Sechrist Industries, Inc. 3200-00Recalls & reported problems
Sechrist Industries, Inc. · HBO CHMBER,32IN,PNMTC,ENGLISH,BLUE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00852682007903), received between Mar 3, 2025 and Jan 12, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Key or Button Unresponsive/not Working” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Key or Button Unresponsive/not Working2
- Fail-Safe Did Not Operate1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2026Key or Button Unresponsive/not Working
CUSTOMER REPORTED THE EMERGENCY VENT BUTTON IS NOT WORKING TO PERFORM EMERGENCY DECOMPRESSION TEST. CUSTOMER ALSO REPORTED THE TIMING IS OUT OF SPEC FOR COMPRESSION. NO PATIENT INCIDENT WAS REPORTED. ISSUE WAS DISCOVERED AT TIME OF ROUTINE TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24048671 ↗MalfunctionOct 21, 2025Protective Measures Problem; Protective Measures Problem
CUSTOMER REPORTED THAT DURING PRE-CLINICAL CHECK THE CHAMBER IS FALLING OUTSIDE THE 120 SECOND TIME FRAME TO BE IN SPEC FOR THE EMERGENCY VENTING. NO PATIENT INVOLVEMENT OR PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23353006 ↗MalfunctionApr 18, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate
DURING PM SERVICE, SECHRIST INDUSTRIES TECHNICIAN DISCOVERED THAT THE EMERGENCY VENT BUTTON WAS NOT WORKING PROPERLY. FACILITY NEVER REPORTED TO SECHRIST. EMERGENCY VENT BUTTON WOULD ONLY WORK WHEN PUSH EM VENT BUTTON AND TURN MASTER VALVE TO EM VENT POSITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21868147 ↗MalfunctionMar 3, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
AS REPORTED BY THE CUSTOMER, WHEN DOING WEEKLY CHECKS, NOTICED EMERGENCY VENT BUTTON WAS NOT WORKING PROPERLY. WHEN BUTTON PUSHED AND HELD DOWN THE DECOMPRESSION DID NOT BEGIN, INSTEAD HAD TO PUSH 4 OR 5 TIMES BEFORE DECOMPRESSION STARTED. NO PATIENT INVOLVEMENT OR INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21509616 ↗
Other Chamber, Hyperbaric devices
- Sechrist Industries, Inc. 3600HR-00-1Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 3600ER-00Sechrist Industries, Inc. · Chamber, Hyperbaric2 reports
- Sechrist Industries, Inc. 3300H-00Sechrist Industries, Inc. · Chamber, Hyperbaric1 death2 reports
- Sechrist Industries, Inc. 4100HRP-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-H-01-1Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600E-00Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
- Vitaeris 320 15C37179Hyperbaric Technologies, Inc. · Chamber, Hyperbaric1 report
- Sechrist Industries, Inc. 3600H-00-0Sechrist Industries, Inc. · Chamber, Hyperbaric1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page