Sechrist Industries, Inc. 3300H-00Recalls & reported problems
Sechrist Industries, Inc. · HBO CHMBR,ASSY,33IN,ENGLISH,BLUE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00852682007637), received between Jan 31, 2025 and Jan 31, 2025 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Use of Device Problem2
Patient problems
- Insufficient Information2
- Burn(s)1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 31, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
ON (B)(6) 2025, SECHRIST BECAME AWARE OF A DEATH, TO A CHILD, THAT WAS BEING TREATED IN A HYPERBARIC CHAMBER. HIS MOM, WHO WAS NEXT TO THE CHAMBER AT THE TIME, ALSO SUFFERED INJURIES TO HER ARMS. OFFICIALS IN TROY, MICHIGAN ARE INVESTIGATING THE MATTER.
Patient problems: Insufficient Information
MAUDE report 21289215 ↗DeathJan 31, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
ON JANUARY 31, 2025, SECHRIST BECAME AWARE OF A DEATH, TO A CHILD, THAT WAS BEING TREATED IN A HYPERBARIC CHAMBER. OFFICIALS IN (B)(6), MICHIGAN CONFIRMED A CHILD DIED AFTER THE INCIDENT.
Patient problems: Burn(s); Insufficient Information
MAUDE report 21289171 ↗
Other Chamber, Hyperbaric devices
- Sechrist Industries, Inc. 3200-00Sechrist Industries, Inc. · Chamber, Hyperbaric4 reports
- Sechrist Industries, Inc. 3600HR-00-1Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 3600ER-00Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 4100HRP-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-H-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600E-00Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Vitaeris 320 15C37179Hyperbaric Technologies, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-00-0Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page