Skip to content

Percussionaire F00008-1 (REF S00018-1)Recalls & reported problems

Percussionaire Corporation · VDR4, W/MONITRON II · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00849436000044), received between Jan 24, 2024 and Dec 16, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System1
  • Display or Visual Feedback Problem1
  • Improper Flow or Infusion1
  • Infusion or Flow Problem1
  • Insufficient Device Problem Information1
  • Key or Button Unresponsive/not Working1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Low Oxygen Saturation1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    ON (B)(6) 2025, IT WAS REPORTED THAT WHILE A VDR-4 DEVICE WAS IN USE ON A PATIENT, THE CUSTOMER REPRESENTATIVE HAD DIFFICULTY ADJUSTING THE CONTROLS ON THE VDR-4 VENTILATOR. IT WAS COMMUNICATED THAT THE DEVICE CONTROLS WERE UNRESPONSIVE AND THE VENTILATOR CONTINUOUSLY ALARMED. THE VENTILATOR HAD TO BE EXCHANGED TO AVOID ANY FURTHER PATIENT DECLINE. DUE TO THE UNINTENDED MEDICAL INTERVENTION, THIS REPORT IS BEING SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23821413 ↗
  • MalfunctionNov 14, 2024No Apparent Adverse Event; Insufficient Device Problem Information; No Apparent Adverse Event; Insufficient Device Problem Information

    A COMPLAINT WAS RECEIVED BY A CUSTOMER USING A VDR-4 STATING THAT WHILE THE DEVICE WAS BEING USED ON A PATIENT, THE DEVICE'S SECONDARY DISCONNECT ALARM FAILED TO ALARM. IT WAS CONFIRMED BY THS CUSTOMER THAT THE ALARM FUNCTION WAS WORKING PROPERLY DURING THE PRE-USE CHECK. NO PATIENT HARM OR INJURY WAS REPORTED DUE TO THIS FAILURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20691535 ↗
  • MalfunctionJan 25, 2024Display or Visual Feedback Problem; Infusion or Flow Problem; Display or Visual Feedback Problem; Infusion or Flow Problem

    PER THE CUSTOMER COMPLAINT, WHILE THE DEVICE WAS IN USE ON A PATIENT, IT WAS STATING THE VENTILATOR DEVICE WAS NOT FUNCTIONING CORRECTLY, AS WELL AS NOT DISPLAYING ACCURATE PATIENT DATA. THE PATIENT WAS STATED TO BE STABLE AND NOT COMPROMISED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18579916 ↗
  • MalfunctionJan 24, 2024Device Alarm System; Device Alarm System

    PER CUSTOMER COMPLAINT RECEIVED, WHILE THE PATIENT WAS USING THE VDR DEVICE FOR VENTILATION, THE PATIENT EXPERIENCED DESAT. THE BLENDER ALARM SOUNDED. THE RESPIRATORY THERAPIST ALSO NOTICED THE VENT PRESSURE DECREASING. THE PATIENT WAS MOVED TO ANOTHER VENTILATOR. NO PATIENT HARM REPORTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 18576305 ↗
  • MalfunctionJan 24, 2024Improper Flow or Infusion; Improper Flow or Infusion

    PER CUSTOMER FEEDBACK, A VDR VENTILATOR FAILED WHILE IN USE ON A PATIENT. IT WAS STATED BY THE CUSTOMER THAT NO OXYGEN WAS BEING DELIVERED TO THE PATIENT. NO PATIENT INJURY OR HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18573398 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.

  • Z-0960-2025 · Dec 23, 2024 · Open, Classified
    VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR
  • Z-0693-2024 · Mar 28, 2023 · Open, Classified
    Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
  • Z-0259-2024 · Sep 14, 2020 · Completed
    Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page