Skip to content

NA QD11-5BRecalls & reported problems

Anspach Effort, The · 5 mm Fluted Ball. Use with QD11 Attachment. · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384006138), received between Nov 19, 2021 and Nov 19, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Dull, Blunt” (1 report).

70 other recalls cover devices cleared under the same FDA submission (K011444); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Dull, Blunt1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2021Dull, Blunt; Separation Problem; Dull, Blunt; Separation Problem

    IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE SHAFT SHOWED HEAVY CORROSION AND THE HEAD OF THE DEVICE SHOWED SIGNS OF REGULAR WEAR. IT WAS NOTED THAT THE CUTTER DEVICE WAS RECEIVED INSIDE A DEDICATED ATTACHMENT DEVICE AND WAS REMOVED FOR A SEPARATE ASSESSMENT. IT WAS NOTED THAT THE CUTTER DEVICE BROKE INTO TWO PIECES WHEN IT WAS REMOVED FROM THE ATTACHMENT DEVICE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT LEVER LOCK OF THE DEDICATED ATTACHMENT DEVICE WAS DEFECTIVE AND WHEN RECEIVED FOR EVALUATION IT WAS OBSERVED THAT THE CUTTING TOOL WAS RECEIVED INSIDE THE ATTACHMENT DEVICE AND COULD NOT BE REMOVED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12845139 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K011444. One submission often covers many models, so check the product and lot details in each record.

  • Z-0922-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
  • Z-0917-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
  • Z-0924-2023 · Dec 7, 2022 · Open, Classified
    Pediatric Craniotome Ref:CRANI-P-G1
  • Z-0916-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
  • Z-0920-2023 · Dec 7, 2022 · Open, Classified
    7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
  • Z-0923-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
  • Z-0921-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-0919-2023 · Dec 7, 2022 · Open, Classified
    6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
  • Z-0918-2023 · Dec 7, 2022 · Open, Classified
    Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
  • Z-2012-2017 · Oct 6, 2016 · Terminated
    8.0cm Medium Attachment Product Usage: Pneumatic system
  • Z-2027-2017 · Oct 6, 2016 · Terminated
    SMALL ATTACHMENT, JACOBS CHUCK
  • Z-1931-2017 · Oct 6, 2016 · Terminated
    AL-III W/ male disconnect and 20ft hose Product Usage: Pneumatic system
  • Z-1990-2017 · Oct 6, 2016 · Terminated
    Rotating Large Craniotome
  • Z-1985-2017 · Oct 6, 2016 · Terminated
    6.8cm Controlled Depth Attachment Product Usage: Pneumatic system
  • Z-1953-2017 · Oct 6, 2016 · Terminated
    Foot Control, SC2100 Product Usage: Electric system

…and 55 more.

Other MOTOR, DRILL, ELECTRIC devices

All MOTOR, DRILL, ELECTRIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page