Skip to content

Titan SGS Standard Gastric Stapler SGS23RbRecalls & reported problems

Standard Bariatrics, Inc · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00851677007133), received between May 5, 2023 and Sep 16, 2025 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Improper or Incorrect Procedure or Method2
  • Material Integrity Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Fire1

Patient problems

  • Insufficient Information2
  • Exsanguination1
  • Failure to Anastomose1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED "AFTER THE DISSECTED STOMACH WAS REMOVED, IT WAS INFLATED AND A LEAK ALONG THE STAPLE LINE WAS NOTICED. ONLY THE EXTRACTED PORTION OF THE STOMACH HAD THIS ISSUE. IT OCCURRED ROUGHLY AT THE 50% MARK OF THE STAPLER". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Insufficient Information

    MAUDE report 23062255 ↗
  • MalfunctionSep 15, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED "AFTER THE DISSECTED STOMACH WAS REMOVED, IT WAS INFLATED AND A LEAK ALONG THE STAPLE LINE WAS NOTICED. ONLY THE EXTRACTED PORTION OF THE STOMACH HAD THIS ISSUE. IT OCCURRED ROUGHLY AT THE 50% MARK OF THE STAPLER". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Insufficient Information

    MAUDE report 23057680 ↗
  • InjuryAug 7, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED "PATIENT HAD A SIGNIFICANT HEMOGLOBIN LOSS (6 POINTS OVERALL) BUT IS STABLE RIGHT NOW. WE DIDN'T WITNESS ANY EVIDENT PRODUCT MALFUNCTION OR MISUSE DURING THE TRIAL".

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 22738775 ↗
  • InjuryJul 29, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PATIENT "SUFFERED SIGNIFICANT BLEEDING FROM THE SUTURE LINE" AND "WENT INTO HEMORRHAGIC SHOCK DURING THE NIGHT, FORCING THE SURGEONS TO PERFORM AN EMERGENCY PROCEDURE TO PUT A STOP TO THE BLEEDING. WE DIDN'T WITNESS ANY EVIDENT PRODUCT MALFUNCTION OR MISUSE DURING THE TRIAL".

    Patient problems: Exsanguination

    MAUDE report 22648104 ↗
  • MalfunctionJul 17, 2025Failure to Fire; Failure to Fire

    IT IS REPORTED AS "TITIAN SGS STAPLER WOULD NOT GO INTO FIRING MODE. AFTER TIME OUT WAS ACKNOWLEDGED AND GREEN BUTTON WAS PUSHED AGAIN IT WOULD GO INTO TIME OUT MODE AGAIN. WE HAD TO OPEN THE STAPLER TO REMOVE THE STOMACH. HOWEVER, AFTER THE STOMACH WAS REMOVED THE DEVICE WOULD NOT CLOSE. AMANUAL BAILOUT CLOSE WAS PERFORMED TO GET THE JAWS TO CLOSE FOR REMOVAL OF THE STAPLER.NO PATIENT INJURY OR CONSEQUENCE REPORTED. CURRENT CONDITION OF THE PATIENT REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22534968 ↗
  • InjuryMay 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT DEVELOPED A LEAK ON (B)(6) 2023 FOLLOWING A SLEEVE GASTRECTOMY PROCEDURE COMPLETED ON (B)(6) 2023, WHICH HAD USED A TITAN SGS23RB STAPLER. A SMALL DEFECT WAS OBSERVED MORE POSTERIOR OF THE FUNDUS. IT WAS REPORTED THAT IT LOOKED LIKE THE FUNDUS WAS POSSIBLY FOLDED. RE-OPERATION/SURGICAL INTERVENTION WAS REQUIRED TO RESOLVE THE LEAK. THE PATIENT WAS RELEASED AFTER RE-OPERATION/SURGICAL INTERVENTION. STANDARD BARIATRICS WAS INFORMED OF THE LEAK ON (B)(6) 2023.

    Patient problems: Failure to Anastomose

    MAUDE report 16882147 ↗

Other Stapler, Surgical devices

All Stapler, Surgical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page