NA 210110Recalls & reported problems
Mako Surgical Corp. · SLED · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00848486031848), received between Aug 15, 2018 and Jun 15, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Mechanical Problem4
- Crack2
- Material Frayed2
- Sharp Edges2
- Device Slipped1
- Non Reproducible Results1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
- Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2023Mechanical Problem; Mechanical Problem
THE BLUE HANDLE WOULD NOT TIGHTEN CAUSING THE LEG TO BE UNSTABLE CASE TYPE / APPLICATION: TKA. PER RESPONSE: "THE BLUE KNOB WOULD NOT TIGHTEN, AND THE PATIENT'S LEG EXTERNALLY ROTATED OFF THE TABLE." ADDITIONAL INFORMATION 6/27/2023: DID THE PATIENT¿S LEG FALL OFF THE TABLE? NO, THE BLUE KNOB WAS NOT TIGHTENING DOWN WHICH ALLOWED THE LEG TO BE UNSTABLE AND ROTATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17140224 ↗MalfunctionNov 12, 2020Mechanical Problem; Device Slipped; Mechanical Problem; Device Slipped
SETTING UP FOR A MAKO PKA CASE, DR WAS ASSEMBLING THE MAKO LEG POSITIONER ONTO THE BED AND PATIENT LEG. WHEN DR PLACED THE BOOT & PATIENT LEG INTO THE SLED- THE LEG POSITIONER SLED WOULD NOT HOLD THE WEIGHT OF THE LEG IN POSITION. THE LEG WOULD KEEP FALLING LATERAL WHENEVER THE LEG WAS LET GO BY THE SURGEON. NO DELAYS TO SURGERY, SURGERY WAS COMPLETED BY ASSISTANT MANUALLY HOLDING THE LEG IN PLACE TO SUPPORT IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10828809 ↗MalfunctionJan 10, 2020Crack; Material Frayed; Sharp Edges; Crack; Material Frayed; Sharp Edges
STRYKER MAKO LEG HOLDERS. NUMEROUS PIECES HAVE EXPOSED FIBERS, TWO PIECES HAVE STRESS FRACTURES AT THE POINT WHERE THE METAL FITTING IS INSERTED. REPORT OF "SOMEONE GETTING STUCK" BY AN EXPOSED FIBER. CASE TYPE: TKA. PATIENT WAS UNDER ANESTHESIA. UPDATE : HOW MANY SURGERIES WERE INVOLVED WITH THE REPORTED ISSUE? APPROXIMATELY 300 PROCEDURES TO GET TO THIS POINT. ALL THE REPORTED DEVICES THAT ARE INCLUDED IN THIS COMPLAINT WERE FOR ONE OR MANY SURGERIES? MANY ¿ SPD CALLED US CONCERNED ABOUT EXPOSED FIBERS AND CRACKS. HAVE YOU REPORTED EACH INCIDENT SEPARATELY THAT HAS THE ISSUES YOU HAVE MENTIONED IN THIS COMPLAINT? (IF NO, PLEASE SUBMIT A COMPLAINT FOR EACH ISSUE.) NO ¿ MULTIPLE LEG HOLDERS ARE EXHIBITING WEAR. THE ITEMS LISTED IN THE PER EITHER HAVE EXPOSED FIBERS OR CRACKS. PLEASE CONFIRM IF THERE WAS A PATIENT THAT WAS ¿STUCK¿ BY THE EXPOSED CARBON FIBER? NOT A PATIENT ¿ A STAFF MEMBER.
Patient problems: Injury; No Known Impact Or Consequence To Patient
MAUDE report 9577650 ↗MalfunctionJan 10, 2020Crack; Material Frayed; Sharp Edges; Crack; Material Frayed; Sharp Edges
STRYKER MAKO LEG HOLDERS. NUMEROUS PIECES HAVE EXPOSED FIBERS, TWO PIECES HAVE STRESS FRACTURES AT THE POINT WHERE THE METAL FITTING IS INSERTED. REPORT OF "SOMEONE GETTING STUCK" BY AN EXPOSED FIBER. CASE TYPE: TKA. PATIENT WAS UNDER ANESTHESIA. UPDATE : HOW MANY SURGERIES WERE INVOLVED WITH THE REPORTED ISSUE? APPROXIMATELY 300 PROCEDURES TO GET TO THIS POINT. ALL THE REPORTED DEVICES THAT ARE INCLUDED IN THIS COMPLAINT WERE FOR ONE OR MANY SURGERIES? MANY ¿ SPD CALLED US CONCERNED ABOUT EXPOSED FIBERS AND CRACKS. HAVE YOU REPORTED EACH INCIDENT SEPARATELY THAT HAS THE ISSUES YOU HAVE MENTIONED IN THIS COMPLAINT? (IF NO, PLEASE SUBMIT A COMPLAINT FOR EACH ISSUE.) NO ¿ MULTIPLE LEG HOLDERS ARE EXHIBITING WEAR. THE ITEMS LISTED IN THE PER EITHER HAVE EXPOSED FIBERS OR CRACKS. PLEASE CONFIRM IF THERE WAS A PATIENT THAT WAS ¿STUCK¿ BY THE EXPOSED CARBON FIBER? NOT A PATIENT ¿ A STAFF MEMBER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9577644 ↗MalfunctionAug 24, 2018Mechanical Problem; Mechanical Problem
LEG KEPT FALLING TO THE SIDE WHILE LOCKED INTO PLACE. CASE TYPE: TKA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7812814 ↗MalfunctionAug 15, 2018Mechanical Problem; Non Reproducible Results; Mechanical Problem; Non Reproducible Results
SLED WOULD NOT TIGHTEN ENOUGH TO KEEP LEG UPRIGHT. LEG KEEP FALLING EITHER INTERNAL OR EXTERNAL. SURGICAL DELAY YES -16-30 MINUTES; CASE TYPE: TKA.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7786656 ↗
Other Operating Room Accessories Table Tray devices
- MAYFIELD® A2101Integra Lifesciences Corporation · Operating Room Accessories Table Tray101 reports
- NA 210080Mako Surgical Corp. · Operating Room Accessories Table Tray74 reports
- ALLEN 104002Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall8 reports
- N/A 7210569Smith & Nephew, Inc. · Accessories, operating-room, table (kit)57 reports
- ALLEN 107004Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall5 reports
- ALLEN 108001Hill-Rom, Inc. · Operating Room Accessories Table Tray1 recall3 reports
- NA 7210551Smith & Nephew, Inc. · Accessories, operating-room, table (kit)41 reports
- MAYFIELD® A1018Integra Lifesciences Corporation · Operating Room Accessories Table Tray36 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page