Skip to content

SOLOPATH STFI-1925Recalls & reported problems

Onset Medical Corporation · SOLOPATH Balloon Expandable TransFemoral System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00852381004265), received between Apr 16, 2019 and Apr 16, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Patient Involvement2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE SOLOPATH DEVICE WAS DAMAGED WHEN THEY TOOK THE SOLOPATH OUT OF THE PACKAGE. THE SILICONE WAS STILL CONNECTED WITH THE SHEATH; HOWEVER, THERE WAS TOO LITTLE SILICONE COVERING THE DISTALLY PART OF THE SHEATH. THERE WERE SOME SHARP EDGES AT THE TRANSITION OF THE SHEATH/TIP. THE DOCTOR TOOK ANOTHER SOLOPATH, WHICH ALSO HAD THE SAME ISSUE. THERE WAS NO PATIENT INVOLVEMENT; THE EVENT OCCURRED BEFORE USE.

    Patient problems: No Patient Involvement

    MAUDE report 8518339 ↗
  • MalfunctionApr 16, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE SOLOPATH DEVICE WAS DAMAGED WHEN THEY TOOK THE SOLOPATH OUT OF THE PACKAGE. THE SILICONE WAS STILL CONNECTED WITH THE SHEATH; HOWEVER, THERE WAS TOO LITTLE SILICONE COVERING THE DISTALLY PART OF THE SHEATH. THERE WERE SOME SHARP EDGES AT THE TRANSITION OF THE SHEATH/TIP. THE DOCTOR TOOK ANOTHER SOLOPATH, WHICH ALSO HAD THE SAME ISSUE. THERE WAS NO PATIENT INVOLVEMENT; THE EVENT OCCURRED BEFORE USE.

    Patient problems: No Patient Involvement

    MAUDE report 8518337 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates