t:slim G4 Insulin Pump, Reconditioned Replacement 008114Recalls & reported problems
Tandem Diabetes Care, Inc. · Insulin Pump, Reconditioned Replacement, t:slim G4 · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00852162004613), received between Dec 23, 2025 and Aug 6, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Key or Button Unresponsive/not Working” (2 reports).
1 other recall cover devices cleared under the same FDA submission (P140015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Key or Button Unresponsive/not Working2
- Display or Visual Feedback Problem1
- Failure to Charge1
- Failure to Power Up1
- Mechanical Problem1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Hyperglycemia1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Key or Button Unresponsive/not Working
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50160330 ↗MalfunctionJul 2, 2026Failure to Power Up
IT WAS REPORTED THAT THE PUMP BATTERY WAS UNRESPONSIVE AND "WOULD NOT START" AFTER A SHUT DOWN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25692471 ↗MalfunctionMar 11, 2026Failure to Charge
THE CUSTOMER REPORTED THAT THEIR OLD G4 PUMP WOULD NOT TURN ON AFTER PLUGGING IN THE CHARGING CABLE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT ADVISED THE CUSTOMER TO KEEP THE PUMP CHARGING WITHOUT TOUCHING THE BUTTON, WHICH RESULTED IN THE PUMP TURNING ON. THE CUSTOMER PLANS TO USE THE G4 PUMP AS A BACKUP UNTIL A REPLACEMENT PUMP IS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24570836 ↗MalfunctionJan 28, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH TURNING ON A LEGACY PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE, INSTEAD DECIDING TO WAIT TO SPEAK WITH A HEALTHCARE PROFESSIONAL, AS THEY WERE IN THE PROCESS OF ACQUIRING ANOTHER PUMP. TECHNICAL SUPPORT THEN CLOSED THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199773 ↗InjuryJan 15, 2026Obstruction of Flow
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS DISPLAYED AS "HIGH"; HOWEVER, SPECIFIC VALUE WAS NOT PROVIDED. ELEVATED BG WAS ADDRESSED BY A CORRECTION BOLUS VIA THE PUMP. CUSTOMER CHANGED THE CARTRIDGE ONLY AND RESUMED INSULIN.
Patient problems: Hyperglycemia
MAUDE report 24089478 ↗MalfunctionJan 8, 2026Mechanical Problem
A CARTRIDGE CHANGE ERROR WAS REPORTED BY THE CUSTOMER WITH THEIR PUMP. THERE WAS NO ADVERSE IMPACT ON THE CUSTOMER'S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT INSTRUCTED THE CUSTOMER TO REMOVE THE CARTRIDGE, INSPECT THE O-RING, AND CLEAN THE PUMP AREA. DESPITE THESE STEPS AND ASSISTANCE FROM TECHNICAL SUPPORT, THE CUSTOMER WAS UNABLE TO SUCCESSFULLY LOAD A NEW CARTRIDGE. THE ISSUE WAS ESCALATED TO ADDITIONAL SUPPORT, AND IT WAS DETERMINED THAT THE CUSTOMER NEEDED A LOANER PUMP. CUSTOMER SERVICE PROVIDED GUIDANCE ON SIGNING THE LOANER AGREEMENT AND CONFIRMED THE RETURN DATE FOR THE TEMPORARY PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24003103 ↗MalfunctionDec 23, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23879211 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140015. One submission often covers many models, so check the product and lot details in each record.
- Z-2471-2018 · Apr 23, 2018 · Open, Classifiedt:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Other Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor devices
- t:slim X2 5.0 Insulin Pump 1000885Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall2 deaths17,224 reports
- t:slim X2 5.0 Insulin Pump 1000911Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls2 deaths15,928 reports
- t:slim X2 5.0 Insulin Pump 1000897Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall2 deaths6,444 reports
- t:slim X2 5.x Insulin Pump 1003808Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls1 death2,526 reports
- t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM 006194Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor4 deaths2,272 reports
- Animas Vibe 11451100Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor2 deaths2,262 reports
- t:slim X2 5.0 Insulin Pump 1000891Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall1,561 reports
- Animas Vibe 11451000Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor1,122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page