NA M-5DC-G1Recalls & reported problems
Synthes GmbH · 5 mm Coarse Diamond Ball. Use with MEDIUM-G1, QD11-G1, MEDIUM, QD11, QD11-S Attachments. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384013037), received between Sep 16, 2021 and Sep 16, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Dull, Blunt” (1 report).
39 other recalls cover devices cleared under the same FDA submission (K113476); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Dull, Blunt1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2021Dull, Blunt; Dull, Blunt
IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE CUTTER DEVICE WOULD NOT CUT/DULL. THE CUTTER WAS REMOVED FROM THE ATTACHMENT DEVICE AND IT WAS DETERMINED THAT THE CUTTING HEAD SHOWED SIGNS OF WEAR. IT WAS FURTHER OBSERVED THAT THE SHAFT HAD CORROSION/RUSTING/PITTING AND THE DEVICE FAILED VISUAL INSPECTION. IT WAS NOTED IN THE SERVICE ORDER THAT THE CUTTER DEVICE HAD AN UNDETERMINED MALFUNCTION, AND THE ATTACHMENT DEVICE DID NOT WORK ANYMORE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12484613 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113476. One submission often covers many models, so check the product and lot details in each record.
- Z-2837-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 7mm FBALL L; For cutting and shaping bone including spine and cranium.
- Z-2839-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 3mm FLUTED BALL; For cutting and shaping bone including spine and cranium.
- Z-2838-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 3mm CRS DBALL, 2mm SHAFT EXP, M; For cutting and shaping bone including spine and cranium.
- Z-2840-2016 · Aug 26, 2016 · TerminatedG1 Dissection Tool - 6MM CRS DBALL, 2mm Shaft EXP, M; For cutting and shaping bone including spine and cranium.
- Z-2306-2015 · Apr 13, 2015 · TerminatedANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
- Z-2285-2015 · Apr 13, 2015 · TerminatedANSPACH - Fluted Drum, Side Cut Only; 7MM x 14MM REF L-7DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2288-2015 · Apr 13, 2015 · TerminatedANSPACH - 8MM x 14MM Fluted Drum, Side Cut Only; REF L-8DRC-G1. Cutting shaping bone including spine and cranium.
- Z-2302-2015 · Apr 13, 2015 · TerminatedANSPACH - 4MM Fluted Ball, Extends 27MM; REF M-4B-27-G1. Cutting shaping bone including spine and cranium.
- Z-2291-2015 · Apr 13, 2015 · TerminatedANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
- Z-2309-2015 · Apr 13, 2015 · TerminatedANSPACH - 5MM Fluted Ball, 2MM Shaft Exposure; REF S-5B-2-G1 Cutting shaping bone including spine and cranium.
- Z-2313-2015 · Apr 13, 2015 · TerminatedANSPACH - 3MM Fluted Matchstick, Extends 8.9MM, Less Aggressive; REF S-8NS-M-G1 Cutting shaping bone including spine and cranium.
- Z-2308-2015 · Apr 13, 2015 · TerminatedANSPACH - 4MM Diamond Ball, Extends 27MM; REF S-4D-L-G1 all lot numbers Cutting shaping bone including spine and cranium.
- Z-2293-2015 · Apr 13, 2015 · TerminatedANSPACH - 3MM Fluted Ball, 2 MM Shaft Exposure; REF M-3SB-2-G1. Cutting shaping bone including spine and cranium.
- Z-2284-2015 · Apr 13, 2015 · TerminatedANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
- Z-2282-2015 · Apr 13, 2015 · TerminatedANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
…and 24 more.
Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices
- NA EMAX2PLUSSynthes GmbH · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)2,623 reports
- MIDAS REX MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)952 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)416 reports
- MIDAS REX MR8 MR8-AS07Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)232 reports
- MIDAS REX MR8 MR8-AA10Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)192 reports
- Midas Rex MR8 MR8-AF02Medtronic Powered Surgical Solutions · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)189 reports
- NA 5407-FA2-000Stryker Corporation · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)181 reports
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page