Percussionaire A50094-DRecalls & reported problems
Percussionaire Corporation · Phasitron Kit, VDR, Single Patient · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00849436000259), received between Jun 3, 2016 and Sep 30, 2026 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Dec 23, 2024Open, Classified
Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
Z-0960-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Infusion or Flow Problem1
- Material Separation1
- Mechanical Problem1
- Particulates1
- Product Quality Problem1
Patient problems
- Low Oxygen Saturation1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pneumothorax1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Mechanical Problem; Particulates
On (b)(6) 2026, Percussionaire was notified that while a ventilated NICU patient was receiving therapy, a piece of blue particulate was identified in the suction port of the patient circuitry. The user ceased use of the breathing circuit on the patient, and replaced it with a different breathing circuit. No patient harm or injury was reported by the customer. As this device malfunction occurred while in use on a ventilated patient, this event is being reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946051 ↗MalfunctionNov 22, 2024Product Quality Problem; Product Quality Problem
IT WAS REPORTED BY A CUSTOMER THAT DURING TREATMENT ON A PATIENT, THE INTERNAL COMPONENT THAT AIDS IN PROVIDING PULSATION WITHIN THE PHASITRON BREATHING CIRCUIT WAS STICKING, INHIBITING PROPER TREATMENT. THE PATIENT EXPERIENCED SLIGHT DESATURATION. NO ADDITIONAL PATIENT INFORMATION PROVIDED.
Patient problems: Low Oxygen Saturation
MAUDE report 20758020 ↗InjuryJan 24, 2024Infusion or Flow Problem; Infusion or Flow Problem
A CUSTOMER COMPLAINT WAS RECEIVED BY PERCUSSIONAIRE'S DISTRIBUTOR AFTER THE CUSTOMER WAS NOTIFIED OF AN URGENT FIELD SAFETY CORRECTIVE ACTION. IN RESPONSE TO THE PERCUSSIONAIRE FIELD SAFETY NOTIFICATION, THE CUSTOMER STATED ON (B)(6) A NICU PATIENT EXPERIENCED A PNEUMOTHORAX AFTER USING THE PHASITRON DEVICE. THE CUSTOMER NOTED THAT THE DEVICE SEEMED TO BE MAKING A HIGH PITCHED NOISE AND EXHIBITING HIGH PRESSURES, DESPITE ATTEMPTING TO REDUCE THE PRESSURE OUTPUT.
Patient problems: Pneumothorax
MAUDE report 18576339 ↗MalfunctionJun 3, 2016Material Separation; Device Operates Differently Than Expected
PHASITRON SEPARATED AT WELD WHILE IN USE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5699445 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.
- Z-0693-2024 · Mar 28, 2023 · Open, ClassifiedDistal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
- Z-0259-2024 · Sep 14, 2020 · CompletedMonitron II Waveform Analyzer, REF: S00015-1, used with ventilators.
Other Ventilator, Continuous, Facility Use devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record