Skip to content

t:slim X2 insulin pump recallZ-1811-2022 · Tandem Diabetes Care Inc · initiated May 24, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tandem Diabetes Care Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

FDA root cause category: Device Design

What the company is doing

On 5/24/22, and 7/7/22, correction notices sent to customers in Australia and New Zealand to update to the latest software. t:slim X2 insulin pump with Basal-IQ technology version 6.6 and Control-IQ technology version 7.6 available with latest software versions. Software update contact: Australia, AMSL: [email protected] or 1300 851 056; New Zealand, NZMS [email protected] (Questions: 0508 634 103) On 7/26/22, correction notices were mailed or emailed in the United States to healthcare providers and distributors. Recalling firm contact: [email protected] Providers asked to instruct patients to update pump software. Software update called version 6.6.0.1 for Basal-IQ technology and version 7.6.0.1 for Control-IQ technology. Check software version on the pump: options > my pump > pump info and arrow down to software version. See user guide and to visit https://support.tandemdiabetes.com/hc/en-us/articles/1500011452302-How-do-I-check-what-software-version-is-on-my-pump-video- Latest software can be downloaded remotely from a computer. Visit https://portal.tandemdiabetes.com and login with their t:connect account credentials. Once logged in, click the MY DEVICES to view available software updates. Perform software update via the Tandem Device Updater (TDU), with precautions and instructions provided in user TDU guide: https://www.tandemdiabetes.com/docs/default-source/product-documents/tandem-device-updater/aw-1004796_c-user-guide-tandem-device-updater-5-4-artwork.pdf. Acknowledge correction notice receipt: https://go.tandemdiabetes.com/Field-Correction-Notice-X.6_US.html Distributors with impacted part numbers asked to contact the recalling firm s distribution team. On 8/4/22, users sent correction notice to update pump software. See above for acknowledgment form link and software update instructions. Prescription and training may be required if pump is out-of-warranty and not currently using Control-IQ technology, can only update to Basal-

Recall details

Recall number
Z-1811-2022
Status
Open, Classified
Initiated by the company
May 24, 2022
Posted by the FDA
Sep 26, 2022
Recalling firm
Tandem Diabetes Care Inc
Quantity
31,100
Product code
PQF
Distribution
US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa

Product description

t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Pump/UDI-DI/Software Versions/Distributor Instructions: t:slim X2 insulin pump/00852162004781, 00852162004798, 00853052007240/ 4.3.5.2, 4.3.6/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Dexcom G5/853052007257, 853052007271, 853052007295, 853052007318, 853052007967, 853052007981/ 5.0.1, 5.0.4, 5.3.1.1/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Basal-IQ technology/853052007264, 853052007288, 853052007301, 853052007325, 850006613373, 850006613380, 850006613700, 850006613717, 850006613724, 850006613731/ 6.0.4, 6.3.1.1, 6.4/ For Distributors: 1) Parts: 1000886, 1000898, 1004219, 1005706, 1006373 or 1006375, Contact Recalling Firm to Switch Out, 2) Parts: 1010004, 1010005,1010006, No Action Required, Software is Up to Date; t:slim X2 insulin pump with Control-IQ technology/ 850006613229, 850006613205, 850006613212, 850006613762, 850006613779, 850006613786, 850006613793/ 7.3.1.1, 7.4/ For Distributors: 1) Parts: 1005011, 1005012, 1006402 or 1006404, Contact Recalling Firm to Switch Out, 2) Parts: 1010009,1010010, 1010011, No Action Required, Software is Up to Date

Devices named in this recall

26 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page