t:slim X2 5.0 Insulin Pump 1000891Recalls & reported problems

Tandem Diabetes Care, Inc. · t:slim X2 5.0 Insulin Pump, Replacement · Class 3

All models of t:slim X2 Insulin Pump (45) · t:slim X2 5.0 Insulin Pump (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem t:slim X2 is an insulin pump. Its 1,561 reports describe many cases of the pump stopping, charging and battery problems, and wrong measurements, and some describe blocked flow. The reports include 123 injuries and no deaths. There is 1 recall, still open, which started in May 2022 and concerns software problems that could lead to blood sugar running high or low.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,561
Deaths reported
0
Injuries reported
123
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1,561 adverse event reports that list this exact device (DI 00853052007271), received between Jul 9, 2018 and Aug 7, 2026 — classified as 123 injuries, 1,438 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (287 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P140015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 3 reportsJan 2024: 6 reportsFeb 2024: 4 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 1 reportsJun 2024: 5 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 24, 2022Class IIOpen, Classified

    Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

    Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped287
  • Charging Problem220
  • Battery Problem211
  • Incorrect Measurement211
  • Obstruction of Flow186
  • Failure to Sense101
  • Adverse Event Without Identified Device or Use Problem77
  • False Alarm76

Patient problems

  • No Consequences Or Impact To Patient664
  • No Clinical Signs, Symptoms or Conditions622
  • No Known Impact Or Consequence To Patient158
  • Hyperglycemia60
  • Hypoglycemia51
  • Elevated Ketones/ Diabetic Ketoacidosis15
  • Insufficient Information4
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 1,561 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177473 ↗
  • MalfunctionJul 6, 2026Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716999 ↗
  • MalfunctionJun 4, 2026Mechanical Problem

    IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING THE LOAD SEQUENCE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25399276 ↗
  • MalfunctionApr 9, 2026Premature Discharge of Battery

    IT WAS REPORTED THAT CUSTOMER EXPERIENCED BATTERY DEPLETION ISSUE WITH PUMP. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED FURTHER FOLLOW-UP AND NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE OUTCOME OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24837599 ↗
  • MalfunctionNov 5, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE CUSTOMER WAS EXPERIENCING UNSPECIFIED ISSUES WITH THE RESPONSIVENESS OF THE WAKE BUTTON ON THE PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476760 ↗
  • MalfunctionNov 5, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23473051 ↗
  • MalfunctionAug 22, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT CARTRIDGE CHANGE ERRORS OCCURRED WITH MULTIPLE CARTRIDGES. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY THE CUSTOMER CONTINUED TO USE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22876879 ↗
  • MalfunctionAug 22, 2025Mechanical Problem; Use of Device Problem; Mechanical Problem; Use of Device Problem

    PATIENT REPORTED EXPERIENCING MULTIPLE CARTRIDGE LOAD ERRORS WITH THEIR CURRENT DEVICE, CAUSING THEM TO GO THROUGH SEVERAL CARTRIDGES BEFORE FINDING ONE THAT WORKS. THERE WAS NO ADVERSE IMPACT TO THE PATIENT¿S BLOOD GLUCOSE LEVEL. THE PATIENT DECLINED FURTHER ASSISTANCE FROM TECHNICAL SUPPORT AS THEY ARE IN THE PROCESS OF RENEWING THEIR DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22873050 ↗
  • MalfunctionJul 13, 2025Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22490634 ↗
  • MalfunctionJul 10, 2025Mechanical Problem; Mechanical Problem

    A CARTRIDGE CHANGE ERROR WAS REPORTED WITH THE PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL, AS IT DID NOT EXCEED 500 MG/DL (27.7 MMOL/L). THE CUSTOMER FOLLOWED INSTRUCTIONS FROM TECHNICAL SUPPORT TO REMOVE THE CARTRIDGE AND INSPECT THE O-RING, AND THEN CLEANED THE PNEUMATIC TAP AREA OF THE PUMP. DESPITE THESE EFFORTS, A MALFUNCTION OCCURRED, AND THE PUMP COULD NOT BE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22468601 ↗
  • MalfunctionJun 7, 2025Failure to Sense; Failure to Sense

    IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. THERE WAS NO ADVERSE IMPACT TO CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22163692 ↗
  • MalfunctionApr 29, 2025Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21928270 ↗
  • MalfunctionMar 17, 2025Incorrect Measurement; Battery Problem; Incorrect Measurement; Battery Problem

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21618230 ↗
  • MalfunctionMar 7, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CONTINUOUS GLUCOSE MONITOR ERROR 42. THE PUMP WAS RESET, AND THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21552037 ↗
  • MalfunctionJan 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A CARTRIDGE DID NOT FIT ONTO THE PUMP. REPORTEDLY, THERE WAS AN O-RING FROM A PREVIOUS CARTRIDGE ON THE PNEUMATIC TAP. THE CUSTOMER REMOVED THE O-RING AND SUCCESSFULLY LOADED THE CARTRIDGE TO ADDRESS THE EVENT. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21134897 ↗
  • MalfunctionNov 19, 2024Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20719891 ↗
  • MalfunctionNov 11, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20652353 ↗
  • MalfunctionNov 7, 2024Incorrect Measurement; Pumping Stopped; Incorrect Measurement; Pumping Stopped

    IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING RESULTING IN THE PUMP SHUTTING DOWN. ADDITIONALLY, THE INSULIN GAUGE WAS INACCURATE AND STATIC. THERE WAS NO ADVERSE IMPACT TO CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20629413 ↗
  • MalfunctionOct 18, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE CUSTOMER HAD AN UNSPECIFIED CARTRIDGE ISSUE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20482044 ↗
  • MalfunctionOct 15, 2024Battery Problem; Charging Problem; Battery Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. ADDITIONALLY, THE PUMP BATTERY WAS REPORTED CHARGING INCONSISTENTLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUES. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20449954 ↗
  • MalfunctionOct 7, 2024Failure to Sense; Battery Problem; Charging Problem; Failure to Sense; Battery Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY AND WAS TAKING LONGER TO CHARGE. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP USB PORT WAS DAMAGED RESULTING IN DIFFICULTIES CHARGING THE PUMP. IT WAS ALSO REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20392697 ↗
  • MalfunctionSep 13, 2024Charging Problem; Charging Problem

    IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED, POSSIBLY DUE TO "BEING WASHED.". THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20215565 ↗
  • MalfunctionAug 28, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20086420 ↗
  • MalfunctionAug 27, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT AN UNSPECIFIED MALFUNCTION ALARM OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20077181 ↗
  • MalfunctionJun 18, 2024Pumping Stopped; Pumping Stopped

    IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19558393 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140015. One submission often covers many models, so check the product and lot details in each record.

  • Z-2471-2018 · Apr 23, 2018 · Open, Classified
    t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.

Other Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor devices

All Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page