Skip to content

Latitude™ EV 0030121Recalls & reported problems

Tornier, Inc. · Ulnar Stem - Coated · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00846832002580), received between Mar 1, 2024 and Nov 26, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Osseointegration Problem2
  • Device Dislodged or Dislocated1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Unspecified Infection3
  • Inadequate Osseointegration2
  • Failure of Implant1
  • Implant Pain1
  • Insufficient Information1
  • Joint Dislocation1
  • Loss of Range of Motion1
  • Swelling/ Edema1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES OF THE ELBOW JOINT REPLACEMENT. THE REPORT DETAILS ANALYSIS PROVIDED FOR REVISION PROCEDURES PERFORMED UNTIL (B)(6) 2024. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2024, A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. SINGLE STAGE REVISION (INCLUDES DAIR WITH MODULAR EXCHANGE AND MODULAR EXCHANGE FOR INDICATIONS OTHER THAN INFECTION).

    Patient problems: Unspecified Infection

    MAUDE report 23654363 ↗
  • InjuryOct 16, 2025Unstable; Unstable

    IT WAS REPORTED THAT, THE PATIENT WOULD REQUIRE A REVISION SURGERY FOR FAILURE OF ELBOW LOCKING MECHANISM. THE PLANNED DATE OF REVISION WAS (B)(6) 2025 2025. AS NO FURTHER INFORMATION COULD BE OBTAINED, THE REVISION SURGERY HAS BEEN CAPTURED AS IT WAS PLANNED.

    Patient problems: Insufficient Information

    MAUDE report 23316758 ↗
  • InjuryOct 10, 2025Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    IT WAS REPORTED THAT, THE PATIENT UNDERWENT OPEN REDUCTION, INTERNAL FIXATION OF FRACTURED LEFT DISTAL HUMERUS, IN 2017. SHE UNDERWENT TOTAL ELBOW REPLACEMENT ON (B)(6) 2017, USING TORNIER LATITUDE TER IMPLANTS. SHE HAS BEEN VERY HAPPY WITH THIS SURGERY UNTIL RECENT MONTHS BUT LATER PRESENTED WITH PAINFUL LEFT ELBOW, DECREASED RANGE OF MOTION, SWELLING. X-RAYS SUGGESTIVE OF POSSIBLE INFECTION AND LOOSENING. SURGERY FAILED DUE TO INFECTION. DURING SURGERY TODAY, GROSS AMOUNT OF SYNOVITIS TISSUE WAS FOUND, AND EXCISED. ULNAR COMPONENTS WERE FRANKLY LOOSE AND REMOVED BY THE SURGEON USING HIS FINGERS ONLY. MULTIPLE TISSUE SPECIMENS WERE TAKEN. HUMERAL COMPONENTS WERE STILL WELL FIXED, BUT WITH OSTEOLYSIS AROUND THE DISTAL PART OF THE PROSTHESIS. SURGEON DECIDED TO REMOVE THIS COMPONENT (WHICH WAS WELL FIXED), CLEAN OUT BOTH MEDULLARY CANALS, COPIOUS LAVAGE, AND IMPLANT REVISION SIZED EXTRA-LONG COMPONENTS. AS REPORTED, EXPLANTED COMPONENTS ARE AVAILABLE FOR RETURN. BOTH REVISION COMPONENTS CEMENTED IN USING ANTIBIOTIC CEMENT. JOINT STABLE ON REDUCTION AND WOUND CLOSED. SURGEON PERFORMED THIS, 1-STAGE REVISION ARTHROPLASTY BECAUSE OF MULTIPLE CO-MORBIDITIES THAT THE PATIENT SUFFERS FROM

    Patient problems: Implant Pain; Inadequate Osseointegration; Loss of Range of Motion; Swelling/ Edema; Unspecified Infection

    MAUDE report 23265553 ↗
  • InjuryJun 26, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE SCREW THAT CONNECTS THE LATITUDE ULNAR CAP IMPLANT TO THE LATITUDE ULNAR STEM IMPLANT DISASSOCIATED AND THEREFORE PATIENT WILL REQUIRE A REVISION SURGERY.

    Patient problems: Joint Dislocation

    MAUDE report 19620147 ↗
  • InjuryMar 1, 2024Naturally Worn; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Naturally Worn; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem

    DURING A SURGICAL PROCEDURE, THE #5 FLEXIBLE REAMER BROKE WHILE BEING USED ON A PATIENT'S ELBOW. BOTH PIECES OF THE REAMER WERE RETRIEVED, AND AN X-RAY WAS TAKEN, WHICH SHOWED NO ABNORMALITIES. THE ORIGINAL INTENDED PROCEDURE WAS THE REMOVAL OF HARDWARE FROM THE RIGHT ELBOW, INSERTION OF A CEMENT ANTIBIOTIC SPACER, IRRIGATION AND DEBRIDEMENT OF THE RIGHT ELBOW, AND BURSECTOMY OF THE OLECRANON BURSA. THE USER ENCOUNTERED PROBLEMS RELATED TO DEVICE FAILURE AND MALFUNCTION.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Unspecified Infection

    MAUDE report 18822906 ↗

Other PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices

All PROSTHESIS, ELBOW, CONSTRAINED, CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page