CERTAIN® BELLATEK® ENCODE® IEEHA455Recalls & reported problems
ZimVie US Corp LLC · CERTAIN® BELLATEK® ENCODE® EMERGENCE HEALING ABUTMENT 4.1MM(D) 5MM(P) 5MM(H) · Class 2 · Implant
All models of CERTAIN BELLATEK ENCODE (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00844868045496), received between Mar 20, 2023 and Sep 25, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Failure5
- Device Damaged by Another Device1
- Mechanical Problem1
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Positioning Failure
It was reported that during the second-stage surgery, the insertion of the Encode Emergence healing screw was impossible. The doctor was unable to screw the device. Tooth site 26.No impact on the patient reported. Procedure was completed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50896488 ↗MalfunctionApr 17, 2026Positioning Failure
IT WAS REPORTED THAT THE HEALING ABUTMENT WAS NOT SEATING CORRECTLY ON THE IMPLANT AT TOOTH SITE #28. THE PROCEDURE WAS COMPLETED WITH ANOTHER HEALING ABUTMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24916125 ↗MalfunctionJan 6, 2026Positioning Failure
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23974484 ↗MalfunctionNov 25, 2025Separation Failure; Separation Failure
NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23646875 ↗InjuryJan 30, 2025Mechanical Problem; Device Damaged by Another Device; Mechanical Problem; Device Damaged by Another Device
CONTACTED ORAL SURGEON FOR ADDITIONAL DETAILS. IT WAS REPORTED THAT THE TOP HEX OF THE ENCODE HEALING ABUTMENT WAS STRIPPED. WHILE TRYING TO DRILL IT THE ABUTMENT, THE INTERNAL THREADS OF THE IMPLANT WERE DAMAGED IN THE PROCESS. THE IMPLANT WAS REMOVED AND SITE GRAFTED. PATIENT WILL RETURN LATER TO REPLACE. SYMPTOMS AS A RESULT OF THE EVENT: PAIN.
Patient problems: Pain
MAUDE report 21277252 ↗MalfunctionMar 20, 2023Positioning Failure; Positioning Failure
IT WAS REPORTED THAT THE ENCODE HEALING ABUTMENT DID NOT SEAT IN THE PATIENT'S MOUTH. THEY WERE UNABLE TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16573007 ↗MalfunctionMar 20, 2023Positioning Failure; Positioning Failure
IT WAS REPORTED THAT THE ENCODE HEALING ABUTMENT DID NOT SEAT IN THE PATIENT'S MOUTH. THEY WERE UNABLE TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16572990 ↗
FDA clearance and classification
The FDA regulates product code NHA (abutment, implant, dental, endosseous) under 21 CFR 872.3630 as a Class II device. 21 CFR 872.3630 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: To be used in conjunction with an endosseous dental implant fixture to aid in prosthetic rehabilitation.
Common questions about the CERTAIN® BELLATEK® ENCODE® IEEHA455
Has the CERTAIN® BELLATEK® ENCODE® IEEHA455 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00844868045496), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CERTAIN® BELLATEK® ENCODE® IEEHA455?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 20, 2023 and Sep 25, 2026. In those reports, the most common specific device problems are “Positioning Failure” (5), “Device Damaged by Another Device” (1) and “Mechanical Problem” (1); the most common patient problem is “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CERTAIN® BELLATEK® ENCODE® IEEHA455?
None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CERTAIN® BELLATEK® ENCODE® IEEHA455 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the CERTAIN® BELLATEK® ENCODE® IEEHA455, and is it still sold?
The FDA's device database lists ZimVie US Corp LLC as the company responsible for it (brand CERTAIN® BELLATEK® ENCODE®, model IEEHA455). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Abutment, Implant, Dental, Endosseous devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS150 reports
- Nobel Biocare® 36662Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS2 recalls5 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous61 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- CERTAIN® BELLATEK® ILDAN3ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page