Titan SGS Standard Gastric Stapler SGS23RbRecalls & reported problems
Standard Bariatrics, Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00851677007133), received between May 5, 2023 and Sep 16, 2025 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method2
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Fire1
Patient problems
- Insufficient Information2
- Exsanguination1
- Failure to Anastomose1
- Hemorrhage/Blood Loss/Bleeding1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED "AFTER THE DISSECTED STOMACH WAS REMOVED, IT WAS INFLATED AND A LEAK ALONG THE STAPLE LINE WAS NOTICED. ONLY THE EXTRACTED PORTION OF THE STOMACH HAD THIS ISSUE. IT OCCURRED ROUGHLY AT THE 50% MARK OF THE STAPLER". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 23062255 ↗MalfunctionSep 15, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED "AFTER THE DISSECTED STOMACH WAS REMOVED, IT WAS INFLATED AND A LEAK ALONG THE STAPLE LINE WAS NOTICED. ONLY THE EXTRACTED PORTION OF THE STOMACH HAD THIS ISSUE. IT OCCURRED ROUGHLY AT THE 50% MARK OF THE STAPLER". THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 23057680 ↗InjuryAug 7, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED "PATIENT HAD A SIGNIFICANT HEMOGLOBIN LOSS (6 POINTS OVERALL) BUT IS STABLE RIGHT NOW. WE DIDN'T WITNESS ANY EVIDENT PRODUCT MALFUNCTION OR MISUSE DURING THE TRIAL".
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 22738775 ↗InjuryJul 29, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PATIENT "SUFFERED SIGNIFICANT BLEEDING FROM THE SUTURE LINE" AND "WENT INTO HEMORRHAGIC SHOCK DURING THE NIGHT, FORCING THE SURGEONS TO PERFORM AN EMERGENCY PROCEDURE TO PUT A STOP TO THE BLEEDING. WE DIDN'T WITNESS ANY EVIDENT PRODUCT MALFUNCTION OR MISUSE DURING THE TRIAL".
Patient problems: Exsanguination
MAUDE report 22648104 ↗MalfunctionJul 17, 2025Failure to Fire; Failure to Fire
IT IS REPORTED AS "TITIAN SGS STAPLER WOULD NOT GO INTO FIRING MODE. AFTER TIME OUT WAS ACKNOWLEDGED AND GREEN BUTTON WAS PUSHED AGAIN IT WOULD GO INTO TIME OUT MODE AGAIN. WE HAD TO OPEN THE STAPLER TO REMOVE THE STOMACH. HOWEVER, AFTER THE STOMACH WAS REMOVED THE DEVICE WOULD NOT CLOSE. AMANUAL BAILOUT CLOSE WAS PERFORMED TO GET THE JAWS TO CLOSE FOR REMOVAL OF THE STAPLER.NO PATIENT INJURY OR CONSEQUENCE REPORTED. CURRENT CONDITION OF THE PATIENT REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22534968 ↗InjuryMay 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT DEVELOPED A LEAK ON (B)(6) 2023 FOLLOWING A SLEEVE GASTRECTOMY PROCEDURE COMPLETED ON (B)(6) 2023, WHICH HAD USED A TITAN SGS23RB STAPLER. A SMALL DEFECT WAS OBSERVED MORE POSTERIOR OF THE FUNDUS. IT WAS REPORTED THAT IT LOOKED LIKE THE FUNDUS WAS POSSIBLY FOLDED. RE-OPERATION/SURGICAL INTERVENTION WAS REQUIRED TO RESOLVE THE LEAK. THE PATIENT WAS RELEASED AFTER RE-OPERATION/SURGICAL INTERVENTION. STANDARD BARIATRICS WAS INFORMED OF THE LEAK ON (B)(6) 2023.
Patient problems: Failure to Anastomose
MAUDE report 16882147 ↗
Other Stapler, Surgical devices
- ECHELON ECH60SEthicon Endo-Surgery, LLC · Stapler, surgical1 death631 reports
- ECHELON ECH60LEthicon Endo-Surgery, LLC · Stapler, surgical467 reports
- ECHELON ECH45SEthicon Endo-Surgery, LLC · Stapler, surgical269 reports
- EEA TRIEEA28XTCovidien LP · Stapler, surgical220 reports
- ECHELON ECH60CEthicon Endo-Surgery, LLC · Stapler, surgical197 reports
- ECHELON ECH45CEthicon Endo-Surgery, LLC · Stapler, surgical138 reports
- EEA TRIEEA31XTCovidien LP · Stapler, surgical73 reports
- ETHICON EC3D60LEthicon Endo-Surgery, LLC · Stapler, surgical39 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.