LUMINELLE DTx Hysteroscopy System LumDTX-ARecalls & reported problems
Uvision 360, Inc. · Hysteroscopy System indicated for both Hysteroscopy and Cystoscopy to permit direct viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00853277008046), received between Mar 21, 2023 and Aug 15, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Erratic or Intermittent Display” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Erratic or Intermittent Display2
- Connection Problem1
- Deformation Due to Compressive Stress1
- Device Handling Problem1
- Image Display Error/Artifact1
- Improper or Incorrect Procedure or Method1
- Material Twisted/Bent1
- No Display/Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2023Material Twisted/Bent; Material Twisted/Bent
THE CUSTOMER REPORTED THAT THEY NOTICED THE BLUE CASING OF THE SCOPE IS STARTING TO TEAR. THEY REPORTED THE SCOPE HAD ONLY BEEN USED IN TWO CASES. THE CUSTOMER PROVIDED PHOTOS OF THE PHYSICAL DAMAGE TO THE SCOPE STICK. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552750 ↗MalfunctionAug 15, 2023Erratic or Intermittent Display; No Display/Image; Deformation Due to Compressive Stress; Erratic or Intermittent Display; No Display/Image; Deformation Due to Compressive Stress
THE CUSTOMER REPORTED THAT THE LOANER SCOPE WAS DAMAGED DURING A PATIENT PROCEDURE. THE CUSTOMER EXPLAINED THAT THERE WAS SEVERE DAMAGE TO THE BLUE SCOPE STICK, BUT DID NOT SPECIFY HOW THE DAMAGE OCCURRED. THERE WERE NO PATIENT ADVERSE EVENTS DUE TO THIS MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552749 ↗MalfunctionAug 15, 2023Erratic or Intermittent Display; Image Display Error/Artifact; Improper or Incorrect Procedure or Method; Erratic or Intermittent Display; Image Display Error/Artifact; Improper or Incorrect Procedure or Method
THE CUSTOMER REPORTED THAT THEY WERE EXPERIENCING INTERMITTENT IMAGES FROM THE SCOPE ON THE MONITOR. THE CUSTOMER SENT IMAGES OF A PHYSICAL DAMAGE THEY NOTICE AS WELL. CLEAR DAMAGE TO THE FIBER OPTIC STICK WAS VISIBLE IN THE IMAGES. THE FIBER OPTIC CABLE WAS VISUALLY AND FUNCTIONALLY TESTED TO CONFIRM THE PHYSICAL AND FUNCTIONAL DAMAGE SEEN BY THE CUSTOMER. THE USAGE COMMUNICATED BY THE CUSTOMER WAS RECREATED IN AN ATTEMPT TO RECREATE THE DAMAGE. THE DAMAGE COULD NOT BE RECREATED USING THE SAME INSTRUMENTATION USED WITH THE DEVICE. THERE WAS NO REPORT OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17552729 ↗MalfunctionMar 21, 2023Device Handling Problem; Device Handling Problem
THE CUSTOMER REPORTED DURING A PHONE CALL THAT THEY WERE UNABLE TO PRODUCE AN IMAGE IN THEIR MONITOR AFTER ASSEMBLING THE SYSTEM. THE CUSTOMER INDICATED THAT THE SYSTEM FAILURE WAS CAUGHT PRIOR TO BE BEING USED WITH THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587911 ↗MalfunctionMar 21, 2023Connection Problem; Premature Separation; Connection Problem; Premature Separation
SCOPE AND SHEATH(S) DO NOT CONSISTENTLY ATTACH TO THE DEVICE. WHILE CONDUCTING PROCEDURES, THE SCOPE AND SHEATH(S) BECOME DISLODGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587901 ↗MalfunctionMar 21, 2023Poor Quality Image; Poor Quality Image
IN A PRE-SALE DEMONSTRATION, THE SURGEON WAS USING THE LUMINELLE DTX SYSTEM FOR A CYSTOSCOPY PROCEDURE AND THE SALES REPRESENTATIVE REPORTED THAT THE VISIBILITY WAS VERY DARK. THE SALES REPRESENTATIVE AND SURGEON SAID THAT IT WAS DIFFICULT TO VIEW THE BLADDER UNLESS THE SCOPE WAS VERY CLOSE TO THE BLADDER WALL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16587900 ↗
Other Hysteroscope (And Accessories) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.