylene and surgical grade stainless steel pins. A protective cover and jaw assembly protect the rings and allow for easy loading onto the Anastomotic Instrument. Both the protective cover and jaw assembly are disposable. The GEM Microvascular Anastomotic COUPLER Device is single-use and available in various sizes · Class 2 · Implant
All models of GEM Microvascular Anastomotic COUPLER (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Synovis GEM coupler is a single-use ring used to join small blood vessels during reconstructive surgery, such as breast reconstruction. Reports mostly describe handling problems in the operating room: rings that would not open or close properly, pins that would not line up, couplers coming apart or falling loose in the sterile field, and rings that would not fit the instrument. There are 28 reports, with 2 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00844735002683), received between Aug 25, 2020 and Sep 15, 2026 — classified as 2 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Difficult to Open or Close” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Open or Close8
- Disconnection5
- Detachment of Device or Device Component2
- Device Dislodged or Dislocated2
- Failure to Disconnect2
- Fitting Problem2
- Material Twisted/Bent2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- No Consequences Or Impact To Patient2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Injury1
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Misconnection; Failure to Disconnect
It was reported that the rings of two (2) 3.0 mm Couplers failed to release completely from the jaws. Additionally, in one coupler, the pins failed to align, resulting in an anastomotic failure. This occurred during patient use. The issue was ultimately resolved by gently dislodging the rings from the jaws using a hemostat, allowing the procedure to be completed successfully. There was no report of patient injury or medical intervention associated with this event. No additional information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50774545 ↗MalfunctionJul 3, 2026Device Dislodged or Dislocated
IT WAS REPORTED THAT A GEM COUPLER,3.0MM WAS DETACHED WHEN ATTEMPTING TO CONNECT A BLOOD VESSEL. THIS OCCURRED DURING BREAST RECONSTRUCTION SURGERY. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25706943 ↗MalfunctionApr 14, 2026Difficult to Open or Close
IT WAS REPORTED THAT A 3MM COUPLER RING MAY OPEN OR FAIL TO MAINTAIN ITS PROPER CLOSURE DURING HANDLING, PREVENTING CORRECT COUPLING OF THE SYSTEM. THIS OCCURRED DURING SURGICAL PROCEDURES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24875291 ↗MalfunctionApr 14, 2026Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM COUPLER RING MAY OPEN OR FAIL TO MAINTAIN ITS PROPER CLOSURE DURING HANDLING, PREVENTING CORRECT COUPLING OF THE SYSTEM. THIS OCCURRED DURING SURGICAL PROCEDURES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24875040 ↗MalfunctionApr 14, 2026Difficult to Open or Close
IT WAS REPORTED THAT TWO (2) 3.0MM COUPLERS RING MAY OPEN OR FAIL TO MAINTAIN ITS PROPER CLOSURE DURING HANDLING, PREVENTING CORRECT COUPLING OF THE SYSTEM. THIS OCCURRED DURING SURGICAL PROCEDURES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24875036 ↗MalfunctionMar 19, 2026
COUPLER WAS DEFECTIVE AND WOULD NOT FIT INTO COUPLER INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24638937 ↗MalfunctionFeb 27, 2026Break
IT WAS REPORTED THAT A 3.0MM COUPLER FELL APART AND WAS UNUSABLE. THIS WAS OBSERVED WHEN THE DEVICE WAS OPENED TO THE STERILE FIELD. A NEW DEVICE WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24470710 ↗MalfunctionJan 30, 2026Disconnection
DURING EVALUATION OF A RETURNED 3.0MM COUPLER, LOOSE RINGS WITHIN THE INNER TRAY WERE OBSERVED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24225356 ↗MalfunctionOct 20, 2025Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE PINS OF A MICROVASCULAR ANASTOMOTIC COUPLER DEVICE WERE NOT ENGAGING THE OPPOSING HOLES WHEN CLOSING. THIS ISSUE OCCURRED DURING USE OF THE DEVICE INTRA-OP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23342994 ↗MalfunctionSep 18, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE RINGS OF TWO (2) 3.0MM COUPLERS FELL OUT OF THE WING WITHOUT ANY SIGNIFICANT PRESSURE. THIS WAS OBSERVED WHILE INSERTING THE COUPLER INTO THE "RIGHT RING" WHILE HOLDING THE DEVICE BY HAND DURING AN INTRA-OPERATION STEP. ANOTHER ATTEMPT WAS MADE TO INSERT FROM THE TIP; HOWEVER, WAS LOOSE AND FELL OUT AGAIN. THE ISSUE WAS RESOLVED BY CHANGING THE SIZE TO A 2.5MM COUPLER. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23089691 ↗MalfunctionJun 2, 2025Disconnection; Disconnection
IT WAS REPORTED THAT THE RINGS OF AN UNSPECIFIED COUPLER DISLODGED FROM THE INSTRUMENT. THIS OCCURRED BEFORE COMPLETION OF ANASTOMOSIS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22127655 ↗MalfunctionMay 22, 2025Material Twisted/Bent; Incomplete or Inadequate Connection; Material Twisted/Bent; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE PINS OF A 3.0MM COUPLER WERE CROOKED AND COULD NOT BE FULLY ATTACHED. THIS WAS OBSERVED DURING PRE-OPERATION SET UP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22074941 ↗InjuryMay 2, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT AN INCOMPLETE ANASTOMOSIS OCCURRED ON A 3.0 MM COUPLER WHICH RESULTED IN "LEAKING" AFTER INSERTION. THIS OCCURRED DURING INTRA OPERATION PROCESS STEP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21953359 ↗MalfunctionApr 30, 2025Disconnection; Disconnection
IT WAS REPORTED THAT A RING OF A 3.0 MM COUPLER "FLEW OUT OF THE APPLICATOR" PRIOR TO ANASTOMOSIS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21938234 ↗MalfunctionMar 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE RING OF A 3.0MM COUPLER FELL OUT OF THE DEVICE. THIS OCCURRED DURING ANASTOMOSIS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21663010 ↗MalfunctionMar 4, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM COUPLER DID NOT CLAMP TIGHTLY ONTO THE VEIN. THIS WAS OBSERVED DURING A FREE FLAP GRAFT PROCEDURE. A NEW COUPLER WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21519002 ↗MalfunctionJan 27, 2025Fitting Problem; Fitting Problem
IT WAS REPORTED THAT THE "NEEDLE HOLES" OF A 3.0MM COUPLER WERE MISALIGNED RESULTING IN A FAILED ANASTOMOSIS. THIS WAS OBSERVED DURING PRE-OPERATION SET UP. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21247990 ↗MalfunctionNov 6, 2024Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT THE WING JAWS WOULD NOT CLOSE DURING PATIENT USE OF THREE (03) 3.0MM COUPLERS. THE ANASTOMOTIC INSTRUMENT WAS INSPECTED AND HAD NO OBSERVED MECHANICAL ISSUES. THE COUPLERS WERE REPLACED TO RESOLVE THE EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20626077 ↗MalfunctionOct 28, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT A 3.0MM MICROVASCULAR ANASTOMOTIC COUPLER DEVICE WAS FAULTY. THE COUPLER WAS NOT SEALING WELL IN THE APPLICATORY AND WAS LOOSE. THIS OCCURRED DURING USE OF THE DEVICE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20546175 ↗MalfunctionAug 12, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PLASTIC RING OF TWO (2) 3.0MM COUPLERS CAME LOOSE FROM THE DEVICE. THE RINGS WHERE THE VEIN ATTACHES CAME LOOSE DURING FREE FLAP PROCEDURE PRIOR TO THE SURGEON READINESS. A THIRD 3.0MM COUPLER WITH A DIFFERENT LOT NUMBER WAS THEN OPENED AND USED WITH NO ISSUE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19960638 ↗MalfunctionAug 7, 2024Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM COUPLER DID NOT CLOSE PROPERLY; FURTHER DESCRIBED AS "MISFIRE. COUPLER DID NOT MEET". THIS OCCURRED PRIOR TO PATIENT USE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934997 ↗MalfunctionApr 25, 2023Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM COUPLER WOULD NOT CLOSE. THE ISSUE WAS DISCOVERED DURING PATIENT USE. A NEW COUPLER WAS USED TO RESOLVE THE EVENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16808284 ↗MalfunctionMar 17, 2023Product Quality Problem; Difficult to Open or Close; Product Quality Problem; Difficult to Open or Close
IT WAS REPORTED THAT A 3.0MM COUPLER WOULD NOT CLOSE; EACH COUPLER (RING) SEEMED A DIFFERENT SIZE. THIS WAS OBSERVED DURING A SURGICAL PROCEDURE ON A PATIENT. A NEW COUPLER WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16565961 ↗MalfunctionJul 12, 2022Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT A 3.0MM COUPLER HAD A BROKEN COMPONENT; THE RING PIN WAS CROOKED SO IT WAS NOT ABLE TO BE FULLY ATTACHED. THIS WAS IDENTIFIED DURING A PATIENT PROCEDURE IN THE OPERATING ROOM. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14988281 ↗MalfunctionFeb 18, 2022Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE RINGS OF A 3.0MM COUPLER WERE ¿A BIT LOOSE¿ AND WERE UNABLE TO FULLY ATTACH. THIS OCCURRED DURING USE. ANOTHER COUPLER WAS USED TO COMPLETE ANASTOMOSIS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13563999 ↗
Other Device,anastomotic,microvascular devices
- GEM Microvascular Anastomotic COUPLER GEM2753Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular44 reports
- GEM Microvascular Anastomotic COUPLER GEM2755Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular13 reports
- GEM Microvascular Anastomotic COUPLER GEM2751Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular13 reports
- GEM Microvascular Anastomotic COUPLER GEM2752Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular10 reports
- GEM Microvascular Anastomotic COUPLER GEM2753 ISynovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular6 reports
- GEM FLOW COUPLER GEM2754 FCSynovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular6 reports
- GEM Microvascular Anastomotic COUPLER GEM2754 ISynovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular5 reports
- GEM Microvascular Anastomotic COUPLER GEM2750Synovis Micro Companies Alliance, Inc · Device,anastomotic,microvascular3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page