NA S-7B-G1Recalls & reported problems
Synthes GmbH · 7 mm Fluted Ball. Use with QD8-G1, SHORT-G1, QD8, QD8-S, SHORT Attachments. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384012658), received between Dec 22, 2023 and Dec 22, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
39 other recalls cover devices cleared under the same FDA submission (K113476); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture1
- Separation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2023Fracture; Separation Problem; Fracture; Separation Problem
IT WAS REPORTED BY SWITZERLAND THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DRILL DEVICE HAD A BROKEN TIP. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL INSPECTION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT FUNCTION. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18382786 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113476. One submission often covers many models, so check the product and lot details in each record.
- Z-2839-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 3mm FLUTED BALL; For cutting and shaping bone including spine and cranium.
- Z-2840-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 6MM CRS DBALL, 2mm Shaft EXP, M; For cutting and shaping bone including spine and cranium.
- Z-2838-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 3mm CRS DBALL, 2mm SHAFT EXP, M; For cutting and shaping bone including spine and cranium.
- Z-2837-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 7mm FBALL L; For cutting and shaping bone including spine and cranium.
- Z-2293-2015 · Apr 13, 2015 · TerminatedANSPACH - 3MM Fluted Ball, 2 MM Shaft Exposure; REF M-3SB-2-G1. Cutting shaping bone including spine and cranium.
- Z-2307-2015 · Apr 13, 2015 · TerminatedANSPACH -4MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-4DC-22-G1 Cutting shaping bone including spine and cranium.
- Z-2312-2015 · Apr 13, 2015 · TerminatedANSPACH - 2.3MM Fluted Matchstick, Extends 22MM; REF S-8N-L-G1 Cutting shaping bone including spine and cranium.
- Z-2297-2015 · Apr 13, 2015 · TerminatedANSPACH - 6MM x 11.3MM Fluted Barrel; REF M-6BA-A-G1 Cutting shaping bone including spine and cranium.
- Z-2306-2015 · Apr 13, 2015 · TerminatedANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
- Z-2296-2015 · Apr 13, 2015 · TerminatedANSPACH - 5MM x 11.3MM Fluted Barrel; REF M-5BA-A-G1 Cutting shaping bone including spine and cranium.
- Z-2284-2015 · Apr 13, 2015 · TerminatedANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
- Z-2313-2015 · Apr 13, 2015 · TerminatedANSPACH - 3MM Fluted Matchstick, Extends 8.9MM, Less Aggressive; REF S-8NS-M-G1 Cutting shaping bone including spine and cranium.
- Z-2282-2015 · Apr 13, 2015 · TerminatedANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
- Z-2291-2015 · Apr 13, 2015 · TerminatedANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2303-2015 · Apr 13, 2015 · TerminatedANSPACH - 5MM Fluted Ball, Extends 27MM; REF M-5B-27-G1. Cutting shaping bone including spine and cranium.
…and 24 more.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page