Skip to content

NA MA15-4DRecalls & reported problems

Synthes GmbH · 4 mm Diamond Ball, 15 cm. Use with MA-15 Bearing Sleeves. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00845384004578), received between Dec 10, 2019 and Nov 20, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

31 other recalls cover devices cleared under the same FDA submission (K042783); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3
  • Break1
  • Excessive Heating1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A TRANSSPHENOIDAL PROCEDURE, IT WAS DISCOVERED THAT THE CUTTER DEVICE FRACTURED DURING USE. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20737744 ↗
  • MalfunctionMar 7, 2021Break; Break

    IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE BURR DEVICE WAS JAMMED INSIDE THE BEARING SLEEVE AND ATTACHMENT DEVICE. DURING THE ASSESSMENT IT WAS NOTED THAT IT WAS POSSIBLE TO REMOVE THE CUTTER AND BEARING SLEEVE FROM THE ATTACHMENT LOCKING MECHANISM. IT WAS FURTHER DETERMINED THAT THE CUTTER SHAFT WAS BENT CLOSE TO THE DIAMOND BALL AND AT ITS LOCKING GROOVE. IT WAS NOT POSSIBLE TO CONNECT THE DEVICE TO THE SERVICE CENTER'S EQUIPMENT DUE TO THE DENT AT THE END OF THE SHAFT. IT WAS FURTHER OBSERVED THAT THE DIAMOND BALL CARRIED RESIDUES, AND ALONG THE CUTTER SHAFT, THERE WERE SIGNS OF USAGE AND CORROSION DETECTED, WHICH MOST LIKELY RESULTED FROM REPROCESSING OF ALL THE THREE PARTS INTERLOCKED. IT WAS DETERMINED THAT THE CUTTER COULD NOT BE INSERTED AND WAS DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION WHILE BEING USED TOGETHER WITH THE ATTACHMENT AND BEARING SLEEVE DEVICE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11426518 ↗
  • MalfunctionDec 3, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED FROM (B)(6) THAT DURING A RESECTION OF A COMBINED LEVEL 5 BRAIN LESION OR SKULL BASE SURGICAL PROCEDURE, THE CUTTER DEVICE WAS SPLIT NEAR THE JUNCTION WITH SPHERICAL PART WHILE STARTING THE DRILLING AT THE ENDONASAL APPROACH. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR WHETHER A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10942268 ↗
  • MalfunctionDec 10, 2019Material Fragmentation; Excessive Heating; Material Fragmentation; Excessive Heating

    THIS IS REPORT 1 OF 3 FOR THE SAME EVENT. IT WAS REPORTED FROM (B)(6) THAT DURING AN ENDOSCOPIC LUMBAR LAMINECTOMY MEPD MILD SURGICAL PROCEDURE FOR A LUMBAR SPINAL CANAL STENOSIS (ARTHROSCOPIC ABLATION FOR ARCH OF LUMBAR VERTEBRA) IT WAS OBSERVED THAT AFTER 3 TO 5 MINUTES OF CUTTING OUT BONE WITH A MINIMAL ACCESS DRIVER DEVICE, BEARING SLEEVE DEVICE, AND CUTTER DEVICE, THE HANDLE OF THE DEVICE HAD BECOME QUITE HEATED, AND THE BAR OF THE CUTTING BURR DEVICE IMMEDIATELY BROKE OFF. IT WAS REPORTED THAT THE DEVICE DROVE AT 80,000 ROTATIONS PER MINUTE (RPM) DURING THE ABLATION. IT WAS REPORTED THAT THERE WAS A FIVE-MINUTE DELAY IN THE PROCEDURE DUE TO THE EVENT. IT WAS NOT REPORTED IF THERE WAS A SPARE DEVICE AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9443923 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K042783. One submission often covers many models, so check the product and lot details in each record.

  • Z-2009-2017 · Oct 6, 2016 · Terminated
    Minimal Access Attachment Product Usage: Pneumatic system
  • Z-2011-2017 · Oct 6, 2016 · Terminated
    Minimal Access Straight Driver Product Usage: Pneumatic system
  • Z-2006-2017 · Oct 6, 2016 · Terminated
    Bearing Sleeve, 15CM Straight Product Usage: Pneumatic system
  • Z-2004-2017 · Oct 6, 2016 · Terminated
    Bearing Sleeve, 10CM Straight Product Usage: Pneumatic system
  • Z-2003-2017 · Oct 6, 2016 · Terminated
    Bearing Sleeve, 10CM Curved Product Usage: Pneumatic system
  • Z-2007-2017 · Oct 6, 2016 · Terminated
    Bearing Sleeve, 15CM Straight Taper Product Usage: Pneumatic system
  • Z-2005-2017 · Oct 6, 2016 · Terminated
    Bearing Sleeve, 15CM Curved Product Usage: Pneumatic system
  • Z-0708-2014 · Oct 26, 2012 · Terminated
    Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, M
  • Z-0756-2014 · Mar 29, 2012 · Terminated
    MCA5-1SB - 1MM Fluted Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochl
  • Z-0768-2014 · Mar 29, 2012 · Terminated
    MCA7-2SB - 2MM Fluted Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochl
  • Z-0761-2014 · Mar 29, 2012 · Terminated
    MCA7-06SD - 0.6MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as c
  • Z-0755-2014 · Mar 29, 2012 · Terminated
    MCA5-15SD - 1.5MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as c
  • Z-0753-2014 · Mar 29, 2012 · Terminated
    MCA5-08SD - 0.8MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as c
  • Z-0751-2014 · Mar 29, 2012 · Terminated
    MCA5-06SD - 0.6MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as c
  • Z-0762-2014 · Mar 29, 2012 · Terminated
    MCA7-07SD - 0.7MM Diamond Ball, 7.3CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as c

…and 16 more.

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates