Z Vent 799-AEB1-00-01Recalls & reported problems

Zoll Medical Corporation · MAIN ASSEMBLY, VENTILATOR, COMMERCIAL, ZVENT, BASIC · Class 2

All models of Z Vent (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zoll Z Vent is a facility ventilator that helps patients breathe. Many of its 188 reports describe error messages during self-tests, the device stopping ventilation, or failing to power up. Several reports say no patient was harmed. There is 1 injury report and 2 death reports. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
193
Deaths reported
2
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 193 adverse event reports that list this exact device (DI 00847946024192), received between Aug 1, 2019 and Sep 30, 2026 — classified as 2 deaths, 2 injuries, 189 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (51 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 2 reportsJan 2024: 3 reportsFeb 2024: 7 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 4 reportsNov 2024: 2 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 4 reportsJun 2025: 5 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 3 reportsNov 2025: 5 reportsDec 2025: 4 reportsJan 2026: 3 reportsFeb 2026: 6 reportsMar 2026: 4 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem51
  • Failure of Device to Self-Test29
  • Improper Flow or Infusion27
  • Failure to Power Up24
  • Calibration Problem18
  • Communication or Transmission Problem17
  • No Display/Image16
  • Infusion or Flow Problem10

Patient problems

  • No Clinical Signs, Symptoms or Conditions150
  • Insufficient Information25
  • No Patient Involvement12
  • No Consequences Or Impact To Patient3
  • Necrosis2
  • Brain Injury1
  • Inadequate Osseointegration1

Recent adverse event reports

The 25 most recent of 193 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Infusion or Flow Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "O2 SUPPLY FAILURE- 1020" ERROR MESSAGE.COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947438 ↗
  • MalfunctionSep 28, 2026Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKOWN), THE DEVICE DISPLAYED LOW SUPPLY O2 SUPPLY ERROR 1020.COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918872 ↗
  • MalfunctionSep 28, 2026Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN.COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915917 ↗
  • InjurySep 28, 2026Improper Flow or Infusion; Insufficient Device Problem Information

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKOWN), THE DEVICE STOPPED VENTILATING AND DISPLAYED AN UNRECOGNIZED "SHORT IN SYSTEM" ERROR MESSAGE.THE PATIENT RAPIDLY DESATURATE TO 89%. POWER CYCLING RESTORED AIRFLOW, AND BIPAP RESUMED AT 100% FIO2 WITHOUT FURTHER ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 50913688 ↗
  • MalfunctionSep 24, 2026Mechanical Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED A "COMPRESSOR SELF CHECK FAILURE - 1001" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50882913 ↗
  • MalfunctionAug 19, 2026Failure of Device to Self-Test

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323633 ↗
  • MalfunctionJun 15, 2026Infusion or Flow Problem

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED AN "02 VALVE FAULT - 2011" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25501498 ↗
  • MalfunctionMay 4, 2026Mechanical Problem

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "COMPRESSOR FAILURE -1001 " ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25073270 ↗
  • MalfunctionApr 23, 2026Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "OFF SET SELF CHECK FAILURE - 1003" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965315 ↗
  • MalfunctionApr 10, 2026Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE STOPPED VENTILATING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24845053 ↗
  • MalfunctionApr 3, 2026No Display/Image; Failure to Power Up; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE INAPPROPRIATELY SHUT DOWN AND THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781196 ↗
  • MalfunctionMar 30, 2026Gas Output Problem

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED AN "FRESH GAS INTAKE FAILURE -1030" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24727404 ↗
  • MalfunctionMar 30, 2026Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE STOPPED VENTILATING. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24727320 ↗
  • MalfunctionMar 30, 2026Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE STOPPED VENTILATING. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24727208 ↗
  • MalfunctionMar 3, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE WOULD NOT POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24498515 ↗
  • MalfunctionFeb 17, 2026Mechanical Problem; Unexpected Shutdown

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAY A "COMPRESSOR FAILURE- 1001" AND INAPPROPRIATELY SHUT DOWN. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24363253 ↗
  • MalfunctionFeb 17, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24363200 ↗
  • MalfunctionFeb 13, 2026No Display/Image

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24338258 ↗
  • MalfunctionFeb 12, 2026Improper Flow or Infusion

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE STOPPED VENTILATING. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24321485 ↗
  • MalfunctionFeb 10, 2026Mechanical Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "COMPRESSOR FAULT-2001 " ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24306559 ↗
  • MalfunctionFeb 4, 2026Mechanical Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE STOPPED FUNCTIONING AND DISPLAYED A "COMPRESSOR FAILURE-1001" ERROR MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259284 ↗
  • MalfunctionJan 26, 2026Calibration Problem

    COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT (AGE & GENDER UNKNOWN), THE DEVICE DISPLAYED A "RUNTIME CALIBRATION FAILURE - 1051" ERROR MESSAGE. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.

    Patient problems: Insufficient Information

    MAUDE report 24185038 ↗
  • MalfunctionJan 15, 2026Failure to Power Up

    COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE WOULD NOT POWER UP. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24081452 ↗
  • MalfunctionJan 12, 2026Mechanical Problem; Failure to Power Up; Failure of Device to Self-Test

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE STOPPED VENTILATING. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24036428 ↗
  • MalfunctionDec 23, 2025No Display/Image; No Display/Image

    COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23878529 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page