NA QD8-5BRecalls & reported problems
Anspach Effort, The · 5 mm Fluted Ball. Use with QD8 Attachment. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384006954), received between Jul 15, 2021 and Jul 15, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
70 other recalls cover devices cleared under the same FDA submission (K011444); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2021Break; Separation Problem; Break; Separation Problem
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE WAS RECEIVED STUCK INSIDE THE ATTACHMENT DEVICE BUT COULD BE EASILY RELEASED DURING THE ASSESSMENT. IT WAS FOUND THAT THE DEVICE FAILED THE VISUAL INSPECTION. IT WAS FURTHER OBSERVED THAT THE DEVICE SHOWED REGULAR SIGNS OF NORMAL WEAR AND THE CUTTING EDGES WERE DAMAGED. IT WAS NOTED IN THE SERVICE ORDER THAT THE CUTTER DEVICE WAS STUCK IN THE ATTACHMENT. IT WAS REPORTED THAT THERE WERE NO DELAYS IN SURGERY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12175629 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011444. One submission often covers many models, so check the product and lot details in each record.
- Z-0916-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems. Ref: CRANI-A Intended for cutting and shaping bone including the spine and cranium
- Z-0920-2023 · Dec 7, 2022 · Open, Classified7.5 cm Large Craniotome Attachment Use with XMax, microMax, and eMax Systems. Ref: CRANI-L Intended for cutting and shaping bone including the spine and cranium
- Z-0922-2023 · Dec 7, 2022 · Open, Classified7.5 cm Large Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-L-R Intended for cutting and shaping bone including the spine and cranium
- Z-0917-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Craniotome, Thin Foot Plate Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-01 intended for cutting and shaping bone including the spine and crani
- Z-0923-2023 · Dec 7, 2022 · Open, Classified6.5 cm Pediatric Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-P Intended for cutting and shaping bone including the spine and cranium
- Z-0919-2023 · Dec 7, 2022 · Open, Classified6.5 cm Adult Rotating Craniotome Attachment. Use with XMax, microMax, and eMax Systems Ref: CRANI-A-R intended for cutting and shaping bone including the spine and cranium
- Z-0924-2023 · Dec 7, 2022 · Open, ClassifiedPediatric Craniotome Ref:CRANI-P-G1
- Z-0918-2023 · Dec 7, 2022 · Open, ClassifiedAdult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-0921-2023 · Dec 7, 2022 · Open, ClassifiedAdult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
- Z-2023-2017 · Oct 6, 2016 · Terminated14cm QD Angle Attachment
- Z-1988-2017 · Oct 6, 2016 · TerminatedRotating Adult Craniotome Product Usage: Pneumatic system
- Z-2013-2017 · Oct 6, 2016 · Terminated8.0CM Heavy Duty Medium Attachment Product Usage: Pneumatic system
- Z-1933-2017 · Oct 6, 2016 · TerminatedFoot control irrigation system Product Usage: Pneumatic system
- Z-1932-2017 · Oct 6, 2016 · TerminatedAL-III W/ male disconnect and 30ft hose Product Usage: Pneumatic system
- Z-1951-2017 · Oct 6, 2016 · TerminatedeMax2 Hybrid Hand Control Product Usage: Electric system
…and 55 more.
Other MOTOR, DRILL, ELECTRIC devices
- NA CRANI-A-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC1 recall358 reports
- NA QD8Synthes GmbH · MOTOR, DRILL, ELECTRIC386 reports
- NA QD11-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC312 reports
- Midas Rex EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC254 reports
- MIDAS REX MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC186 reports
- Midas Rex MR8 EM800Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC158 reports
- NA SHORT-G1Synthes GmbH · MOTOR, DRILL, ELECTRIC134 reports
- Midas Rex® EM200Medtronic Powered Surgical Solutions · MOTOR, DRILL, ELECTRIC123 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page