Percussionaire F00038-2Recalls & reported problems
Percussionaire Corporation · Sinusoidal Bronchotron and Universal Monitron Transport. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00849436000235), received between Jun 28, 2016 and Feb 13, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Infusion or Flow Problem1
- Leak/Splash1
- Noise, Audible1
- Unable to Obtain Readings1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2024Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED BY THE END USER THAT ON JANUARY 2, 2020, THE CARE TEAM ARRIVED TO THE PATIENT'S LOCATION AND NOTICED THE PATIENT WAS IN A STATE OF EXTREMIS. WHEN ATTEMPTING TO USE THE VENTILATOR, IT FAILED PRIOR TO USE ON THE PATIENT. IT WAS STATED THE DEVICE WAS UNABLE TO REGISTER PRESSURE ON THE DEVICE MULTIMETER AND A LOUD LEAK WAS HEARD.THE PATIENT HAD TO BE MANUALLY VENTILATED DURING TRANSPORT TO THE HOSPITAL. A SECOND INCIDENT OCCURRED ON (B)(6) 2020, IN WHICH AGAIN THE SAME VENTILATOR DEVICE FAILED PRIOR TO USE ON A PATIENT IN THE SAME MANNER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18699241 ↗MalfunctionJan 24, 2024Infusion or Flow Problem; Noise, Audible; Infusion or Flow Problem; Noise, Audible
PER THE CUSTOMER, THE RESPIRATORY THERAPIST ACCOMPANYING THE PATIENT WHILE IN TRANSPORT, INDICATED THAT THE DEVICE STARTED CHUGGING/HESITATING AND THEN STOPPED WORKING COMPLETELY. THE PATIENT HAD TO BE MANUALLY VENTILATED.
Patient problems: Insufficient Information
MAUDE report 18571872 ↗MalfunctionJun 28, 2016Leak/Splash; Device Operates Differently Than Expected
POSSIBLE LEAK IN DEVICE. DEVICE USED A FULL M TANK IN 2 AND A HALF HOURS. BIOMED PERSONNEL DID A CHECK WITH A FLUKE DEVICE AND SAW SOME LEAKAGE WHILE THEY WERE RUNNING THE MACHINE. SHORTLY AFTER THIS, THE MACHINE WASN'T FUNCTIONING CORRECTLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5757068 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.
- Z-0960-2025 · Dec 23, 2024 · Open, ClassifiedVDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR
- Z-0693-2024 · Mar 28, 2023 · Open, ClassifiedDistal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
- Z-0259-2024 · Sep 14, 2020 · CompletedMonitron II Waveform Analyzer, REF: S00015-1, used with ventilators.
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page