TourGuide™ Steerable Sheath TG0655509Recalls & reported problems

Oscor Inc. · Steerable Sheath, 6.5F · Class 2

All models of TourGuide Steerable Sheath (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00850232006031), received between Jan 20, 2023 and Sep 7, 2023 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak4
  • Detachment of Device or Device Component3
  • Mechanical Problem2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 7, 2023Mechanical Problem; Mechanical Problem

    THE CUSTOMER REPORTED THE DILATOR DOES NOT "CLICK' SHUT INTO THE TOURGUIDE. THUS, IT DOES NOT STAY IN PLACE WHILE ADVANCING INTO THE PATIENT. THE EVENT OCCURRED DURING THE PROCEDURE AND THE PATIENT RECEIVED TREATMENT. THE CUSTOMER STATED THE ISSUE HAS NOT RESOLVED. THE HEALTHCARE PROFESSIONAL STATED THE CAUSE OF THE EVENT WAS ERROR OF GUIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17699702 ↗
  • MalfunctionMar 23, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE EVENT OCCURRED DURING PRE-OP, THERE WAS NO PATIENT INVOLVED IN THIS EVENT. DEVICE LEAKED FROM THE VALVE, AT THE BEGINNING OF THE HOSE ON THE SIDE WHERE THE TAP IS LOCATED AND AT THE TURNING MECHANISM. THE ISSUE WAS RESOLVED USING A NEW PRODUCT. THE PATIENT DID NOT RECEIVE TREATMENT. THE DEVICE IS EXPECTED TO BE RETURNED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16600008 ↗
  • MalfunctionMar 8, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    IT WAS REPORTED DURING A PROCEDURE FOR SMA STENTING, THE SIDEARM OF THIS 6.5F TOUR GUIDE SHEATH FELL OFF AND STARTED LEAKING BLOOD (APPROXIMATELY 100MLS). NO BLOOD TRANSFUSION WAS REQUIRED. IN ORDER TO CONTROL THE BLOOD LOSS, THE PHYSICIAN PLACED FINGER OVER HOLE IN SHEATH, WHILE AN EXCHANGE FOR A NEW 6.5F TOURGUIDE SHEATH (SAME MODEL) WAS COMPLETED. THE PROCEDURE WAS DELAYED FOR 10 MINUTES, TO FIND AND EXCHANGE A NEW SHEATH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE CAUSE OF THE EVENT IS UNKNOWN. THERE WERE NO REPORTED PATIENT SYMPTOMS, OR COMPLICATIONS ASSOCIATED WITH THIS EVENT (ALIVE, NO INJURY).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16503372 ↗
  • MalfunctionMar 6, 2023Fluid/Blood Leak; Detachment of Device or Device Component; Fluid/Blood Leak; Detachment of Device or Device Component

    HEMOSTATIC VALVE FELL OUT AND BLOOD LEAKED ONTO THE TABLE. USED NEW DEVICE TO FINISH PROCEDURE AS NORMAL. THE PATIENT RECEIVED TREATMENT. NO INJURY TO PATIENT. ADDITIONAL PATIENT INFORMATION RECEIVED: PATIENT RECEIVED COELIAC ARTERY ANEURYSM STENT GRAFTING. BLOOD LOST REPORTED AS 100MLS - ISH. 5F STANDARD SHEATH INTRODUCED THROUGH THE END OF THE DEVICE (WHERE THE HEMOSTATIC VALVE WAS) TO STOP THE BLEEDING AND ALLOW A SUPPORTIVE WIRE TO BE INTRODUCED TO FACILITATE EXCHANGE. SAME TYPE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 10 MINS DELAY FINDING A NEW SHEATH - PUT IN SUPPORTING GUIDEWIRE AND EXCHANGE. NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16488615 ↗
  • MalfunctionFeb 14, 2023Mechanical Problem; Mechanical Problem

    THE DILATOR DOES NOT CONNECT TO THE GUIDE. THERE IS NOTHING THAT IT CAN STAY CONNECTED TO. THE DILATOR IS COMPLETELY LOOSE AND FREE RIDING AND IMMEDIATELY UPON TOUCHING ANYTHING AT THE END IT WILL SLIDE INSIDE OF THE GUIDE. THE PATIENT HAS NOT AND WILL NOT RECEIVE TREATMENT. THE HANDLE IS OUTSIDE OF THE PATIENT AND THE COMPLAINT IS ABOUT THE DILATOR-TO-GUIDE CONNECTION OUTSIDE OF THE PATIENT´S BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16375486 ↗
  • MalfunctionJan 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE PHYSICIANS WERE USING A 6.5F TOURGUIDE IN A LEAD INTERVENTIONAL RADIOLOGY (IR) CASE AND REPORTED THE SHEATH WAS LEAKING AT THE BACKEND WIRE PORT. ANOTHER 6.5F TOURGUIDE FROM SAME LOT NUMBER, WAS OPENED AND THE GREEN CAP FELL OFF. PATIENT RECEIVED TREATMENT. REPORTED EVENT WAS RESOLVED AS CASE WAS COMPLETED. IT WAS REPORTED NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. REFERENCE REPORT # 1035166-2023-00006.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16225639 ↗
  • MalfunctionJan 23, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE PHYSICIANS WERE USING A 6.5F TOURGUIDE IN A LEAD INTERVENTIONAL RADIOLOGY (IR) CASE AND REPORTED THE SHEATH WAS LEAKING AT THE BACKEND WIRE PORT. ANOTHER 6.5F TOURGUIDE FROM SAME LOT NUMBER, WAS OPENED AND THE GREEN CAP FELL OFF. PATIENT RECEIVED TREATMENT. REPORTED EVENT WAS RESOLVED AS CASE WAS COMPLETED. IT WAS REPORTED NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. REFERENCE REPORT#: 1035166-2023-00007.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16225559 ↗
  • MalfunctionJan 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING PROCEDURE THE SIDE ARM FELL OFF AND NEW DEVICE WAS REQUIRED. THE PATIENT RECEIVED TREATMENT. PROCEDURE CONTINUED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16219111 ↗

FDA clearance and classification

Cleared through 510(k) K120459 on Jun 29, 2012 (decision: Substantially Equivalent). Submitted by Oscor, Inc. as “Steerable Guiding Sheath, Model Adelante Destino”. The FDA review took 135 days. FDA record ↗

The FDA regulates product code DYB (introducer, catheter) under 21 CFR 870.1340 as a Class II device. 21 CFR 870.1340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TourGuide™ Steerable Sheath TG0655509

Has the TourGuide™ Steerable Sheath TG0655509 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00850232006031), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TourGuide™ Steerable Sheath TG0655509?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jan 20, 2023 and Sep 7, 2023. In those reports, the most common specific device problems are “Fluid/Blood Leak” (4), “Detachment of Device or Device Component” (3) and “Mechanical Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TourGuide™ Steerable Sheath TG0655509?

None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TourGuide™ Steerable Sheath TG0655509 FDA approved?

It was cleared through 510(k) K120459 on Jun 29, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the TourGuide™ Steerable Sheath TG0655509, and is it still sold?

The FDA's device database lists Oscor Inc. as the company responsible for it (brand TourGuide™ Steerable Sheath, model TG0655509). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page