Skip to content

NA SHD-7B-G1Recalls & reported problems

Synthes GmbH · 7 mm Fluted Ball. Use with SHORT-HD-G1, SHORT-HD Attachments. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845384013563), received between Aug 9, 2023 and Aug 9, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Problem” (1 report).

39 other recalls cover devices cleared under the same FDA submission (K113476); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 9, 2023Separation Problem; Separation Problem

    IT WAS REPORTED THAT THE CUTTER DEVICE WAS STUCK IN THE MOTOR DEVICE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE COULD NOT BE DISASSEMBLED (REMOVED) AND WAS FRACTURED. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17490431 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113476. One submission often covers many models, so check the product and lot details in each record.

  • Z-2837-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 7mm FBALL L; For cutting and shaping bone including spine and cranium.
  • Z-2838-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 3mm CRS DBALL, 2mm SHAFT EXP, M; For cutting and shaping bone including spine and cranium.
  • Z-2840-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 6MM CRS DBALL, 2mm Shaft EXP, M; For cutting and shaping bone including spine and cranium.
  • Z-2839-2016 · Aug 26, 2016 · Terminated
    G1 Dissection Tool - 3mm FLUTED BALL; For cutting and shaping bone including spine and cranium.
  • Z-2282-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.
  • Z-2305-2015 · Apr 13, 2015 · Terminated
    ANSPACH -3MM Coarse Diamond Ball,72MM Shaft Exposure; REF S-3SDC-22-G1. Cutting shaping bone including spine and cranium.
  • Z-2311-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 5MM Coarse Diamond Ball, 22MM Shaft Exposure; REF S-5DC-22-G1 Cutting shaping bone including spine and cranium.
  • Z-2291-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Fluted Drum, Side Cut Only; 9MM x 14 MM REF L-9DRC-G1 Cutting shaping bone including spine and cranium.
  • Z-2281-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Fluted Drum, Side Cut only 5MM x 14MM , REF L-5DRC-G1. Cutting shaping bone including spine and cranium.
  • Z-2290-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Fluted Ball; 9MM REF L-9B-G1 Cutting shaping bone including spine and cranium.
  • Z-2284-2015 · Apr 13, 2015 · Terminated
    ANSPACH - Fluted Acorn, 7MM REF L-7AC-G1 Cutting shaping bone including spine and cranium.
  • Z-2306-2015 · Apr 13, 2015 · Terminated
    ANSPACH -4MM Carbide Fluted Ball, Extends 27MM; REF S-4BL-C-G1. Cutting shaping bone including spine and cranium.
  • Z-2292-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 3MM Course Diamond Ball, 2 MM Shaft Exposure; REF M-3DC-2-G1 Cutting shaping bone including spine and cranium.
  • Z-2297-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 6MM x 11.3MM Fluted Barrel; REF M-6BA-A-G1 Cutting shaping bone including spine and cranium.
  • Z-2293-2015 · Apr 13, 2015 · Terminated
    ANSPACH - 3MM Fluted Ball, 2 MM Shaft Exposure; REF M-3SB-2-G1. Cutting shaping bone including spine and cranium.

…and 24 more.

Other DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices

All DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page