Skip to content

BioHorizons PGMU172Recalls & reported problems

Biohorizons Implant Systems, Inc · 4.5mm Multi-unit Abutment, 17-degree, 2.25mm Collar · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847236006426), received between Jan 31, 2023 and Apr 29, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K103691, K223697); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Break1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Fracture

    THIS REPORT COVERS FOUR (4) MALFUNCTION EVENT. A REVIEW OF THE EVENT INVOLVED THE DEVICE(S) EXPERIENCING A BROKEN ABUTMENT . NO PATIENT ADVERSE EVENT WAS REPORTED. NO INFORMATION REGARDING PATIENT DEMOGRAPHICS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 25035944 ↗
  • MalfunctionJan 30, 2026Break; Insufficient Device Problem Information

    THIS REPORT INDICATES ONE (1) MALFUNCTION EVENT. A REVIEW OF THE EVENT INVOLVED THE DEVICE(S) BEING A BROKEN ABUTMENT. NO PATIENT ADVERSE EVENT WAS REPORTED. NO INFORMATION REGARDING PATIENT DEMOGRAPHICS WERE PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 24223145 ↗
  • MalfunctionJan 31, 2023Fracture; Fracture

    THIS REPORT IS A SUMMARY OF ONE (1) MALFUNCTION EVENT. A REVIEW OF THE EVENT(S) INVOLVED A DEVICE EXPERIENCING A BROKEN HEX. NO ADVERSE EVENT PATIENT INFORMATION WAS REPORTED. NO INFORMATION REGARDING PATIENT DEMOGRAPHICS HAVE BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 16275260 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K103691, K223697. One submission often covers many models, so check the product and lot details in each record.

  • Z-1111-2025 · Dec 6, 2024 · Open, Classified
    Temporary Titanium Abutments

Other Abutment, Implant, Dental, Endosseous devices

All Abutment, Implant, Dental, Endosseous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page