Skip to content

T15 Hexalobe Drive 201.047.15Recalls & reported problems

Jace Medical, LLC · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00851599007655), received between Mar 11, 2024 and Dec 30, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Insufficient Device Problem Information1
  • Material Twisted/Bent1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2025Material Twisted/Bent; Insufficient Device Problem Information; Material Twisted/Bent; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING RESTOCK WITHIN THE STERILIZATION DEPARTMENT AT THE HOSPITAL, THE DEVICE LOOKED MALFUNCTIONED AND POSSIBLY WARPED. THERE WAS NO PATIENT INVOLVEMENT. HOWEVER, ONCE PRODUCT WAS RETURNED IT WAS NOTED THAT THE TIP WAS TWISTED AND CRACKED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23923431 ↗
  • MalfunctionNov 14, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT A DRIVER WAS WORN BUT WAS DISCOVERED TO BE FRACTURED POST INVESTIGATION. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23554292 ↗
  • MalfunctionMar 18, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE BLADE FRACTURED WITH THE TIP BEING FRACTURED OFF. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633038 ↗
  • MalfunctionMar 18, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE BLADE FRACTURED WITH THE TIP BEING FRACTURED OFF. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21633028 ↗
  • InjuryMar 11, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A CABG PROCEDURE, A TORQUE LIMITER WAS NOT PLACED ON THE DRILL AND THE TIP OF THE SCREW BLADE BROKE OFF AND WAS LATER DISCOVERED TO BE LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED AS PLANNED AND IT WAS NOT NOTED UNTIL DAYS LATER THAT THE BLADE HAD FRACTURED. THERE WAS NO MEDICAL INTERVENTION AND IT WAS DEEMED THAT THE RISK OF REMOVING THE BROKEN PIECE WAS A HIGHER RISK THAN TO LEAVE IT INSIDE THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18878122 ↗

Other Driver, Prosthesis devices

All Driver, Prosthesis devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page