Skip to content

Streamline MIS Spinal Fixation System 05-FINALDRIVER2Recalls & reported problems

Xtant Medical Holdings, Inc. · FINAL DRIVER · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00846468088453), received between Jul 2, 2024 and Jul 2, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL ONE (1) LEVEL TRANSFORAMINAL LUMBAR INTERBODY FUSION PROCEDURE WITH POSTERIOR FIXATION USING AN MIS APPROACH. DURING THE FINAL TIGHTENING STEP IN THE PROCEDURE, IT WAS NOTED THAT THE SET SCREW MAY HAVE BEEN TIGHTENED BEYOND THE TORQUE LIMIT, CAUSING THE DISTAL TIP OF THE DRIVER TO FRACTURE OFF. NO PIECES WERE RETAINED BY THE PATIENT. A SPARE FINAL DRIVER WAS USED TO FINISH THE REMAINDER OF THE SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19656516 ↗
  • MalfunctionJul 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TWO (2) LEVEL TRANSFORAMINAL LUMBAR INTERBODY FUSION PROCEDURE WITH POSTERIOR FIXATION USING AN MIS APPROACH. DURING THE FINAL TIGHTENING STEP IN THE PROCEDURE, IT WAS NOTED THAT THE SET SCREW MAY HAVE BEEN TIGHTENED BEYOND THE TORQUE LIMIT, CAUSING THE DISTAL TIP OF THE DRIVER TO FRACTURE OFF. NO PIECES WERE RETAINED BY THE PATIENT. A SPARE FINAL DRIVER WAS USED TO FINISH THE REMAINDER OF THE SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19656481 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page