NA LONG-HDRecalls & reported problems
Synthes GmbH · 12.4 cm Heavy Duty Long Attachment. Use with XMax, eMax Systems. · Class 2
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00845384003595), received between May 9, 2018 and Jan 15, 2025 — classified as 22 malfunctions.
The most frequently reported specific device problem is “Excessive Heating” (14 reports).
35 other recalls cover devices cleared under the same FDA submission (K080802); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating14
- Vibration5
- Break2
- Detachment of Device or Device Component2
- Fail-Safe Did Not Operate1
- Noise, Audible1
- Overheating of Device1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- No Consequences Or Impact To Patient5
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2025Excessive Heating; Excessive Heating
IT WAS REPORTED THAT MAINTENANCE CHECK, IT WAS OBSERVED THAT THE MOTOR DEVICE OVERHEATED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21156837 ↗MalfunctionJan 13, 2025Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING A MAINTENANCE CHECK, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21138940 ↗MalfunctionFeb 7, 2023Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING EQUIPMENT INSPECTION, IT WAS NOTED THAT THE LONG ATTACHMENT DEVICE WAS RUNNING HOT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16311395 ↗MalfunctionOct 31, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING EQUIPMENT INSPECTION, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15697947 ↗MalfunctionAug 19, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING INSPECTION, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15260355 ↗MalfunctionAug 10, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING INSPECTION, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15197050 ↗MalfunctionJun 22, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING INSPECTION IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14773713 ↗MalfunctionJun 17, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING INSPECTION, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14734253 ↗MalfunctionJun 15, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING INSPECTION IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14701956 ↗MalfunctionFeb 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT BEARING OF THE ATTACHMENT DEVICE WAS OUT. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13616435 ↗MalfunctionFeb 3, 2022Break; Unintended System Motion; Fail-Safe Did Not Operate; Break; Unintended System Motion; Fail-Safe Did Not Operate
IT WAS REPORTED FROM (B)(4) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MOTOR DEVICE WOULD RUN IN THE LOCKED POSITION AND HAD UNINTENDED ACTIVATION/MOTION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR SAFETY ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT THE LOCK ON THE DEVICE WAS DAMAGED DURING A DEMONSTRATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13444900 ↗MalfunctionNov 15, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE ATTACHMENT DEVICE FELL APART. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12811654 ↗MalfunctionNov 3, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING MAINTENANCE AND INSPECTION IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12747339 ↗MalfunctionJul 12, 2021Break; Break
IT WAS REPORTED THAT DURING A MAINTENANCE CHECK, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE BEARING WAS WORN. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE COLOR BANDING HAD CHIPPED/FADED, HAD COMPONENT DAMAGE AND WAS CORRODED. THE DEVICE ALSO FAILED PRETESTS FOR VISUAL ASSESSMENT AND TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THE DATE OF EVENT WAS UNKNOWN. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12152365 ↗MalfunctionMar 29, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED FROM THE (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE WAS GENERATING HEAT, AND THE ETCHING TRIANGLE WITH EXCLAMATION MARK WAS MISSING (ILLEGIBLE ETCH AND MISSING LABEL). IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR LABELING ASSESSMENT AND TEMPERATURE ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11585678 ↗MalfunctionFeb 21, 2021Vibration; Noise, Audible; Vibration; Noise, Audible
IT WAS REPORTED THAT DURING PRE-SURGERY TESTING IT WAS OBSERVED THAT THE ATTACHMENT DEVICE HAD ¿SHAKING METHOD¿ AND RATTLE. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE HAD VIBRATION AND WAS UNABLE TO BE DISASSEMBLED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11356689 ↗MalfunctionJan 28, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT THE ATTACHMENT DEVICE GENERATED TOO MUCH HEAT DURING TESTING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11244683 ↗MalfunctionOct 20, 2020Vibration; Excessive Heating; Vibration; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD HEAT, WORN BEARING AND VIBRATION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VIBRATION AND TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10707654 ↗MalfunctionAug 31, 2020Vibration; Excessive Heating; Vibration; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD A WORN BEARING, MISSING LABEL, VIBRATION, HEAT AND COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL, VIBRATION AND TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10471934 ↗MalfunctionJul 8, 2019Vibration; Vibration
IT WAS REPORTED THAT DURING TESTING IT WAS OBSERVED THAT THE BEARING OF THE LONG ATTACHMENT DEVICE WAS BROKEN. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS OBSERVED THAT THE DEVICE WAS VIBRATING TOO MUCH. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8767220 ↗MalfunctionApr 12, 2019Vibration; Vibration
IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE FAILED VIBRATION ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2019. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8510389 ↗MalfunctionMay 9, 2018Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING INSPECTION, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS RUNNING HOT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7498151 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080802. One submission often covers many models, so check the product and lot details in each record.
- Z-2026-2017 · Oct 6, 2016 · TerminatedReciprocating MICRO SAW
- Z-1955-2017 · Oct 6, 2016 · TerminatedConsole, eMax2 Plus Product Usage: Electric system
- Z-1957-2017 · Oct 6, 2016 · TerminatedSystem Console, Single Port, no irrigation
- Z-2032-2017 · Oct 6, 2016 · Terminated20.2cm Heavy Duty Long Heavy Short Attachment
- Z-2015-2017 · Oct 6, 2016 · TerminatedSagittal Saw Product Usage: Pneumatic system
- Z-2001-2017 · Oct 6, 2016 · Terminated12.4cm Heavy Duty Long Heavy Short Attachment Product Usage: Pneumatic system
- Z-2019-2017 · Oct 6, 2016 · TerminatedOscillating MICRO-SAW Product Usage: Pneumatic system
- Z-2031-2017 · Oct 6, 2016 · TerminatedSagittal MICRO-SAW
- Z-1949-2017 · Oct 6, 2016 · TerminatedElectronic Foot Control with Direction or Irrigation Switches Product Usage: Electric system
- Z-1952-2017 · Oct 6, 2016 · TerminatedeMax2 PLUS Product Usage: Electric system
- Z-1948-2017 · Oct 6, 2016 · TerminatedElectronic Foot Control with Direction Switch Product Usage: Electric system
- Z-1956-2017 · Oct 6, 2016 · TerminatedSystem Console, Single Port, with irrigation Product Usage: Electric system
- Z-0728-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs & Drums: 1.5MM x 7.5MM Cylindrical Carbide Drum part # S-15DRMC Anspach cutting burrs are intended for cutting and shaping bone, including the spine
- Z-0726-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs: 2MM Carbide Fluted Ball, Standard, 8 Flutes - part #QD8-2SB-C-8F, 4MM Carbide Fluted Ball, 10 Flutes - part # QD8-4B-C-10F 5MM Carbide Fluted Ball, 1
- Z-0727-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs: 1.5MMx7.5MM Cylindrical Carbide Drum part # S-15DRMC-L, 1MM Carbide Fluted ball, Extended part # S-1B-C, 1MM Carbide Fluted ball part # S-SB-C, 2MM C
…and 20 more.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) devices
- CODMAN® 261221Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall223 reports
- Zyphr 5100060001Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall199 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)67 reports
- CODMAN® 261222Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall7 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)54 reports
- CODMAN® 261230Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall3 reports
- CODMAN® 261223Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall2 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)37 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.