NA MEDIUM-HDRecalls & reported problems
Synthes GmbH · 8 cm Medium Heavy Duty Attachment. Use with XMax, eMax Systems. · Class 2
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00845384004905), received between Feb 25, 2019 and Oct 22, 2024 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Excessive Heating” (11 reports).
70 other recalls cover devices cleared under the same FDA submission (K011444); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating11
- Vibration5
- Device Reprocessing Problem3
- Appropriate Device Problem Term/Code Not Available2
- Detachment of Device or Device Component2
- Break1
- Device Slipped1
- Device-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2024Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating
IT WAS REPORTED BY GERMANY THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE GENERATED HEAT. IT WAS FURTHER OBSERVED THAT THE DEVICE EXHIBITED VIBRATION AND THE COLOR BAND WAS CHIPPING/FADING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR COLOR BAND VISUAL ASSESSMENT, VIBRATION, AND TEMPERATURE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20507896 ↗MalfunctionNov 2, 2023Excessive Heating; Excessive Heating
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE PRODUCED HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VERIFICATION: TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18053961 ↗MalfunctionSep 19, 2023Appropriate Device Problem Term/Code Not Available; Excessive Heating; Appropriate Device Problem Term/Code Not Available; Excessive Heating
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE PRODUCED HEAT. IT WAS FURTHER OBSERVED THAT THE DEVICE HAD VIBRATION, A WORN BEARING, AND THE COLOR BAND WAS CHIPPING/FADING. IT WAS DETERMINED THAT THE CUTTER COULD NOT BE INSERTED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR COLOR BAND VISUAL ASSESSMENT, CUTTER INSERTION, VIBRATION AND TEMPERATURE. IT WAS NOTED IN THE SERVICE ORDER THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE DEVICE WAS NOT WORKING PROPERLY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17773648 ↗MalfunctionJun 2, 2023Excessive Heating; Excessive Heating
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE PRODUCED HEAT, EXHIBITED VIBRATION, HAD ILLEGIBLE LABELING ETCH AND COULD NOT INSERT CUTTER. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR LABELING ASSESSMENT, CUTTER INSERTION, VIBRATION AND TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17048800 ↗MalfunctionNov 30, 2021Detachment of Device or Device Component; Device-Device Incompatibility; Detachment of Device or Device Component; Device-Device Incompatibility
IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE BEARINGS ON THE ATTACHMENT DEVICE WERE OUT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12894442 ↗MalfunctionJul 8, 2021Vibration; Vibration
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD A WORN BEARING, VIBRATION AND ILLEGIBLE ETCH. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR LABELING ASSESSMENT AND VIBRATION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12134317 ↗MalfunctionJul 6, 2021Vibration; Vibration
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD A WORN BEARING, EXHIBITED VIBRATION, MISSING LABELLING AND AN ILLEGIBLE LABELING ETCH. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR LABELING ASSESSMENT AND VIBRATION. IT WAS NOTED THAT THE TRIANGLE SYMBOL WAS MISSING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12122977 ↗MalfunctionJul 6, 2021Vibration; Vibration
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS VIBRATING, THE ETCHING WAS ILLEGIBLE, THE TRIANGLE SYMBOL WAS MISSING AND THE BEARING WAS WORN. THE DEVICE ALSO FAILED PRETESTS FOR VIBRATION AND LABELING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12121234 ↗MalfunctionJun 3, 2021Device Reprocessing Problem; Excessive Heating; Device Reprocessing Problem; Excessive Heating
IT WAS REPORTED THAT THE ATTACHMENT DEVICE HAD WATER IN IT. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE WAS GENERATING HEAT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11934669 ↗MalfunctionMay 27, 2021Device Reprocessing Problem; Excessive Heating; Device Reprocessing Problem; Excessive Heating
IT WAS REPORTED THAT THE ATTACHMENT DEVICE HAD WATER IN IT. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE COLOR BAND WAS CHIPPING/FADING, AND THE DEVICE GENERATED HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND TEMPERATURE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11893652 ↗MalfunctionMay 14, 2021Device Reprocessing Problem; Excessive Heating; Device Reprocessing Problem; Excessive Heating
IT WAS REPORTED THAT THE ATTACHMENT DEVICE HAD WATER IN IT. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD HEAT, THE COLOR BAND WAS CHIPPING AND FADING AND FAILED PRE-TEST FOR TEMPERATURE VERIFICATION. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11825133 ↗MalfunctionApr 8, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRETESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11634286 ↗MalfunctionJan 27, 2021Break; Device Slipped; Excessive Heating; Break; Device Slipped; Excessive Heating
IT WAS REPORTED THAT THE ATTACHMENT DEVICE HAD LOOSE BEARINGS. DURING AN IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD HEAT, OLD SPACERS, THE COLOR BAND WAS CHIPPING/FADING AND FAILED PRE-TEST FOR TEMPERATURE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11239966 ↗MalfunctionNov 17, 2020Vibration; Vibration
IT WAS REPORTED FROM SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD MISSING LABELING, THE BEARING WAS WORN, AND THE DEVICE HAD VIBRATION AND COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR LABELING ASSESSMENT, NOSE CONE VERIFICATION, AND VIBRATION VERIFICATION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10856162 ↗MalfunctionOct 20, 2020Vibration; Excessive Heating; Vibration; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE HAD HEAT, WORN BEARING AND VIBRATION. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VIBRATION AND TEMPERATURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10707719 ↗MalfunctionAug 31, 2020Excessive Heating; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS OVERHEATING, THE WARNING TRIANGLE LABELING WAS MISSING AND THERE WAS COMPONENT DAMAGE. THE DEVICE ALSO FAILED PRETESTS FOR VISUAL AND TEMPERATURE ASSESSMENT. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10472423 ↗MalfunctionFeb 25, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING PRE-SURGERY IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS BROKEN. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS OBSERVED THAT THERE WERE TEETH MARKS IN THE NOSE CONE (FAILED PRETEST FOR VISUAL ASSESSMENT) AND THE LOCTITE HAD BROKEN APART AT THE SECONDARY LOCK PROTECTION. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8365817 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011444. One submission often covers many models, so check the product and lot details in each record.
- Z-0918-2023 · Dec 7, 2022 · Open, ClassifiedAdult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-0920-2023 · Dec 7, 2022 · Open, Classified7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
- Z-0919-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
- Z-0922-2023 · Dec 7, 2022 · Open, Classified7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
- Z-0924-2023 · Dec 7, 2022 · Open, ClassifiedPediatric Craniotome Ref:CRANI-P-G1
- Z-0923-2023 · Dec 7, 2022 · Open, Classified6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
- Z-0921-2023 · Dec 7, 2022 · Open, ClassifiedAdult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-0917-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
- Z-0916-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
- Z-2025-2017 · Oct 6, 2016 · Terminated7.5cm QD Angle Attachment
- Z-2000-2017 · Oct 6, 2016 · Terminated10.5cm Long Attachment, Non-Tapered End Product Usage: Pneumatic system
- Z-1987-2017 · Oct 6, 2016 · Terminated6.5CM ADLT CRANI, THIN FT PLATE Product Usage: Pneumatic system
- Z-1930-2017 · Oct 6, 2016 · Terminated14cm QD Angle Attachment Product Usage: Product attachment
- Z-2027-2017 · Oct 6, 2016 · TerminatedSMALL ATTACHMENT, JACOBS CHUCK
- Z-2029-2017 · Oct 6, 2016 · Terminated5.0cm Short Attachment
…and 55 more.
Other MOTOR, DRILL, ELECTRIC devices
- NA CRANI-A-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC1 recall358 reports
- NA QD8Synthes GmbH · MOTOR, DRILL, ELECTRIC386 reports
- NA QD11-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC312 reports
- Midas Rex EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC254 reports
- MIDAS REX MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC186 reports
- Midas Rex MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC158 reports
- NA SHORT-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC134 reports
- Midas Rex® EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC123 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.