Skip to content

EMPOWER™ INTRODUCER SHEATH WITH HYDROPHILIC COATING 668782-100Recalls & reported problems

Freudenberg Medical, LLC · The EMPOWER™ Introducer Sheath with Hydrophilic Coating consists of a single-use disposable introducer sheath and dilator that is inserted into the peripheral vasculature to support the introduction of diagnostic or interventional devices, while minimizing blood loss associated with such insertions. The EMPOWER™ Introducer Sheath is coil-reinforced and coated with a lubricious hydrophilic coating. The EMPOWER™ hemostasis valve is integrated into the sheath hub and features a variable aperture seal for enhanced control over device insertion while maintaining hemostasis. The valve is normally closed until the user squeezes the actuator levers to proportionally open the valve and receive any device up to its labeled size. A sideport extends from the EMPOWER™ valve housing with a 3-way stopcock. The radiopaque dilator is designed to accommodate an .035” (0.89mm) guidewire and includes a proximal hub grip with a female luer taper fitting for flushing. Prior to insertion, the dilator hub can be secured to the EMPOWER™ housing with a quarter-turn twist. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K190628); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 21, 2022Class IITerminated

    Non-intended glue from the hub/shaft bond on the inside ID of the catheter shaft may cause Embolism leading to peripheral ischemia

    Z-1016-2022 · Root cause: Process control · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190628. One submission often covers many models, so check the product and lot details in each record.

  • Z-1017-2022 · Feb 21, 2022 · Terminated
    Empower Introducer Sheath with Hydrophilic Coating 21F x 56cm Model: 668782-200 (OUS)

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates