CORTICAL BONE SCREW 3.5mm x 28mm 8110-028Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00848665033212), received between Feb 2, 2024 and Mar 6, 2025 — classified as 8 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Patient Device Interaction Problem4
Patient problems
- Failure of Implant3
- Foreign Body In Patient2
- Hypersensitivity/Allergic reaction2
- Implant Pain2
- Malunion of Bone2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryMar 6, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A PATIENT UNDERWENT A PROXIMAL HUMERUS FRACTURE PROCEDURE ON (B)(6) 2023. A 3024-011 LEFT ALPHA PLATE, AN AR-8935-24 LOW PROFILE SCREW, TITANIUM, NON-LOCKING, AN 8110-026 CORTICAL BONE SCREW, AN 8110-024 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREW, TWO 8124-036 CANCELLOUS LOCKING SCREW, TWO 8124-044 CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, AN 8124-048 CANCELLOUS LOCKING SCREW, AN 8124-046 CANCELLOUS LOCKING SCREW, THREE 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-042 CANCELLOUS LOCKING SCREW, AN 8124-038 CANCELLOUS LOCKING SCREW, AND AN 8124-030 CANCELLOUS LOCKING SCREW WERE IMPLANTED. DUE TO NONUNION, THE PATIENT UNDERWENT A SECOND SURGERY ON (B)(6) 2024 TO HAVE AN AR-18827P-16 20 HOLE REINFORCED STRAIGHT PLATE, AN AR-18827-30 CORTICAL SCREW, AR-18827-18 CORTICAL SCREW, AR-18827V-18, AN AR-18827V-16 VAL SCREW, AN AR-18827V-22 VAL SCREW, TWO AR-18827V-20 VAL SCREW AN AR-18827V-10 VAL SCREW, AND AN ABS-2010-10 ALLOSYNC PURE IMPLANTED. ALL THE PREVIOUSLY IMPLANTED PRODUCTS DURING THE INITIAL SURGERY ON (B)(6) 2023 REMAINED IN THE PATIENT.
Patient problems: Malunion of Bone
MAUDE report 21536507 ↗InjuryMar 6, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT A PATIENT UNDERWENT A PROXIMAL HUMERUS FRACTURE PROCEDURE ON (B)(6) 2023. A 3024-011 LEFT ALPHA PLATE, AN AR-8935-24 LOW PROFILE SCREW, TITANIUM, NON-LOCKING, AN 8110-026 CORTICAL BONE SCREW, AN 8110-024 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREW, TWO 8124-036 CANCELLOUS LOCKING SCREW, TWO 8124-044 CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, AN 8124-048 CANCELLOUS LOCKING SCREW, AN 8124-046 CANCELLOUS LOCKING SCREW, THREE 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-042 CANCELLOUS LOCKING SCREW, AN 8124-038 CANCELLOUS LOCKING SCREW, AND AN 8124-030 CANCELLOUS LOCKING SCREW WERE IMPLANTED. DUE TO NONUNION, THE PATIENT UNDERWENT A SECOND SURGERY ON (B)(6) 2024 TO HAVE AN AR-18827P-16 20 HOLE REINFORCED STRAIGHT PLATE, AN AR-18827-30 CORTICAL SCREW, AR-18827-18 CORTICAL SCREW, AR-18827V-18, AN AR-18827V-16 VAL SCREW, AN AR-18827V-22 VAL SCREW, TWO AR-18827V-20 VAL SCREW AN AR-18827V-10 VAL SCREW, AND AN ABS-2010-10 ALLOSYNC PURE IMPLANTED. ALL THE PREVIOUSLY IMPLANTED PRODUCTS DURING THE INITIAL SURGERY ON (B)(6) 2023 REMAINED IN THE PATIENT.
Patient problems: Malunion of Bone
MAUDE report 21536472 ↗InjuryMar 3, 2025Break; Break
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN ALPHA PLATE BROKE POSTOPERATIVELY. THE PATIENT WAS PICKING UP THE CAT AND FELT A BREAK. X-RAYS CONFIRM THE PLATE BROKE THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. NO ADDITIONAL INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED ON 2/6/2025: ON (B)(6) 2024, THE PATIENT UNDERWENT REVISION SURGERY DUE TO THE LEFT ALPHA PLATE BREAKING AND CAUSING THE REINFORCED STRAIGHT PLATE TO DISLODGE. A 3024-011 LEFT ALPHA PLATE, AN AR-8935-24 LOW PROFILE SCREW, TITANIUM, NON-LOCKING, AN 8110-026 CORTICAL BONE SCREW, AN 8110-024 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREW, TWO 8124-036 CANCELLOUS LOCKING SCREW, TWO 8124-044 CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, AN 8124-048 CANCELLOUS LOCKING SCREW, AN 8124-046 CANCELLOUS LOCKING SCREW, THREE 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-042 CANCELLOUS LOCKING SCREW, AN 8124-038 CANCELLOUS LOCKING SCREW, AND AN 8124-030 CANCELLOUS LOCKING SCREW, AN AR-18827P-16 20 HOLE REINFORCED STRAIGHT PLATE, AN AR-18827-30 CORTICAL SCREW, AR-18827-18 CORTICAL SCREW, AR-18827V-18, AN AR-18827V-16 VAL SCREW, AN AR-18827V-22 VAL SCR
Patient problems: Foreign Body In Patient
MAUDE report 21503903 ↗InjuryMar 3, 2025Break; Break
ON 12/19/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN ALPHA PLATE BROKE POSTOPERATIVELY. THE PATIENT WAS PICKING UP THE CAT AND FELT A BREAK. X-RAYS CONFIRM THE PLATE BROKE THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. NO ADDITIONAL INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION RECEIVED ON 2/6/2025: ON (B)(6) 2024, THE PATIENT UNDERWENT REVISION SURGERY DUE TO THE LEFT ALPHA PLATE BREAKING AND CAUSING THE REINFORCED STRAIGHT PLATE TO DISLODGE. A 3024-011 LEFT ALPHA PLATE, AN AR-8935-24 LOW PROFILE SCREW, TITANIUM, NON-LOCKING, AN 8110-026 CORTICAL BONE SCREW, AN 8110-024 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREW, TWO 8124-036 CANCELLOUS LOCKING SCREW, TWO 8124-044 CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, AN 8124-048 CANCELLOUS LOCKING SCREW, AN 8124-046 CANCELLOUS LOCKING SCREW, THREE 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-042 CANCELLOUS LOCKING SCREW, AN 8124-038 CANCELLOUS LOCKING SCREW, AND AN 8124-030 CANCELLOUS LOCKING SCREW, AN AR-18827P-16 20 HOLE REINFORCED STRAIGHT PLATE, AN AR-18827-30 CORTICAL SCREW, AR-18827-18 CORTICAL SCREW, AR-18827V-18, AN AR-18827V-16 VAL SCREW, AN AR-18827V-22 VAL SCRE
Patient problems: Foreign Body In Patient
MAUDE report 21503896 ↗InjuryDec 12, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-01 11-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND REVISION SURGERY IS STILL TO BE DETERMINED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: DURING THE INITIAL SURGERY ON (B)(6) 2024, A 3024-011 11-HOLE LEFT ALPHA PLATE, TWO 8110-026 CORTICAL BONE SCREW, AN 8110-028 CORTICAL BONE SCREW, THREE 8110-030 CORTICAL BONE SCREWS, AN 8114-024 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-028 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-030 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-028 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 CANCELLOUS LOCKING SCREW, FULLY THREADED, AN 8124-038 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-046 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-048 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-050 FULLY THREADED CANCELLOUS LOCKING SCREW, A
Patient problems: Failure of Implant
MAUDE report 20914358 ↗MalfunctionNov 27, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-016 16-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND ON (B)(6) 2024, THE PLATE WAS REMOVED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: THE CORRECT ORIGINAL SURGERY DATE IS (B)(6) 2024. THE PATIENT WAS IMPLANTED WITH THE 3024-016 16-HOLE LEFT ALPHA PLATE, AN 8110-026 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREWS, FOUR 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-036 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-030 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-036 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-042 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO AR-7268T TIGERTAPE CERCLAGE WITH FIBERLINK SHUTTLE SUTURE, AND AN ABS-2010-10 ALLOSYNC PURE. DURING THE REVISION SURGERY, ALL THE HARDWARE WAS REMOVED. THE REVISION SURGERY WAS COMPLETED USI
Patient problems: Failure of Implant
MAUDE report 20791074 ↗InjuryNov 26, 2024Break; Break
ON 10/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A 3024-016 16-HOLE LEFT ALPHA PLATE BROKE POSTOPERATIVELY THROUGH THE OBLONG HOLE IN THE FIRST BEND OF THE PLATE. ON (B)(6) 2024, THE PATIENT WAS IMPLANTED WITH THE PLATE, AND ON (B)(6) 2024, THE PLATE WAS REMOVED. PER THE SALES REPRESENTATIVE, THE PATIENT DID NOT REPORT ANY FALLS OR INJURIES. NO ADDITIONAL INFORMATION HAS BEEN REPORTED. ADDITIONAL INFORMATION RECEIVED ON 11/13/2024: THE CORRECT ORIGINAL SURGERY DATE IS (B)(6) 2024. THE PATIENT WAS IMPLANTED WITH THE 3024-016 16-HOLE LEFT ALPHA PLATE, AN 8110-026 CORTICAL BONE SCREW, TWO 8110-028 CORTICAL BONE SCREWS, FOUR 8114-026 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8114-036 DOUBLE LEAD LOCK CORTICAL SCREW, AN 8124-030 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-032 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-036 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW, AN 8124-042 FULLY THREADED CANCELLOUS LOCKING SCREW, TWO AR-7268T TIGERTAPE CERCLAGE WITH FIBERLINK SHUTTLE SUTURE, AND AN ABS-2010-10 ALLOSYNC PURE. DURING THE REVISION SURGERY, ALL THE HARDWARE WAS REMOVED. THE REVISION SURGERY WAS COMPLETED USI
Patient problems: Failure of Implant
MAUDE report 20787348 ↗InjuryFeb 2, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ON 01/19/2024, IT WAS REPORTED BY A PATIENT WHO UNDERWENT A PROCEDURE WHERE ARTHREX DEVICES WERE IMPLANTED WAS EXPERIENCING A PAINFUL REACTION. THE INITIAL PROCEDURE OCCURRED ON (B)(6) 2023. THE PATIENT BEGAN TO EXPERIENCE PAIN 3-4 WEEKS POST. THE PATIENT WAS REFERRED TO A PAIN SPECIALIST WHO SUGGESTED SHE WAS HAVING AN ALLERGIC REACTION AS THE SCAR WAS RED AND PUFFY. SHE NOTED SHE FEELS TERRIBLE PAIN AND THROBBING, AND WHEN ATTEMPTING TO LIFT, IT'S PAINFUL. THE SURGEON MENTIONED SHE COULD HAVE THE DEVICES REMOVED IN 6 MONTHS FROM HER SURGERY DATE. HOWEVER, SHE IS NOT SURE SHE CAN TAKE THE PAIN THAT MUCH LONGER. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.
Patient problems: Hypersensitivity/Allergic reaction; Implant Pain
MAUDE report 18634341 ↗InjuryFeb 2, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2024, IT WAS REPORTED BY A PATIENT WHO UNDERWENT A PROCEDURE WHERE ARTHREX DEVICES WERE IMPLANTED WAS EXPERIENCING A PAINFUL REACTION. THE INITIAL PROCEDURE OCCURRED ON (B)(6) 2023. THE PATIENT BEGAN TO EXPERIENCE PAIN 3-4 WEEKS POST. THE PATIENT WAS REFERRED TO A PAIN SPECIALIST WHO SUGGESTED SHE WAS HAVING AN ALLERGIC REACTION AS THE SCAR WAS RED AND PUFFY. SHE NOTED SHE FEELS TERRIBLE PAIN AND THROBBING, AND WHEN ATTEMPTING TO LIFT, IT'S PAINFUL. THE SURGEON MENTIONED SHE COULD HAVE THE DEVICES REMOVED IN 6 MONTHS FROM HER SURGERY DATE. HOWEVER, SHE IS NOT SURE SHE CAN TAKE THE PAIN THAT MUCH LONGER. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.
Patient problems: Hypersensitivity/Allergic reaction; Implant Pain
MAUDE report 18632802 ↗
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page