CORTICAL BONE SCREW 3.5mm x 32mm 8110-032Recalls & reported problems
Advanced Orthopaedic Solutions, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00848665033236), received between Jul 24, 2025 and Nov 26, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem2
- Break1
Patient problems
- Nonunion/Delayed-union Bone Fracture2
- Pain2
- Unspecified Infection2
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2025Break; Break
ADDITIONAL INFORMATION: DURING THE PROCEDURE, A 3025-018 ALPHA PLATE, RIGHT, 18-HOLE WAS USED WHEN AN 8110-032 CORTICAL BONE SCREW (3.5 MM X 32 MM) FRACTURED DURING INSERTION. NO OTHER SCREWS WERE IMPLICATED, AND NO DAMAGE OR MALFUNCTION WAS REPORTED WITH THE ALPHA PLATE. TO COMPLETE THE CASE, ADDITIONAL SCREWS WERE USED, INCLUDING TWO AR-8935-26 LOW PROFILE SCREW (3.5 X 26 MM) AND ONE AR-8935 LOW PROFILE SCREW FROM THE TITANIUM ANKLE FRACTURE SET. THE SURGERY WAS DELAYED BY APPROXIMATELY 45 MINUTES, AND IT IS UNCLEAR WHETHER ADDITIONAL ANESTHESIA WAS ADMINISTERED DURING THIS TIME. THE PATIENT¿S BONE QUALITY WAS ASSESSED AS GOOD, AND BONE PREPARATION FOR IMPLANT INSERTION WAS PERFORMED USING AN AR-8943-30 2.5 MM DRILL BIT FROM THE TITANIUM ANKLE FRACTURE SET. NO FURTHER INTERVENTION WAS CONSIDERED OR PLANNED TO RETRIEVE THE SCREW FRAGMENT.
Patient problems: Foreign Body In Patient; Insufficient Information
MAUDE report 23659359 ↗InjuryJul 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON 6/27/2025, A PATIENT'S LEGAL REPRESENTATIVE REPORTED VIA EMAIL THAT THE PATIENT UNDERWENT OPEN REDUCTION AND INTERNAL FIXATION OF A COMMINUTED FRACTURE OF THE LEFT HUMERAL NECK ON (B)(6) 2024. FOLLOWING THE PROCEDURE, THE PATIENT EXPERIENCED ONGOING COMPLICATIONS, INCLUDING PERSISTENT PAIN, NONUNION OF THE FRACTURE, AND SIGNIFICANT DISCOMFORT WITH ANY MOVEMENT OF THE LEFT SHOULDER. ON (B)(6) 2024, THE PATIENT WAS INFORMED THAT A TARGETER/DRILL GUIDE HAD BEEN INADVERTENTLY RETAINED FROM THE INITIAL PROCEDURE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 TO REMOVE THE RETAINED INSTRUMENT, DEBRIDE THE INFECTED AREA, AND REPLACE THE SURGICAL HARDWARE. ADDITIONAL INFORMATION RECEIVED ON 7/2/2025: DURING THE REVISION SURGERY ON (B)(6) 2024, THE FOLLOWING COMPONENTS WERE REMOVED: ONE 3022-005 PROXIMAL HUMERAL PLATE; TWO 8110-032 CORTICAL SCREWS; TWO 8110-035 CORTICAL SCREWS; ONE 8110-040 CORTICAL SCREW; ONE 8110-045 CORTICAL SCREW; ONE 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW; ONE 8124-044 FULLY THREADED CANCELLOUS LOCKING SCREW; THREE 8124-046 FULLY THREADED CANCELLOUS LOCKING SCREWS; TWO 8124-048 FULLY THREADED CANCELLOUS LOCKING SCREWS; AND ONE 8124-050 FULLY THREADED CAN
Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Unspecified Infection
MAUDE report 22586364 ↗InjuryJul 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ON 6/27/2025, A PATIENT'S LEGAL REPRESENTATIVE REPORTED VIA EMAIL THAT THE PATIENT UNDERWENT OPEN REDUCTION AND INTERNAL FIXATION OF A COMMINUTED FRACTURE OF THE LEFT HUMERAL NECK ON (B)(6) 2024. FOLLOWING THE PROCEDURE, THE PATIENT EXPERIENCED ONGOING COMPLICATIONS, INCLUDING PERSISTENT PAIN, NONUNION OF THE FRACTURE, AND SIGNIFICANT DISCOMFORT WITH ANY MOVEMENT OF THE LEFT SHOULDER. ON (B)(6) 2024, THE PATIENT WAS INFORMED THAT A TARGETER/DRILL GUIDE HAD BEEN INADVERTENTLY RETAINED FROM THE INITIAL PROCEDURE. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2024 TO REMOVE THE RETAINED INSTRUMENT, DEBRIDE THE INFECTED AREA, AND REPLACE THE SURGICAL HARDWARE. ADDITIONAL INFORMATION RECEIVED ON 7/2/2025: DURING THE REVISION SURGERY ON (B)(6) 2024, THE FOLLOWING COMPONENTS WERE REMOVED: ONE 3022-005 PROXIMAL HUMERAL PLATE; TWO 8110-032 CORTICAL SCREWS; TWO 8110-035 CORTICAL SCREWS; ONE 8110-040 CORTICAL SCREW; ONE 8110-045 CORTICAL SCREW; ONE 8124-040 FULLY THREADED CANCELLOUS LOCKING SCREW; ONE 8124-044 FULLY THREADED CANCELLOUS LOCKING SCREW; THREE 8124-046 FULLY THREADED CANCELLOUS LOCKING SCREWS; TWO 8124-048 FULLY THREADED CANCELLOUS LOCKING SCREWS; AND ONE 8124-050 FULLY THREADED CAN
Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Unspecified Infection
MAUDE report 22586266 ↗
Other Screw, Fixation, Bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page