t:slim X2 5.0 Insulin Pump 1000897Recalls & reported problems
Tandem Diabetes Care, Inc. · t:slim X2 5.0 Insulin Pump, Reconditioned · Class 3
All models of t:slim X2 Insulin Pump (45) · t:slim X2 5.0 Insulin Pump (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Tandem t:slim X2 insulin pump is used by people with diabetes. Of its 6,444 reports, 2 describe deaths and 857 describe injuries; many describe charging problems, pumping that stopped, inaccurate readings and blocked flow. Some mention high or low blood sugar. There is 1 open recall from May 2022, covering software issues such as memory faults and an inaccurate battery display, which could result in low blood sugar, high blood sugar or diabetic ketoacidosis.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 6,444
- Deaths reported
- 2
- Injuries reported
- 857
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6,444 adverse event reports that list this exact device (DI 00853052007295), received between Jul 9, 2018 and Jun 8, 2026 — classified as 2 deaths, 857 injuries, 5,585 malfunctions.
The most frequently reported specific device problem is “Charging Problem” (1,064 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P140015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- May 24, 2022Class IIOpen, Classified
Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
Z-1811-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Charging Problem1,064
- Pumping Stopped943
- Incorrect Measurement927
- Obstruction of Flow614
- Adverse Event Without Identified Device or Use Problem609
- Filling Problem379
- Battery Problem373
- Physical Resistance/Sticking315
Patient problems
- No Consequences Or Impact To Patient2,771
- No Clinical Signs, Symptoms or Conditions2,398
- No Known Impact Or Consequence To Patient452
- Hyperglycemia438
- Hypoglycemia384
- Elevated Ketones/ Diabetic Ketoacidosis87
- Loss of consciousness12
- Insufficient Information9
Recent adverse event reports
The 25 most recent of 6,444 reports, in the reporter's own words as published by the FDA.
InjuryJun 8, 2026Use of Device Problem
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ELEVATED BLOOD GLUCOSE (BG) LEVEL DISPLAYED AS "HIGH"; HOWEVER, SPECIFIC VALUE WAS NOT PROVIDED. REPORTEDLY THE CUSTOMER DID NOT DELIVER A MEAL BOLUS. A CORRECTION BOLUS WAS DELIVERED TO ADDRESS BG. RECOMMENDATION WAS MADE TO CONSULT WITH A HEALTHCARE PROVIDER REGARDING DIABETES MANAGEMENT.
Patient problems: Hyperglycemia
MAUDE report 25429512 ↗MalfunctionMar 20, 2026Pumping Stopped
CALLER REPORTED A MALFUNCTION ON THE PUMP, SPECIFICALLY IDENTIFIED AS MALF 0, WITH A FAULT ID OF 9-0X20F1. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT PROVIDED PUMP RESET INFORMATION AND ASSISTED THE CALLER IN RESETTING THE PUMP, AFTER WHICH THE MALFUNCTION CODE DID NOT RECUR. CALLER WAS ADVISED TO CONTINUE USING THE PUMP AND TO CALL BACK IF THE ISSUE ARISES AGAIN, WHICH THEY ACKNOWLEDGED AND UNDERSTOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24648203 ↗MalfunctionMar 9, 2026Data Problem
IT WAS REPORTED THAT THE PUMP INDICATED A DATA LOG CORRUPTION. THERE WAS NO ADVERSE IMPACT ON THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24544124 ↗InjuryMar 4, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CONTROL-IQ ALGORITHM SUSPENDED BASAL DELIVERY IN RESPONSE TO THE CONTINUOUS GLUCOSE MONITOR (CGM) SENSOR READING; HOWEVER, THE CGM SENSOR READING WAS INACCURATE. REPORTEDLY, THE CGM BLOOD GLUCOSE (BG) READING WAS A LEVEL THAT WAS DISPLAYED AS ¿LOW¿; HOWEVER, A SPECIFIC VALUE WAS NOT PROVIDED, AND THE CONTACT WAS UNABLE TO PROVIDE A METER BG READING AT THE TIME OF THE EVENT. AS A RESULT OF THE BASAL SUSPENSION AND CUSTOMER TREATING A FALSE LOW, CUSTOMER'S BLOOD GLUCOSE (BG) ROSE AND EXPERIENCED DIABETIC KETOSIS. HIGHEST BG LEVEL WAS NOT PROVIDED AT TIME OF THE EVENT. DUE TO THE ELEVATED BG LEVEL THE CUSTOMER LOSS CONSCIOUSNESS. CUSTOMER WENT TO THE EMERGENCY ROOM (ER) AND WAS ADMITTED INTO THE INTENSIVE CARE UNIT. CUSTOMER WAS TREATED WITH INTRAVENOUS FLUIDS OF INSULIN AND SALINE AND WAS RELEASED FROM THE HOSPITAL ON (B)(6) 2025. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE ISSUE; HOWEVER, THE CUSTOMER DID NOT RESPOND. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: Hyperglycemia; Loss of consciousness
MAUDE report 24511858 ↗MalfunctionNov 17, 2025Pumping Stopped; Pumping Stopped
A MALFUNCTION WAS REPORTED ON THE PUMP. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL, AS IT DID NOT EXCEED 500 MG/DL. THE CUSTOMER RECEIVED INFORMATION ON HOW TO RESET THE PUMP FROM TECHNICAL SUPPORT, SUCCESSFULLY RESET THE DEVICE, AND CONFIRMED THAT THE MALFUNCTION DID NOT RECUR. THE CUSTOMER WAS ADVISED TO CONTINUE USING THE PUMP AND TO CONTACT TECHNICAL SUPPORT IF THE ISSUE RECURS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23575806 ↗MalfunctionOct 30, 2025Pumping Stopped; Pumping Stopped
A CALLER REPORTED A MALFUNCTION ON THE PUMP, IDENTIFIED BY MALFUNCTION CODE MALF 9, WHICH WAS RESETTABLE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. TECHNICAL SUPPORT PROVIDED INFORMATION ON RESETTING THE PUMP AND ASSISTED THE CALLER WITH THE PROCESS, ENSURING THE MALFUNCTION CODE DID NOT RECUR. THE CUSTOMER WAS ADVISED TO CONTINUE USING THE PUMP AND INSTRUCTED TO CALL BACK IF ANY MALFUNCTION CODE RECURS, WHICH THE CALLER ACKNOWLEDGED AND UNDERSTOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23429604 ↗MalfunctionJul 8, 2025Charging Problem; Material Twisted/Bent; Charging Problem; Material Twisted/Bent
IT WAS REPORTED THAT THE PUMP'S USB PORT WAS BENT RESULTING IN DIFFICULTIES CHARGING THE PUMP. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22440961 ↗MalfunctionJul 8, 2025False Alarm; False Alarm
AN ALTITUDE ALARM WAS REPORTED FOR A PUMP, CAUSING THE CUSTOMER'S INSULIN TO BE SUSPENDED. THE CUSTOMER'S BLOOD GLUCOSE LEVEL DID NOT EXCEED 500 MG/DL, INDICATING THERE WAS NO ADVERSE IMPACT. TECHNICAL SUPPORT ADVISED THE CUSTOMER TO CLEAN THE SPEAKER HOLES AND ATTEMPT CARTRIDGE RECONNECTION, BUT INSULIN DELIVERY COULD NOT BE RESUMED IMMEDIATELY. EVENTUALLY, THE CUSTOMER WAS INSTRUCTED TO LOAD A NEW CARTRIDGE AND RESUME INSULIN DELIVERY, BUT THE ALARM PERSISTED. AN ATTEMPT WAS MADE BY TANDEM TECHNICAL SUPPORT TO FOLLOW-UP WITH THE CUSTOMER ON THE REPORTED ISSUE, BUT NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22437132 ↗MalfunctionJun 30, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE. THE CUSTOMER DECLINED FOLLOW UP WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22366641 ↗MalfunctionJun 26, 2025Failure to Charge; Failure to Charge
IT WAS REPORTED THAT THE PUMP WAS NOT HOLDING A CHARGE WELL, REQUIRING CHARGING EVERY TWO DAYS, THE SLIDE FOR INSERTING THE CARTRIDGE WAS DEGRADING, AND THE SCREEN HAD SCRATCHES. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FURTHER TROUBLESHOOTING AND DID NOT SEEK ADDITIONAL FOLLOW-UP OR ASSISTANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22335524 ↗MalfunctionJun 10, 2025Measurement System Incompatibility; Measurement System Incompatibility
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT WHILE USING THE TANDEM-PROVIDED USB CABLE AND WALL ADAPTER. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING AN ALTERNATE WALL ADAPTER AND USB CABLE. THERE WAS NO ADVERSE IMPACT THE CUSTOMER¿S BLOOD GLUCOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22177415 ↗MalfunctionMay 29, 2025Failure to Sense; Failure to Sense
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS UNRESPONSIVE. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. REPORTEDLY, THE CUSTOMER REVERTED TO MANUAL INJECTIONS FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22108427 ↗InjuryMar 14, 2025Obstruction of Flow; Obstruction of Flow
IT WAS REPORTED OCCLUSION ALARMS OCCURRED. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS DISPLAYED AS "HIGH"; A SPECIFIC VALUE WAS NOT PROVIDED. ELEVATED BLOOD GLUCOSE WAS ADDRESSED WITH A MANUAL DOSE INJECTION. THE CUSTOMER CHANGED SUPPLIES AND RESUMED INSULIN THERAPY.
Patient problems: Hyperglycemia
MAUDE report 21606058 ↗MalfunctionMar 11, 2025Charging Problem; Charging Problem
IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED ADVERSE IMPACT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21573707 ↗MalfunctionJan 24, 2025Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21231005 ↗InjuryJan 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A BLOOD GLUCOSE (BG) LEVEL THAT DISPLAYED AS "HIGH", SPECIFIC VALUE NOT PROVIDED. CAUSE OF BG LEVEL WAS NOT KNOWN. CUSTOMER ADMINISTERED A BOLUS OF INSULIN VIA THE PUMP AND A MANUAL INJECTION TO ADDRESS THE ISSUE AND WENT TO THE EMERGENCY ROOM. CUSTOMER WAS TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND WAS NOT SURE IF THEY WERE GIVEN INSULIN. CUSTOMER WAS DISCHARGED THE SAME DAY WITHOUT THE ISSUE RESOLVED AND NO PERMANENT DAMAGE. TANDEM TECHNICAL SUPPORT DID A FULL PUMP SYSTEM CHECK AND CONFIRMED THAT PUMP WAS FUNCTIONING AS INTENDED.
Patient problems: Hyperglycemia
MAUDE report 21219745 ↗MalfunctionJan 2, 2025Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE WAKE BUTTON WAS LESS RESPONSIVE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21056964 ↗MalfunctionDec 13, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A CARTRIDGE ALARM OCCURRED DURING THE LOAD PHASE. THE CUSTOMER REVERTED TO AN ALTERNATE PUMP FOR INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926164 ↗MalfunctionDec 6, 2024Incorrect Measurement; Pumping Stopped; Incorrect Measurement; Pumping Stopped
IT WAS REPORTED THAT THE BATTERY GAUGE WAS FLUCTUATING. ADDITIONALLY, IT WAS REPORTED THAT THE PUMP INTERMITTENTLY SHUT OFF UNEXPECTEDLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20864616 ↗MalfunctionNov 18, 2024Pumping Stopped; Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20707103 ↗MalfunctionNov 13, 2024Date/Time-Related Software Problem; Date/Time-Related Software Problem
IT WAS REPORTED THAT THE PUMP TIME WAS INCORRECT, CAUSE WAS UNKNOWN. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20676197 ↗MalfunctionNov 6, 2024Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20620535 ↗MalfunctionOct 16, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. CUSTOMER DECLINED FOLLOW UP FROM TANDEM TECHNICAL SUPPORT. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20460701 ↗MalfunctionSep 18, 2024Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CONTINUOUS GLUCOSE MONITOR ERROR 42. THE PUMP WAS RESET, AND THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20246386 ↗MalfunctionSep 6, 2024Battery Problem; Battery Problem
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20164285 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P140015 on Sep 8, 2015. Submitted by Tandem Diabetes Care, Inc. as “T:Slim G4 Insulin Pump With Dexcom G4 Platinum Cgm”. The FDA review took 418 days. The FDA has since approved 29 supplements (changes to the device or its labeling), most recently on Dec 19, 2019. FDA record ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
FDA definition of this device type: The insulin pump is intended to be used as a component of an invasive glucose device. The pump is designed to pump fluid (insulin) into a patient in a controlled manner as an aid in the mangement of diabetes mellitus in persons requiring insulin. It is intended to accept and display data from the glucose sensor.
Recalls under the same FDA submission
These recalls involve devices cleared under P140015. One submission often covers many models, so check the product and lot details in each record.
- Z-2471-2018 · Apr 23, 2018 · Open, Classifiedt:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.
Common questions about the t:slim X2 5.0 Insulin Pump 1000897
Has the t:slim X2 5.0 Insulin Pump 1000897 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1811-2022 (Class II), began on May 24, 2022: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the t:slim X2 5.0 Insulin Pump 1000897?
The FDA's MAUDE database lists 6,444 adverse event reports for this device, received between Jul 9, 2018 and Jun 8, 2026. In those reports, the most common specific device problems are “Charging Problem” (1,064), “Pumping Stopped” (943) and “Incorrect Measurement” (927); the most common patient problems are “Hyperglycemia” (438), “Hypoglycemia” (384) and “Elevated Ketones/ Diabetic Ketoacidosis” (87). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the t:slim X2 5.0 Insulin Pump 1000897?
2 reports are classified as deaths, 857 as injuries and 5,585 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the t:slim X2 5.0 Insulin Pump 1000897 FDA approved?
It was approved through premarket approval (PMA) P140015 on Sep 8, 2015.
Is the t:slim X2 5.0 Insulin Pump 1000897 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the t:slim X2 5.0 Insulin Pump 1000897, and is it still sold?
The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 5.0 Insulin Pump, model 1000897). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor devices
- t:slim X2 5.0 Insulin Pump 1000885Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall2 deaths17,224 reports
- t:slim X2 5.0 Insulin Pump 1000911Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls2 deaths15,927 reports
- t:slim X2 5.x Insulin Pump 1003808Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor2 recalls1 death2,525 reports
- t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM 006194Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor4 deaths2,272 reports
- Animas Vibe 11451100Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor2 deaths2,262 reports
- t:slim X2 5.0 Insulin Pump 1000891Tandem Diabetes Care, Inc. · Pump, Infusion, Insulin, To Be Used With Invasive Glucose Sensor1 recall1,561 reports
- Animas Vibe 11451000Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor1,122 reports
- Animas Vibe 11451200Animas Corporation · Pump, infusion, insulin, to be used with invasive glucose sensor1,064 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page