NA R-SAWRecalls & reported problems
Anspach Effort, The · Reciprocating microSaw Attachment. Use with all Anspach Systems. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00845384007173), received between Dec 1, 2017 and Mar 21, 2024 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Excessive Heating” (10 reports).
35 other recalls cover devices cleared under the same FDA submission (K080802); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating10
- Complete Loss of Power7
- Detachment of Device or Device Component2
- Overheating of Device2
- Vibration1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2024Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED BY GERMANY THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING MICROSAW ATTACHMENT DEVICE PRODUCED HEAT. IT WAS FURTHER OBSERVED THAT THE DRIVE SHAFT WAS BENT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THAT THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN JANUARY 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18951427 ↗MalfunctionJan 25, 2024Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING SAW ATTACHMENT DEVICE WAS HEATING UP DURING TESTING. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED VISUAL INSPECTION DUE TO DAMAGED CONNECTOR/COUPLER, THE CUTTER DEVICE COULD NOT BE INSERTED, CLAMPING JAWS ON THE DRUM WERE BROKEN OFF AND CORROSION WAS DETECTED ON THE BREAKING POINT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, BLADE INSERTION CHECK AND TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2023. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18579518 ↗MalfunctionAug 3, 2023Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED THAT ATTACHMENT DEVICE DID NOT WORK. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS OBSERVED THAT THE DEVICE HAD HEAT. IT WAS FURTHER DETERMINED THAT THE DEVICE HAD MISSING COMPONENTS AND COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, BLADE INSERTION CHECK, AND TEMPERATURE ASSESSMENT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17451204 ↗MalfunctionNov 22, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED FROM ISRAEL THAT THE HAND PIECE CABLE WAS TURN BETWEEN THE ATTACHMENT AND THE CONSOLE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15842016 ↗MalfunctionNov 4, 2022Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED FROM SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING MICRO SAW DEVICE HAD HEAT, BEARING DAMAGE, THE LOCKING MECHANISM DID NOT MOVE SMOOTHLY, LOCKING COMPONENTS WERE DAMAGED, AND THE LOCKING MECHANISM WAS STIFF/STUCK, WORN TOOL COUPLING, COSMETIC DAMAGE, AND WAS MAKING EXCESSIVE NOISE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, BLADE INSERTION CHECK, AND TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT PRE-PROCEDURE IT WAS OBSERVED THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15728974 ↗MalfunctionAug 4, 2022Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED FROM SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE MOTOR ON THE DEVICE WOULD NOT SECURE THE CUTTER DEVICE, HEATED UP AND WAS WORN. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETESTS FOR BLADE INSERTION CHECK AND TEMPERATURE ASSESSMENT IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK PRIOR TO SURGERY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15164753 ↗MalfunctionJun 7, 2022Complete Loss of Power; Excessive Heating; Complete Loss of Power; Excessive Heating
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING MICRO SAW DEVICE PRODUCED HEAT, COULD NOT SECURE/LOCK CUTTER AND HAD COMPONENT DAMAGE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT, BLADE INSERTION CHECK AND TEMPERATURE ASSESSMENT. IT WAS NOTED THAT THE DEVICE AND LOCKING MECHANISM DAMAGE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14615625 ↗MalfunctionApr 7, 2022Excessive Heating; Excessive Heating
IT WAS REPORTED THAT THE RECIPROCATING MICRO SAW ATTACHMENT DEVICE WAS OVERHEATING. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A DELAY TO A PLANNED PROCEDURE. IT WAS NOT REPORTED IF THERE WAS A SPARE DEVICE AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14047320 ↗MalfunctionSep 9, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT THE SAW DEVICE GENERATED HEAT. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12445142 ↗MalfunctionAug 16, 2021Excessive Heating; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND REPAIR/PRE-TESTING, IT WAS OBSERVED THAT THE RECIPROCATING MICRO SAW DEVICE WAS OVERHEATING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12328085 ↗MalfunctionMar 29, 2021Overheating of Device; Complete Loss of Power; Overheating of Device; Complete Loss of Power
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE RECIPROCATING MICRO SAW DEVICE HAD A COMPONENT DAMAGE, AND IT GENERATED HEAT. IT WAS FURTHER OBSERVED THAT THE JAW WAS LOOSE FROM THE CHUCK. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT DURING PRE-SURGERY, IT WAS DISCOVERED THAT THE DEVICE DID NOT WORK. IT WAS REPORTED THAT THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2021. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11586821 ↗MalfunctionJul 30, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE LOCKING PINS OF THE RECIPROCATING MICRO SAW ATTACHMENT DEVICE CAME OUT AND WOULD NOT LOCK THE CUTTER DEVICE PROPERLY. IT WAS REPORTED THAT THERE WAS A TWENTY MINUTE DELAY TO THE SURGICAL PROCEDURE. IT WAS UNKNOWN IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10343332 ↗MalfunctionMay 1, 2020Excessive Heating; Excessive Heating
IT WAS REPORTED FROM (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SAW DEVICE LOCKING MECHANISM WAS STIFF/STUCK AND THE TOOL COULD NOT BE CLAMPED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR TEMPERATURE ASSESSMENT. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10017230 ↗MalfunctionDec 1, 2017Overheating of Device; Vibration; Overheating of Device; Vibration
IT WAS REPORTED THAT DURING A TEACHING PROCEDURE IN A CADAVER LAB, IT WAS OBSERVED THAT THE RECIPROCATING SAW DEVICE WAS MAKING A BROKEN SOUND AND HAD EXCESSIVE VIBRATION. IT WAS FURTHER REPORTED THAT THE DEVICE BECAME EXTREMELY HOT WITH SMOKE AND METAL SHAVINGS WERE COMING OUT OF IT. IT WAS CONFIRMED THAT THE EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7075222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K080802. One submission often covers many models, so check the product and lot details in each record.
- Z-2032-2017 · Oct 6, 2016 · Terminated20.2cm Heavy Duty Long Heavy Short Attachment
- Z-2031-2017 · Oct 6, 2016 · TerminatedSagittal MICRO-SAW
- Z-1956-2017 · Oct 6, 2016 · TerminatedSystem Console, Single Port, with irrigation Product Usage: Electric system
- Z-1948-2017 · Oct 6, 2016 · TerminatedElectronic Foot Control with Direction Switch Product Usage: Electric system
- Z-2001-2017 · Oct 6, 2016 · Terminated12.4cm Heavy Duty Long Heavy Short Attachment Product Usage: Pneumatic system
- Z-1952-2017 · Oct 6, 2016 · TerminatedeMax2 PLUS Product Usage: Electric system
- Z-2015-2017 · Oct 6, 2016 · TerminatedSagittal Saw Product Usage: Pneumatic system
- Z-2019-2017 · Oct 6, 2016 · TerminatedOscillating MICRO-SAW Product Usage: Pneumatic system
- Z-1955-2017 · Oct 6, 2016 · TerminatedConsole, eMax2 Plus Product Usage: Electric system
- Z-2026-2017 · Oct 6, 2016 · TerminatedReciprocating MICRO SAW
- Z-1957-2017 · Oct 6, 2016 · TerminatedSystem Console, Single Port, no irrigation
- Z-1949-2017 · Oct 6, 2016 · TerminatedElectronic Foot Control with Direction or Irrigation Switches Product Usage: Electric system
- Z-0727-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs: 1.5MMx7.5MM Cylindrical Carbide Drum part # S-15DRMC-L, 1MM Carbide Fluted ball, Extended part # S-1B-C, 1MM Carbide Fluted ball part # S-SB-C, 2MM C
- Z-0726-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs: 2MM Carbide Fluted Ball, Standard, 8 Flutes - part #QD8-2SB-C-8F, 4MM Carbide Fluted Ball, 10 Flutes - part # QD8-4B-C-10F 5MM Carbide Fluted Ball, 1
- Z-0729-2014 · Dec 17, 2013 · TerminatedAnspach Carbide Cutting Burrs & Drums: 2MM Carbide Fluted Ball, Extended part # S-2B-C and 7MM Carbide Fluted Ball part # S-7B-C Anspach cutting burrs are intended for cutting and
…and 20 more.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) devices
- CODMAN® 261221Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall223 reports
- Zyphr 5100060001Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall199 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)67 reports
- CODMAN® 261222Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall7 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)54 reports
- CODMAN® 261230Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall3 reports
- CODMAN® 261223Integra Lifesciences Production Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)1 recall2 reports
- EasyDrill Cranial Perforator Ø14 x Ø11 x 3mm DM0010FAAMicromar Indústria e Comércio Ltda · Drills, Burrs, Trephines & Accessories (Compound, Powered)37 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page