iFuse Implant System 7060M-100Recalls & reported problems
Si-Bone Inc. · Class 2 · Implant
All models of iFuse Implant System (28)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00851085007534), received between Jul 19, 2019 and Nov 2, 2021 — classified as 5 injuries.
The most frequently reported specific device problem is “Malposition of Device” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Malposition of Device5
Patient problems
- Undesired Nerve Stimulation5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 2, 2021Malposition of Device; Malposition of Device
THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2021 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF RIGHT SIDE RADICULAR PAIN FOLLOWING THE INITIAL PROCEDURE. THE SURGEON DETERMINED THAT THE RIGHT SIDE CRANIAL POSITIONED IMPLANT HAD BREACHED THE NEUROFORAMEN. IN (B)(6) 2021, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE CRANIAL POSITIONED IMPLANT AND REPLACED IT WITH A SHORTER IMPLANT OF THE SAME TYPE. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE STATUS OF THE PATIENT FOLLOWING THE REVISION PROCEDURE IS NOT KNOWN.
Patient problems: Undesired Nerve Stimulation
MAUDE report 12738736 ↗InjuryMay 7, 2020Malposition of Device; Malposition of Device
THE PATIENT HAD LEFT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2020 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF LEFT SIDE RADICULAR PAIN AFTER THE INITIAL PROCEDURE. THE SURGEON DETERMINED THAT THE SUPERIOR POSITIONED IMPLANT WAS IMPINGING ON THE NEUROFORAMEN. IN (B)(6) 2020, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE IMPLANT. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. NO OTHER HARDWARE OR BONE GRAFT WAS ADDED. THE PATIENT'S RADICULAR PAIN SYMPTOMS RESOLVED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 10037062 ↗InjuryFeb 26, 2020Malposition of Device; Malposition of Device
THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2020 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF RIGHT SIDE RADICULAR PAIN FOLLOWING THE PROCEDURE. THE SURGEON DETERMINED THAT THE SUPERIOR POSITIONED IMPLANT HAD BREACHED THE NEUROFORAMEN. TWO DAYS AFTER THE INITIAL PROCEDURE, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE IMPLANT AND REPLACED IT WITH A SHORTER IMPLANT OF THE SAME TYPE TO RELIEVE THE NERVE IMPINGEMENT. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE PATIENT'S RADICULAR PAIN SYMPTOMS RESOLVED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 9753704 ↗InjuryDec 23, 2019Malposition of Device; Malposition of Device
THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2019 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF RIGHT SIDE RADICULAR PAIN FOLLOWING THE PROCEDURE. THE SURGEON DETERMINED THAT THE SUPERIOR POSITIONED IMPLANT HAD BREACHED THE NEUROFORAMEN. THE DAY AFTER THE INITIAL PROCEDURE, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE REMOVED THE IMPLANT. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE PATIENT'S RADICULAR PAIN SYMPTOMS RESOLVED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 9510900 ↗InjuryJul 19, 2019Malposition of Device; Malposition of Device
THE PATIENT HAD RIGHT SIDE SI JOINT ARTHRODESIS IN (B)(6) 2019 WHERE THREE IMPLANTS WERE INSTALLED. THE PATIENT COMPLAINED OF RADICULAR PAIN FOLLOWING THE INITIAL PROCEDURE. THE SURGEON DETERMINED THAT THE MIDDLE-POSITIONED IMPLANT HAD BREACHED THE NEUROFORAMEN AND WAS IMPINGING ON THE NERVE ROOT. THE DAY AFTER THE INITIAL PROCEDURE, THE SURGEON PERFORMED A REVISION PROCEDURE WHERE HE ADDED AN ADDITIONAL IMPLANT OF THE SAME TYPE IN A MORE DORSAL POSITION AND THEN REMOVED THE MIDDLE-POSITIONED IMPLANT. NO OTHER PREEXISTING IMPLANTS WERE ADJUSTED OR REMOVED. THE PATIENT'S RADICULAR PAIN SYMPTOMS RESOLVED FOLLOWING THE REVISION PROCEDURE.
Patient problems: Undesired Nerve Stimulation
MAUDE report 8809662 ↗
Other Sacroiliac Joint Fixation devices
- Firebird SI Fusion 18-1035SPOrthofix US LLC · Sacroiliac Joint Fixation1 recall1 report
- iFuse TORQ Implant System 10065TSi-Bone Inc. · Sacroiliac Joint Fixation1 recall1 report
- iFuse TORQ Implant System 10060TSi-Bone Inc. · Sacroiliac Joint Fixation1 recall1 report
- Firebird SI Fusion 18-2050SPOrthofix US LLC · Sacroiliac Joint Fixation1 recall0 reports
- Firebird SI Fusion 18-1065SPOrthofix US LLC · Sacroiliac Joint Fixation1 recall0 reports
- Firebird SI Fusion 18-2065SPOrthofix US LLC · Sacroiliac Joint Fixation1 recall0 reports
- Firebird SI Fusion 18-2045SPOrthofix US LLC · Sacroiliac Joint Fixation1 recall0 reports
- iFuse TORQ Implant System 10055TSi-Bone Inc. · Sacroiliac Joint Fixation1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page