CODMAN® 261222Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® Disposable Perforator 11mm · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10381780513605), received between May 17, 2022 and Jun 18, 2025 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Fail-Safe Did Not Operate” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 26, 2025Open, Classified

    Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.

    Z-0374-2026 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Fail-Safe Did Not Operate3
  • Dull, Blunt2
  • Melted1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Ambulation or Postural Difficulties1
  • Brain Injury1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    A PHYSICIAN REPORTED A DISPOSABLE PERFORATOR (ID 261222) FAILED TO DISENGAGE CAUSING DURAL DAMAGE, REQUIRING DURAL RECONSTRUCTION TO TREAT THE DAMAGE. A SURGICAL DELAY OF LESS THAN 30 MINUTES WAS REPORTED. THE PERFORATOR WAS USED DURING A BRAIN TUMOR REMOVAL SURGERY. THE DRILL USED WITH THE PERFORATOR WAS AN ANSPACH. ACCORDING TO INFORMATION PROVIDED, IT IS UNKNOWN IF THE DRILL WAS ELECTRIC OR PNEUMATIC, OR IF THE PERFORATOR CLICKED IN PLACE ON THE DRILL. IT IS ALSO UNKNOWN IF THE RECOMMENDED SPRING TESTS WERE PERFORMED BETWEEN EACH BURR HOLE. ADDITIONALLY, IT IS UNKNOWN IF A PERPENDICULAR APPROACH WAS MAINTAINED THROUGHOUT THE DRILLING PROCESS, WHETHER ANY ROCKING MOTION OCCURRED, OR IF CONSTANT DOWNWARD PRESSURE WAS APPLIED.

    Patient problems: Brain Injury

    MAUDE report 22283716 ↗
  • MalfunctionSep 14, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    A FACILITY REPORTED: "DURING THE SURGICAL PROCEDURE, WHEN PERFORATOR (ID 261222) IS ASSEMBLED IN THE COUPLING, IT IS EVIDENT THAT IT IS UNADJUSTED (LOOSE). THE SURGEON PERFORMS THE ADJUSTMENT MANUALLY. WHEN PERFORMING THE BURR PERFORATION, THE DRILL DOES NOT PERFORM THE BLOCKING WHEN PASSING THE BONE TABLE AND CONTINUES UNTIL THE DURA OF THE PATIENT." NO PATIENT INJURY REPORTED AND THE EVENT LED TO LESS THAN 30 MINUTES SURGICAL DELAY. THE DRILL WAS AN ELECTRIC PIDRIVE ENGINE (STRYKER). THE DRILL CLICKED INTO PLACE ON THE DRILL AND IT WAS HEARD WHEN ASSEMBLING IT WITH THE COUPLING AND THE BIT. RECOMMENDED RESOURCE TESTS BEING PERFORMED BETWEEN EACH HOLE AND SEVERAL TESTS WERE CARRIED OUT AT THE TIME OF FINISHING ASSEMBLING THE COUPLER WITH THE MOTOR AND CONFIGURING THE CONSOLE. THE SURGEON PERFORMED THE CRANIOTOMY AT A 90 DEGREE ANGLE. AT ALL TIMES, THEY TRIED TO MAINTAIN THE ANGLE, HOWEVER THE DRILL BECAME DISLODGED FROM THE COUPLING AT THE BEGINNING OF THE CRANIOTOMY PROCESS AND HAD TO BE REARRANGED. THEY CONFIRMED THAT CONSTANT PRESSURE WAS MAINTAINED AT ALL TIMES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17745379 ↗
  • InjuryAug 15, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    N/A.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 17549887 ↗
  • MalfunctionMar 29, 2023Peeled/Delaminated; Peeled/Delaminated

    A FACILITY REPORTED THE LABEL OF A PERFORATOR (ID 261222) CAME OFF DURING USE AND THUS FALL IN THE WOUND. NO PATIENT INJURY. THE PROCEDURE WAS DELAYED CAUSE THEY HAD TO SEARCH FOR THE PIECES OF PAPER IN THE BODY OF PATIENT BUT THERE IS NO INFORMATION ABOUT EXACT TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16635722 ↗
  • MalfunctionNov 10, 2022Dull, Blunt; Dull, Blunt

    THIS IS 2 OF 2 REPORTS LINKED TO MFG REPORT NUMBER: 3014334038-2022-00254. A PHYSICIAN REPORTED A PERFORATOR (ID (B)(4)) COULDN'T DRILL THE BONE. THE PHYSICIAN FELT THE PERFORATOR WAS TOO DULL AND HAD DIFFICULTY IN PERFORATING DURING A PROCEDURE. ACCORDING TO INFORMATION RECEIVED, IT IS UNKNOWN IF WHAT DRILL WAS USED AND IF THE PERFORATOR CLICKED IN PLACE ON THE DRILL. IT IS ALSO UNKNOWN IF THE RECOMMENDED SPRING TESTS WERE PERFORMED BETWEEN EACH BURR HOLE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15772752 ↗
  • MalfunctionNov 9, 2022Dull, Blunt; Dull, Blunt

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15765206 ↗
  • MalfunctionMay 17, 2022Melted; Melted

    A FACILITY REPORTED THE TOP OF THE GREEN PLASTIC OF A PERFORATOR MELTED AND WAS PRODUCING SMOKE WHILE IN USE. THE PERFORATOR WAS INSPECTED PRIOR TO USE AS INSTRUCTED; THE SPRING TESTS WERE PERFORMED; THE PLASTIC SLEEVE WAS CHECKED PRIOR TO USE. NO PATIENT INJURY REPORTED AND THE EVENT DID NOT LED TO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14414923 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.