Percussionaire A50094-DRecalls & reported problems

Percussionaire Corporation · Phasitron Kit, VDR, Single Patient · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00849436000259), received between Jun 3, 2016 and Sep 30, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 23, 2024Open, Classified

    Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

    Z-0960-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Operates Differently Than Expected1
  • Infusion or Flow Problem1
  • Material Separation1
  • Mechanical Problem1
  • Particulates1
  • Product Quality Problem1

Patient problems

  • Low Oxygen Saturation1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Pneumothorax1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem; Particulates

    On (b)(6) 2026, Percussionaire was notified that while a ventilated NICU patient was receiving therapy, a piece of blue particulate was identified in the suction port of the patient circuitry. The user ceased use of the breathing circuit on the patient, and replaced it with a different breathing circuit. No patient harm or injury was reported by the customer. As this device malfunction occurred while in use on a ventilated patient, this event is being reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946051 ↗
  • MalfunctionNov 22, 2024Product Quality Problem; Product Quality Problem

    IT WAS REPORTED BY A CUSTOMER THAT DURING TREATMENT ON A PATIENT, THE INTERNAL COMPONENT THAT AIDS IN PROVIDING PULSATION WITHIN THE PHASITRON BREATHING CIRCUIT WAS STICKING, INHIBITING PROPER TREATMENT. THE PATIENT EXPERIENCED SLIGHT DESATURATION. NO ADDITIONAL PATIENT INFORMATION PROVIDED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 20758020 ↗
  • InjuryJan 24, 2024Infusion or Flow Problem; Infusion or Flow Problem

    A CUSTOMER COMPLAINT WAS RECEIVED BY PERCUSSIONAIRE'S DISTRIBUTOR AFTER THE CUSTOMER WAS NOTIFIED OF AN URGENT FIELD SAFETY CORRECTIVE ACTION. IN RESPONSE TO THE PERCUSSIONAIRE FIELD SAFETY NOTIFICATION, THE CUSTOMER STATED ON (B)(6) A NICU PATIENT EXPERIENCED A PNEUMOTHORAX AFTER USING THE PHASITRON DEVICE. THE CUSTOMER NOTED THAT THE DEVICE SEEMED TO BE MAKING A HIGH PITCHED NOISE AND EXHIBITING HIGH PRESSURES, DESPITE ATTEMPTING TO REDUCE THE PRESSURE OUTPUT.

    Patient problems: Pneumothorax

    MAUDE report 18576339 ↗
  • MalfunctionJun 3, 2016Material Separation; Device Operates Differently Than Expected

    PHASITRON SEPARATED AT WELD WHILE IN USE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5699445 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.

  • Z-0693-2024 · Mar 28, 2023 · Open, Classified
    Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
  • Z-0259-2024 · Sep 14, 2020 · Completed
    Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Other Ventilator, Continuous, Facility Use devices