Skip to content

Lapiplasty Lesser TMT Fixation Pack SK28Recalls & reported problems

Treace Medical Concepts, Inc. · Anatomic Fixation · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Treace Lapiplasty fixation plate and screw pack is used to hold bones together in bunion surgery. Reports mostly describe pain and swelling after surgery, hardware that had to be removed because it was irritating the foot, and screws backing out or breaking, with one broken plate found on follow-up imaging. There are 33 reports, 20 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00853114006204), received between May 12, 2022 and Mar 12, 2026 — classified as 20 injuries, 13 malfunctions.

The most frequently reported specific device problem is “Break” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem19
  • Break12
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Migration or Expulsion of Device2

Patient problems

  • Pain13
  • Swelling/ Edema6
  • Foreign Body In Patient5
  • No Clinical Signs, Symptoms or Conditions5
  • Nonunion/Delayed-union Bone Fracture5
  • Discomfort2
  • Skin Inflammation/ Irritation2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 12, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2025, ALL HARDWARE WAS REMOVED ON (B)(6) 2026 DUE TO PAIN. NO DEFICIENCIES OR MALFUNCTIONS WERE ALLEGED/FOUND WITH ANY TMC DEVICE, AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 24578898 ↗
  • InjuryFeb 11, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, ALL HARDWARE WAS REMOVED ON (B)(6) 2025 DUE TO PAIN. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 24318347 ↗
  • InjuryJan 20, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, ALL HARDWARE WAS REMOVED ON (B)(6) 2025 DUE TO PAIN AND SWELLING. THE PATIENT'S BONES WERE COMPLETELY FUSED/HEALED. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 24117224 ↗
  • InjuryAug 29, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, ALL HARDWARE WAS REMOVED ON (B)(6) 2025 DUE TO HARDWARE PAIN AND DISCOMFORT AT THE PLANTAR AND DORSAL MIDFOOT. THE SCREWS WERE BACKING OUT SLIGHTLY AND THERE WAS HYPERTROPHIC BONE FORMATION AROUND THE HARDWARE CAUSING PAIN AT THE FIRST TMT INCISION. NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 22926860 ↗
  • MalfunctionFeb 3, 2025Break; Break

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, A BROKEN PLATE AT THE THIRD TMT WAS DISCOVERED VIA A RADIOGRAPHIC IMAGE FROM A 6-MONTH POST-OP FOLLOW-UP. NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT. ALTHOUGH THERE HAS BEEN NO INJURY REPORTED AND NO INFORMATION HAS BEEN RECEIVED INDICATING THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY TO DATE, THERE HAVE BEEN SIMILAR EVENTS WHERE THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY. THEREFORE, AN MDR WILL BE FILED TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21300706 ↗
  • InjurySep 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ALL HARDWARE WAS REMOVED ON (B)(6) 2024 DUE TO HARDWARE IRRITATION. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 20165302 ↗
  • InjurySep 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ALL HARDWARE WAS REMOVED ON (B)(6) 2024 DUE TO HARDWARE IRRITATION. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 20165301 ↗
  • MalfunctionJul 13, 2024Break; Break

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, THE PATIENT IS EXPERIENCING A POTENTIAL NON-UNION, AND HAS PAIN AND SWELLING AT THE FIRST TMT JOINT. RADIOGRAPHIC IMAGES FROM A POST-OP FOLLOW-UP SHOWS A SCREW IS BACKING OUT OF THE PLATE AND IS ALSO BROKEN. NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT. ALL HARDWARE CURRENTLY REMAINS IMPLANTED. ALTHOUGH THERE HAS BEEN NO INJURY REPORTED AND THIS MALFUNCTION HAS NOT RESULTED IN A REVISION SURGERY TO DATE, THERE HAVE BEEN SIMILAR EVENTS WHERE THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY. THEREFORE, A MDR WILL BE FILED TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Swelling/ Edema

    MAUDE report 19737148 ↗
  • MalfunctionJul 13, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2024, THE PATIENT IS EXPERIENCING A POTENTIAL NON-UNION, AND HAS PAIN AND SWELLING AT THE FIRST TMT JOINT. RADIOGRAPHIC IMAGES FROM A POST-OP FOLLOW-UP SHOWS A SCREW IS BACKING OUT OF THE PLATE. NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT. ALL HARDWARE CURRENTLY REMAINS IMPLANTED. ALTHOUGH THERE HAS BEEN NO INJURY REPORTED AND THIS MALFUNCTION HAS NOT RESULTED IN A REVISION SURGERY TO DATE, THERE HAVE BEEN SIMILAR EVENTS WHERE THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY. THEREFORE, A MDR WILL BE FILED TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain; Swelling/ Edema

    MAUDE report 19737146 ↗
  • MalfunctionMar 29, 2024Break; Migration or Expulsion of Device; Break; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ONE BROKEN PLATE AND TWO PLATES THAT MAY POTENTIALLY BE MIGRATING WERE DISCOVERED DURING A POST-SURGERY FOLLOW-UP ON (B)(6) 2024. ALTHOUGH THERE HAS BEEN NO INJURY REPORTED AND NO INFORMATION HAS BEEN RECEIVED INDICATING THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY TO DATE, THERE HAVE BEEN SIMILAR EVENTS WHERE THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY. THEREFORE, AN MDR WILL BE FILED TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19004523 ↗
  • MalfunctionMar 29, 2024Break; Migration or Expulsion of Device; Break; Migration or Expulsion of Device

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ONE BROKEN PLATE AND TWO PLATES THAT MAY POTENTIALLY BE MIGRATING WERE DISCOVERED DURING A POST-SURGERY FOLLOW-UP ON (B)(6) 2024. ALTHOUGH THERE HAS BEEN NO INJURY REPORTED AND NO INFORMATION HAS BEEN RECEIVED INDICATING THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY TO DATE, THERE HAVE BEEN SIMILAR EVENTS WHERE THIS MALFUNCTION HAS RESULTED IN A REVISION SURGERY. THEREFORE, AN MDR WILL BE FILED TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19004522 ↗
  • InjuryJan 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ALL HARDWARE WAS REMOVED ON (B)(6) 2023 DUE TO PAIN AT THE IMPLANT SITE. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 18438269 ↗
  • InjuryDec 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ALL HARDWARE WAS REMOVED ON 11-27-2023 DUE TO UNCOMFORTABLE HARDWARE. ALTHOUGH AN ADDITIONAL PROCEDURE WAS PERFORMED AT THE SAME TIME TO ADDRESS A RUPTURED TENDON, IT WAS UNRELATED TO THE INITIAL BUNION SURGERY. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort

    MAUDE report 18394693 ↗
  • InjuryDec 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023, ALL HARDWARE WAS REMOVED ON (B)(6) 2023 DUE TO UNCOMFORTABLE HARDWARE. ALTHOUGH AN ADDITIONAL PROCEDURE WAS PERFORMED AT THE SAME TIME TO ADDRESS A RUPTURED TENDON, IT WAS UNRELATED TO THE INITIAL BUNION SURGERY. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort

    MAUDE report 18394686 ↗
  • InjuryNov 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2022, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON (B)(6) 2023 DUE TO PAIN. NO DEFICIENCIES OR MALFUNCTIONS WERE ALLEGED WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 18148685 ↗
  • MalfunctionAug 2, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF AN OLIVE WIRE DRILL BROKE OFF AND WAS RETAINED IN THE PATIENT'S BONE DURING A BUNION PROCEDURE ON (B)(6) 2023. ALTHOUGH THE COMPANY HAS DETERMINED THAT THE SUBJECT EVENT IN THIS MDR IS UNLIKELY TO RESULT IN A SERIOUS INJURY IF IT RECURRED, THE COMPANY HAS DECIDED TO FILE THE MDR IN AN ABUNDANCE OF CAUTION AND TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17445807 ↗
  • MalfunctionJun 22, 2023Break; Break

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2023 HARDWARE (1 PLATE AND 4 SCREWS) WAS REMOVED DUE TO A BROKEN PLATE. THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17181596 ↗
  • InjuryJun 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON (B)(6) 2023 DUE TO SWELLING. THE PATIENT'S BONES WERE COMPLETELY FUSED/HEALED. UPON REMOVAL, THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Swelling/ Edema

    MAUDE report 17172530 ↗
  • InjuryJun 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A BUNION SURGERY ON (B)(6) 2022, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON (B)(6) 2023 DUE TO 3 BROKEN SCREWS. UPON HARDWARE REMOVAL, IT WAS CONFIRMED THAT 2 OUT OF THE 3 INTENDED BONES WERE COMPLETELY FUSED/HEALED WITH A NON-UNION IN ONLY ONE. THE PATIENT WAS REVISED WITH TMC HARDWARE, ALONG WITH THE USE OF A BONE GRAFT DOWEL. THERE WAS NO OTHER REPORT OF ANY PATIENT IMPACT OR INJURY AS A RESULT OF THIS EVENT.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17091098 ↗
  • InjuryJun 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A BUNION SURGERY ON (B)(6) 2022, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON (B)(6) 2023 DUE TO 3 BROKEN SCREWS. UPON HARDWARE REMOVAL, IT WAS CONFIRMED THAT 2 OUT OF THE 3 INTENDED BONES WERE COMPLETELY FUSED/HEALED WITH A NON-UNION IN ONLY ONE. THE PATIENT WAS REVISED WITH TMC HARDWARE, ALONG WITH THE USE OF A BONE GRAFT DOWEL. THERE WAS NO OTHER REPORT OF ANY PATIENT IMPACT OR INJURY AS A RESULT OF THIS EVENT.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 17090833 ↗
  • InjuryApr 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON (B)(6) 2023 DUE TO PAIN. UPON REMOVAL, THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 16789982 ↗
  • InjuryApr 21, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY, ALL HARDWARE WAS REMOVED IN A REVISION SURGERY ON ON (B)(6) 2023 DUE TO PAIN. UPON REMOVAL, THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 16789910 ↗
  • MalfunctionApr 7, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF A PLATE TACK BROKE OFF AND WAS RETAINED IN THE PATIENT'S BONE DURING A BUNION PROCEDURE ON (B)(6) 2023. ALTHOUGH THE COMPANY HAS DETERMINED THAT THE SUBJECT EVENT IN THIS MDR IS UNLIKELY TO RESULT IN A SERIOUS INJURY IF IT RECURRED, THE COMPANY HAS DECIDED TO FILE THE MDR IN AN ABUNDANCE OF CAUTION AND TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: Foreign Body In Patient

    MAUDE report 16704858 ↗
  • InjuryApr 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER AN INITIAL BUNION SURGERY ON (B)(6) 2021, A PLATE WAS REMOVED IN A REVISION SURGERY ON 03/08/2023 DUE TO PAIN. UPON REMOVAL, THERE WERE NO DEFICIENCIES OR MALFUNCTIONS ALLEGED/FOUND WITH ANY TMC DEVICE AND NO OTHER PATIENT OUTCOMES WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain

    MAUDE report 16704781 ↗
  • MalfunctionJan 28, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF AN OLIVE WIRE DRILL BROKE OFF AND WAS RETAINED IN THE PATIENT'S BONE DURING A BUNION PROCEDURE ON (B)(6) 2022. ALTHOUGH THE COMPANY HAS DETERMINED THAT THE SUBJECT EVENT IN THIS MDR IS UNLIKELY TO RESULT IN A SERIOUS INJURY IF IT RECURRED, THE COMPANY HAS DECIDED TO FILE THE MDR IN AN ABUNDANCE OF CAUTION AND TO ENSURE FULL COMPLIANCE WITH 21 CFR PART 803.

    Patient problems: Foreign Body In Patient

    MAUDE report 16261640 ↗