Vitaeris 320 15C37179Recalls & reported problems
Hyperbaric Technologies, Inc. · hyperbaric chamber · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10856039006011), received between Jan 11, 2018 and Jan 11, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Mechanics Altered” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Mechanics Altered1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2018Mechanics Altered
THIS INVOLVES THE OXYHEALTH VITAERIS320 PORTABLE HYPERBARIC OXYGEN CHAMBER. THERE WAS NO PROBLEM WITH THE CHAMBER'S FUNCTIONALITY BUT A DESIGN FLAW WITH THE COVER CAUSED WHAT COULD HAVE BEEN A DANGEROUS SITUATION, AT THE END OF A PERIOD OF SELF-TREATING I ATTEMPTED TO EMERGE FROM THE CHAMBER BUT COULD NOT TURN THE PRESSURE RELEASE VALVE. ULTIMATELY I HAD TO EXERT NEARLY MY FULL STRENGTH TO OPEN THE VALVE WHILE LYING ON MY BACK. AFTER EXITING I DISCOVERED THAT SEVERAL THREADS WHERE THE COVER HAD BEEN SEWN TOGETHER HAD COME LOOSE AND CONTACTED THE VALVE STEM. WHEN I CLOSED THE VALVE FROM INSIDE THE CHAMBER, THE THREADS FROM THE COVER WRAPPED AROUND THE VALVE STEM WHICH, AS I CLOSED IT CREATED A TIGHT A SEAL AROUND THE STEM WHICH MADE IT ALMOST IMPOSSIBLE TO RELEASE THE VALVE. A PERSON OF LESSER STRENGTH THAN I, A SMALL WOMAN FOR EXAMPLE, WOULD POSSIBLY HAVE NOT BEEN ABLE TO EXIT THE CHAMBER UNTIL SOMEONE FROM OUTSIDE THE CHAMBER WORKED THE VALVE LOOSE. THIS IS A DANGER FOR A PERSON WHO CHOOSES TO SELF TREAT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7186485 ↗
Other Chamber, Hyperbaric devices
- Sechrist Industries, Inc. 3200-00Sechrist Industries, Inc. · Chamber, Hyperbaric4 reports
- Sechrist Industries, Inc. 3600HR-00-1Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 3600ER-00Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 3300H-00Sechrist Industries, Inc. · Chamber, Hyperbaric1 death2 reports
- Sechrist Industries, Inc. 4100HRP-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-H-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600E-00Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-00-0Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page