Skip to content

Percussionaire F00008-1 (REF S00018-2)Recalls & reported problems

Percussionaire Corporation · VDR4 without Monitron II · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00849436000556), received between Nov 28, 2024 and Nov 28, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Increase in Pressure” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K905235); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Increase in Pressure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2024Increase in Pressure; Increase in Pressure

    PERCUSSIONAIRE WAS NOTIFIED ON NOVEMBER 1 REGARDING A POTENTIAL MALFUNCTION WITH A VDR VENTILATOR. IT WAS REPORTED BY THE CUSTOMER THAT THE VDR WAS MAKING UNUSUAL NOISES WHILE IN USE ON A PATIENT. THE VDR WAS CLOSELY MONITORED AFTER THIS INITIAL OBSERVATION. SHORTLY AFTER, IT WAS NOTICED ON THE VDR MONITOR THAT THE DISPLAYED PRESSURE ON THE OSCILLATORY CPAP WOULD STOP OSCILLATING AND UNEXPECTEDLY INITIATE A MANUAL BREATH AUTOMATICALLY. THE PRESSURE THEN ROSE INTO THE 70S BRIEFLY, AND THEN RETURNED TO THE OSCILLATORY CPAP MODE. THE CUSTOMER COMMUNICATED THAT THERE WAS AN AUDIBLE INCREASE IN FLOW, WHICH SUGGESTED THE PRESSURE WAS DELIVERED TO THE PATIENT. THE DEVICE WAS THEN REMOVED FROM THE PATIENT. THE PATIENT WAS CONFIRMED TO HAVE NO ADVERSE REACTIONS DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20805839 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K905235. One submission often covers many models, so check the product and lot details in each record.

  • Z-0960-2025 · Dec 23, 2024 · Open, Classified
    VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR
  • Z-0693-2024 · Mar 28, 2023 · Open, Classified
    Distal Phasitron, Part: S20020, for use with Percussionaire VDR-4 control driver (or control unit)
  • Z-0259-2024 · Sep 14, 2020 · Completed
    Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.

Other Ventilator, Continuous, Facility Use devices

All Ventilator, Continuous, Facility Use devices →