Skip to content

CODMAN® 261230Recalls & reported problems

Integra Lifesciences Production Corporation · CODMAN® Craniotomy Kit Disposable Perforator Cranio-blade Wire Pass Drill 14mm · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10381780513629), received between Aug 2, 2022 and Jun 9, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Fail-Safe Did Not Operate” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Apr 11, 2025Open, Classified

    Inadequate weld that can potentially cause the product to disassemble.

    Z-1717-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Fail-Safe Did Not Operate3

Patient problems

  • Brain Injury3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 9, 2023Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    A FACILITY REPORTED A PERFORATOR (ID: 261230) FAILED TO DISENGAGE CAUSING A DURAL INJURY REQUIRING TO BE REPAIRED WITH COAGULATION OF THE VESSEL ON THE SURFACE. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT PRODUCT AVAILABLE.

    Patient problems: Brain Injury

    MAUDE report 17100888 ↗
  • InjuryDec 21, 2022Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    N/A.

    Patient problems: Brain Injury

    MAUDE report 16022588 ↗
  • InjuryAug 2, 2022Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    A FACILITY REPORTED A PERFORATOR DID NOT DISENGAGE CAUSING A TEAR IN THE DURA MATTER. SUTURE WAS NECESSARY TO STOP THE BLEEDING OF THE SUPERIOR SAGITTAL SINUS. THE PERFORATOR WAS USED IN A SURGERY FOR MENINGIOMA.

    Patient problems: Brain Injury

    MAUDE report 15148291 ↗

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) devices