Skip to content

Percussionaire F00038-2Recalls & reported problems

Percussionaire Corporation · Sinusoidal Bronchotron and Universal Monitron Transport. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849436000235), received between Jun 28, 2016 and Feb 13, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Infusion or Flow Problem1
  • Leak/Splash1
  • Noise, Audible1
  • Unable to Obtain Readings1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2024Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED BY THE END USER THAT ON JANUARY 2, 2020, THE CARE TEAM ARRIVED TO THE PATIENT'S LOCATION AND NOTICED THE PATIENT WAS IN A STATE OF EXTREMIS. WHEN ATTEMPTING TO USE THE VENTILATOR, IT FAILED PRIOR TO USE ON THE PATIENT. IT WAS STATED THE DEVICE WAS UNABLE TO REGISTER PRESSURE ON THE DEVICE MULTIMETER AND A LOUD LEAK WAS HEARD.THE PATIENT HAD TO BE MANUALLY VENTILATED DURING TRANSPORT TO THE HOSPITAL. A SECOND INCIDENT OCCURRED ON (B)(6) 2020, IN WHICH AGAIN THE SAME VENTILATOR DEVICE FAILED PRIOR TO USE ON A PATIENT IN THE SAME MANNER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18699241 ↗
  • MalfunctionJan 24, 2024Infusion or Flow Problem; Noise, Audible; Infusion or Flow Problem; Noise, Audible

    PER THE CUSTOMER, THE RESPIRATORY THERAPIST ACCOMPANYING THE PATIENT WHILE IN TRANSPORT, INDICATED THAT THE DEVICE STARTED CHUGGING/HESITATING AND THEN STOPPED WORKING COMPLETELY. THE PATIENT HAD TO BE MANUALLY VENTILATED.

    Patient problems: Insufficient Information

    MAUDE report 18571872 ↗
  • MalfunctionJun 28, 2016Leak/Splash; Device Operates Differently Than Expected

    POSSIBLE LEAK IN DEVICE. DEVICE USED A FULL M TANK IN 2 AND A HALF HOURS. BIOMED PERSONNEL DID A CHECK WITH A FLUKE DEVICE AND SAW SOME LEAKAGE WHILE THEY WERE RUNNING THE MACHINE. SHORTLY AFTER THIS, THE MACHINE WASN'T FUNCTIONING CORRECTLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5757068 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.

  • Z-0960-2025 · Dec 23, 2024 · Open, Classified
    VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR
  • Z-0693-2024 · Mar 28, 2023 · Open, Classified
    Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
  • Z-0259-2024 · Sep 14, 2020 · Completed
    Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →

Sources and updates